盐酸金刚烷胺胶囊%高效液相色谱-蒸发光检测器%含量测定
鹽痠金剛烷胺膠囊%高效液相色譜-蒸髮光檢測器%含量測定
염산금강완알효낭%고효액상색보-증발광검측기%함량측정
amantadine Hydrochloride capsules%High performance liquid
目的:探索盐酸金刚烷胺胶囊的含量测定方法。方法应用 HPlc-elsd 法对盐酸金刚烷胺胶囊含量进行测定,色谱柱为 c18(4.6 mm×250 mm,5μm),流动相为乙腈-0.06%的三氟乙酸(20:80),柱温为30℃,进样量为10μl,流速为1.0 ml· min-1。检测器:elsd ,漂移管温度:65℃,雾化气体:空气,载气流量:2.8 l·min-1。结果盐酸金刚烷胺在(39.62-237.72μg·ml-1)范围内进样量与峰面积的对数线性关系好(r=0.9993);平均回收率:100.2%,(rsd=1.2%)。结论 HPlc-elsd 法样品处理便捷,准确,专属性强,是盐酸金刚烷胺胶囊的质量控制的优选。
目的:探索鹽痠金剛烷胺膠囊的含量測定方法。方法應用 HPlc-elsd 法對鹽痠金剛烷胺膠囊含量進行測定,色譜柱為 c18(4.6 mm×250 mm,5μm),流動相為乙腈-0.06%的三氟乙痠(20:80),柱溫為30℃,進樣量為10μl,流速為1.0 ml· min-1。檢測器:elsd ,漂移管溫度:65℃,霧化氣體:空氣,載氣流量:2.8 l·min-1。結果鹽痠金剛烷胺在(39.62-237.72μg·ml-1)範圍內進樣量與峰麵積的對數線性關繫好(r=0.9993);平均迴收率:100.2%,(rsd=1.2%)。結論 HPlc-elsd 法樣品處理便捷,準確,專屬性彊,是鹽痠金剛烷胺膠囊的質量控製的優選。
목적:탐색염산금강완알효낭적함량측정방법。방법응용 HPlc-elsd 법대염산금강완알효낭함량진행측정,색보주위 c18(4.6 mm×250 mm,5μm),류동상위을정-0.06%적삼불을산(20:80),주온위30℃,진양량위10μl,류속위1.0 ml· min-1。검측기:elsd ,표이관온도:65℃,무화기체:공기,재기류량:2.8 l·min-1。결과염산금강완알재(39.62-237.72μg·ml-1)범위내진양량여봉면적적대수선성관계호(r=0.9993);평균회수솔:100.2%,(rsd=1.2%)。결론 HPlc-elsd 법양품처리편첩,준학,전속성강,시염산금강완알효낭적질량공제적우선。
Objective to establish a method for determination of the content of amantadine Hydrochloride capsules. Methods HPlc-elsd method was used to determine the content of amantadine Hydrochloride capsules. the column was c18 (4.6 mm * 250 mm, 5 m), the mobile phase was three -0.06% trifluoroacetic acid acetonitrile (20:80), the column temperature was 30 ℃ , the injection volume was 10 L, the flow rate was 1 mL - min - 1. Detector: ELSD, drift tube temperature: 65 ℃ , atomization gas: air, carrier gas flow rate: 2.8 L - min - 1. Results there was a good log linear relationship (r=0.9993) between the sample volume and peak area of amantadine hydrochloride in (39.62-237.72 G - ml - 1); the average recovery rate: 100.2%, (rsd=1.2%). Conclusion Treating sample with HPLC-ELSD method is convenient, accurate, specific, top exclusive and it is preferred for quality control in amantadine Hydrochloride capsules.