眼前段光学相干断层扫描仪%裂隙灯型眼前段光学相干断层扫描仪%OrbscanⅡz角膜地形图/角膜测厚系统%A型超声测量仪%中央角膜厚度
眼前段光學相榦斷層掃描儀%裂隙燈型眼前段光學相榦斷層掃描儀%OrbscanⅡz角膜地形圖/角膜測厚繫統%A型超聲測量儀%中央角膜厚度
안전단광학상간단층소묘의%렬극등형안전단광학상간단층소묘의%OrbscanⅡz각막지형도/각막측후계통%A형초성측량의%중앙각막후도
Visante OCT%SL-OCT%OrbscanⅡz%A-mode ultrasound pachymetry%central corneal thickness
目的:探讨眼前段光学相干断层扫描仪( Visante OCT )、裂隙灯型眼前段光学相干断层扫描仪( SL-OCT )、OrbscanⅡz角膜地形图/角膜测厚系统( OrbscanⅡz)和A型超声测量仪测量近视人群中央角膜厚度( central corneal thickness ,CCT)的差异及相关性,以评价三种光学测量仪器在测量中央角膜厚度的准确性,为其临床应用提供依据。<br> 方法:对近视患者43例86眼分别用Visante OCT、SL-OCT、OrbscanⅡz和A型超声测量仪进行CCT的测量,并对四种仪器测量的CCT值采用单因素方差分析, LSD-t法两两比较,Pearson相关分析。<br> 结果:Visante OCT、SL-OCT、OrbscanⅡz和A型超声测量仪测量 CCT 的平均值分别为(521.05±29.88)μm、(533.73±31.83)μm、(534.02±32.96)μm、(540.60±31.11)μm,95%可信区间分别为(514.57~527.54),(526.83~540.65),(533.85~547.35),(536.75~544.45)。 Visante OCT 与SL-OCT(P=0.009)、A型超声测量仪(P<0.01)、OrbscanⅡz(P=0.008)的CCT测量值差异均有统计学意义。 Visante OCT 的CCT测量值分别低于SL-OCT的CCT测量值(12.69±9.32)μm,低于A型超声测量仪的 CCT 测量值(19.55±8.64)μm,低于OrbscanⅡz的CCT测量值(12.96±12.21)μm。 SL-OCT与A型超声测量仪( P=0.158)、OrbscanⅡz ( P=0.954)的CCT测量值差异均无统计学意义。 OrbscanⅡz与A型超声测量仪的 CCT 测量值差异无统计学意义( P=0.176)。四种方法CCT测量值均呈正相关( P<0.01)。<br> 结论:三种光学测量仪器测量CCT值与A超测量值均有很好的相关性。 OrbscanⅡz和SL-OCT测量CCT值与A超的测量值基本一致。三种光学测量仪器与A超相比具有非接触性、无创性、测量时间短等特点,更易被患者接受。
目的:探討眼前段光學相榦斷層掃描儀( Visante OCT )、裂隙燈型眼前段光學相榦斷層掃描儀( SL-OCT )、OrbscanⅡz角膜地形圖/角膜測厚繫統( OrbscanⅡz)和A型超聲測量儀測量近視人群中央角膜厚度( central corneal thickness ,CCT)的差異及相關性,以評價三種光學測量儀器在測量中央角膜厚度的準確性,為其臨床應用提供依據。<br> 方法:對近視患者43例86眼分彆用Visante OCT、SL-OCT、OrbscanⅡz和A型超聲測量儀進行CCT的測量,併對四種儀器測量的CCT值採用單因素方差分析, LSD-t法兩兩比較,Pearson相關分析。<br> 結果:Visante OCT、SL-OCT、OrbscanⅡz和A型超聲測量儀測量 CCT 的平均值分彆為(521.05±29.88)μm、(533.73±31.83)μm、(534.02±32.96)μm、(540.60±31.11)μm,95%可信區間分彆為(514.57~527.54),(526.83~540.65),(533.85~547.35),(536.75~544.45)。 Visante OCT 與SL-OCT(P=0.009)、A型超聲測量儀(P<0.01)、OrbscanⅡz(P=0.008)的CCT測量值差異均有統計學意義。 Visante OCT 的CCT測量值分彆低于SL-OCT的CCT測量值(12.69±9.32)μm,低于A型超聲測量儀的 CCT 測量值(19.55±8.64)μm,低于OrbscanⅡz的CCT測量值(12.96±12.21)μm。 SL-OCT與A型超聲測量儀( P=0.158)、OrbscanⅡz ( P=0.954)的CCT測量值差異均無統計學意義。 OrbscanⅡz與A型超聲測量儀的 CCT 測量值差異無統計學意義( P=0.176)。四種方法CCT測量值均呈正相關( P<0.01)。<br> 結論:三種光學測量儀器測量CCT值與A超測量值均有很好的相關性。 OrbscanⅡz和SL-OCT測量CCT值與A超的測量值基本一緻。三種光學測量儀器與A超相比具有非接觸性、無創性、測量時間短等特點,更易被患者接受。
목적:탐토안전단광학상간단층소묘의( Visante OCT )、렬극등형안전단광학상간단층소묘의( SL-OCT )、OrbscanⅡz각막지형도/각막측후계통( OrbscanⅡz)화A형초성측량의측량근시인군중앙각막후도( central corneal thickness ,CCT)적차이급상관성,이평개삼충광학측량의기재측량중앙각막후도적준학성,위기림상응용제공의거。<br> 방법:대근시환자43례86안분별용Visante OCT、SL-OCT、OrbscanⅡz화A형초성측량의진행CCT적측량,병대사충의기측량적CCT치채용단인소방차분석, LSD-t법량량비교,Pearson상관분석。<br> 결과:Visante OCT、SL-OCT、OrbscanⅡz화A형초성측량의측량 CCT 적평균치분별위(521.05±29.88)μm、(533.73±31.83)μm、(534.02±32.96)μm、(540.60±31.11)μm,95%가신구간분별위(514.57~527.54),(526.83~540.65),(533.85~547.35),(536.75~544.45)。 Visante OCT 여SL-OCT(P=0.009)、A형초성측량의(P<0.01)、OrbscanⅡz(P=0.008)적CCT측량치차이균유통계학의의。 Visante OCT 적CCT측량치분별저우SL-OCT적CCT측량치(12.69±9.32)μm,저우A형초성측량의적 CCT 측량치(19.55±8.64)μm,저우OrbscanⅡz적CCT측량치(12.96±12.21)μm。 SL-OCT여A형초성측량의( P=0.158)、OrbscanⅡz ( P=0.954)적CCT측량치차이균무통계학의의。 OrbscanⅡz여A형초성측량의적 CCT 측량치차이무통계학의의( P=0.176)。사충방법CCT측량치균정정상관( P<0.01)。<br> 결론:삼충광학측량의기측량CCT치여A초측량치균유흔호적상관성。 OrbscanⅡz화SL-OCT측량CCT치여A초적측량치기본일치。삼충광학측량의기여A초상비구유비접촉성、무창성、측량시간단등특점,경역피환자접수。
AIM:To compare the difference and correlation of central corneal thickness ( CCT ) measured by Zeiss Visante Anterior Segment Optical Coherence Tomography ( Visante OCT ) , Heidelberg Slit -Lamp Anterior Segment Optical Coherence Tomography ( SL-OCT) , OrbscanⅡz Computerized Topography ( OrbscanⅡz ) and A-mode ultrasound pachymetry. <br> METHODS:Forty-three patients with myopia (86 eyes) were included. Sequential measurements of CCT was performed by the same operator with two different anterior segment OCT devices ( VisanteTM , Carl Zeiss, Meditec and Heidelberg EngineeringTM ), OrbscanⅡz Computerized Topography ( Bausch and Lomb IncTM ) and A-mode ultrasound pachymetry.The consistency of measurements was statistically analyzed by one-way ANOVA and LSD pair-wise comparisons.The differences were analyzed by Pearson correlation coefficients. <br> RESULTS:The average CCT measured by Visante OCT, SL -OCT, Orbscan Ⅱ z, and A -mode ultrasound pachymetry were (521.05 ±29.88)μm, (533.73 ±31.83)μm, (534.02 ±32.96 )μm, and ( 540.60 ±31.11 )μm, the 95%confidence interval were (514.57 ~527.54), (526.83 ~540.65), (533.85~547.35), (536.75~544.45).There was statistically significant difference of CCT between using of Visante OCT and SL-OCT ( P =0.009 ). There was statistically significant difference of CCT using of Visante OCT and A-mode ultrasound pachymetry ( P<0.01 ), OrbscanⅡ z ( P =0.008 ). The Visante OCT CCT measurements was significantly less than CCT with SL-OCT ( 12.69 ±9.32 ) μm. Compare with A -mode ultrasound pachymetry and OrbscanⅡz, the Visante OCT measurements significantly underestimated central corneal thickness by a mean of ( 19.55 ±8.64 )μm and (12.96±12.21)μm.There was no significant difference of CCT between using of SL-OCT and A-mode ultrasound pachymetry (P=0.158), OrbscanⅡz (P=0.954).There was no significant difference of CCT between using of OrbscanⅡz and A-mode ultrasound pachymetry ( P=0.176 ) . There were positive correlationship among Visante OCT, SL-OCT, OrbscanⅡz, and A-mode ultrasound pachymetry in CCT measurement ( P <0.01 ) . <br> CONCLUSION: There are excellent correlationship among Visante OCT, SL-OCT,OrbscanⅡz, and A-mode ultrasound pachymetry in CCT measurement.The data shows that the CCT obtained by SL-OCT and OrbscanⅡz do not have significant difference as compared with that from the A-mode ultrasound pachymetry.Three optical measurement devices is more likely to be accepted by patients because of its non-contact, non-invasive, and short measurement time as compared with the A-mode ultrasound pachymetry.