国际检验医学杂志
國際檢驗醫學雜誌
국제검험의학잡지
INTERNATIONAL JOURNAL OF LABORATORY MEDICINE
2014年
9期
1177-1178,1180
,共3页
潘婕%罗浔阳%胡晓菡%张葵
潘婕%囉潯暘%鬍曉菡%張葵
반첩%라심양%호효함%장규
精密度%准确度%线性范围%生化分析仪%交叉污染
精密度%準確度%線性範圍%生化分析儀%交扠汙染
정밀도%준학도%선성범위%생화분석의%교차오염
precision%accuracy%linear range%biochemical analyzer%cross-contamination
目的:评价OLYMPUS AU5421型全自动生化分析仪电解质模块测定K+、Na+、Cl-的性能。方法采用OLYM-PUS AU5421型全自动生化分析仪电解质模块测定K+、Na+、Cl-,并进行其检测精密度、准确度、线性范围、交叉污染评价及仪器检测结果的比对。依据美国临床实验室标准化协会(CLSI)EP6-A、EP9-A2文件方法,计算线性回归方程和回归系数。结果K+、Na+、Cl-的批内精密度分别为0.328%、0.306%、0.343%,批间精密度分别为0.780%、1.012%、0.985%,线性方程回归系数均大于0.999;其交叉污染率分别为1.066%、0.812%及0.858%。结论 OLYMPUS AU5421型全自动生化分析仪电解质模块具有精密度、准确度、线性范围优良,交叉污染率低的优点。
目的:評價OLYMPUS AU5421型全自動生化分析儀電解質模塊測定K+、Na+、Cl-的性能。方法採用OLYM-PUS AU5421型全自動生化分析儀電解質模塊測定K+、Na+、Cl-,併進行其檢測精密度、準確度、線性範圍、交扠汙染評價及儀器檢測結果的比對。依據美國臨床實驗室標準化協會(CLSI)EP6-A、EP9-A2文件方法,計算線性迴歸方程和迴歸繫數。結果K+、Na+、Cl-的批內精密度分彆為0.328%、0.306%、0.343%,批間精密度分彆為0.780%、1.012%、0.985%,線性方程迴歸繫數均大于0.999;其交扠汙染率分彆為1.066%、0.812%及0.858%。結論 OLYMPUS AU5421型全自動生化分析儀電解質模塊具有精密度、準確度、線性範圍優良,交扠汙染率低的優點。
목적:평개OLYMPUS AU5421형전자동생화분석의전해질모괴측정K+、Na+、Cl-적성능。방법채용OLYM-PUS AU5421형전자동생화분석의전해질모괴측정K+、Na+、Cl-,병진행기검측정밀도、준학도、선성범위、교차오염평개급의기검측결과적비대。의거미국림상실험실표준화협회(CLSI)EP6-A、EP9-A2문건방법,계산선성회귀방정화회귀계수。결과K+、Na+、Cl-적비내정밀도분별위0.328%、0.306%、0.343%,비간정밀도분별위0.780%、1.012%、0.985%,선성방정회귀계수균대우0.999;기교차오염솔분별위1.066%、0.812%급0.858%。결론 OLYMPUS AU5421형전자동생화분석의전해질모괴구유정밀도、준학도、선성범위우량,교차오염솔저적우점。
Objective To evaluate the performance of electrolyte module in OLYMPUS AU5421 Automatic Biochemical Analy-zer in K+ ,Na+ ,Cl- detection .Methods Electrolyte module in OLYMPUS AU5421 Automatic Biochemical Analyzer was employed to detect K+ ,Na+ ,Cl- ,and their precision ,accuracy ,linearity range ,cross-contamination and the test results of instruments were compared .The linear regression equation and regression coefficients were calculated according to the methods of Clinical and Labo-ratory Standards Institute(CLSI) EP6-A ,EP9-A2 documents .Results The within-run precisions of K + ,Na+ ,Cl- were 0 .328% , 0 .306% ,0 .343% ,respectively ,and their between-run precisions were 0 .780% ,1 .012% ,0 .985% ,respectively .Their regression coefficients of linear equation were all greater than 0 .999 ,and their cross-contamination rate were 1 .066% ,0 .812% and 0 .858% , respectively .Conclusion Electrolyte module in OLYMPUS AU5421 Automatic Biochemical Analyzer modules has advantages of good precision ,accuracy ,linear range ,and low cross-contamination rate .