国际眼科杂志
國際眼科雜誌
국제안과잡지
INTERNATIONAL JOURNAL OF OPHTHALMOLOGY
2013年
12期
2475-2478
,共4页
徐肃仲%叶欣杰%许哲%杨纯%沈梅晓
徐肅仲%葉訢傑%許哲%楊純%瀋梅曉
서숙중%협흔걸%허철%양순%침매효
自动分层%超高分辨率OCT%角膜上皮
自動分層%超高分辨率OCT%角膜上皮
자동분층%초고분변솔OCT%각막상피
automatic hierarchy%ultra-high resolution OCT%corneal epithelium
目的:研究基于自动分层算法的超高分辨率谱域 OCT ( UHR-OCT)用于角膜上皮厚度测量的重复性。<br> 方法:使用自行搭建的UHR-OCT系统采集18例健康受试者(年龄23.2±0.9岁)左眼的图像,应用一种基于轴向梯度信息和最短路径的分层算法对得到的SD-OCT图像进行边界探测获取角膜上皮厚度。<br> 结果:自动与手动的上皮分层的位置有很好的相关性(P>0.05,配对t检验)。自动算法获取的6 mm范围平均上皮厚度为52.1±2.1μm,范围从50.3±1.8μm 到53.2±2.4μm,与手动分层之间的角膜上皮厚度差异为0.4±0.3μm。UHR-OCT自动算法重复测量上皮厚度的组内相关系数ICC和可重复性系数CoR分别为0.90±0.05,1.44±0.01μm。<br> 结论:对于超高分辨率OCT,自动分层方法在角膜上皮厚度测量中有着很好的准确性和可重复性。
目的:研究基于自動分層算法的超高分辨率譜域 OCT ( UHR-OCT)用于角膜上皮厚度測量的重複性。<br> 方法:使用自行搭建的UHR-OCT繫統採集18例健康受試者(年齡23.2±0.9歲)左眼的圖像,應用一種基于軸嚮梯度信息和最短路徑的分層算法對得到的SD-OCT圖像進行邊界探測穫取角膜上皮厚度。<br> 結果:自動與手動的上皮分層的位置有很好的相關性(P>0.05,配對t檢驗)。自動算法穫取的6 mm範圍平均上皮厚度為52.1±2.1μm,範圍從50.3±1.8μm 到53.2±2.4μm,與手動分層之間的角膜上皮厚度差異為0.4±0.3μm。UHR-OCT自動算法重複測量上皮厚度的組內相關繫數ICC和可重複性繫數CoR分彆為0.90±0.05,1.44±0.01μm。<br> 結論:對于超高分辨率OCT,自動分層方法在角膜上皮厚度測量中有著很好的準確性和可重複性。
목적:연구기우자동분층산법적초고분변솔보역 OCT ( UHR-OCT)용우각막상피후도측량적중복성。<br> 방법:사용자행탑건적UHR-OCT계통채집18례건강수시자(년령23.2±0.9세)좌안적도상,응용일충기우축향제도신식화최단로경적분층산법대득도적SD-OCT도상진행변계탐측획취각막상피후도。<br> 결과:자동여수동적상피분층적위치유흔호적상관성(P>0.05,배대t검험)。자동산법획취적6 mm범위평균상피후도위52.1±2.1μm,범위종50.3±1.8μm 도53.2±2.4μm,여수동분층지간적각막상피후도차이위0.4±0.3μm。UHR-OCT자동산법중복측량상피후도적조내상관계수ICC화가중복성계수CoR분별위0.90±0.05,1.44±0.01μm。<br> 결론:대우초고분변솔OCT,자동분층방법재각막상피후도측량중유착흔호적준학성화가중복성。
AIM:To assess the repeatability of thickness profiles of corneal epithelium produced by aautomatic hierarchic algorithm that was applied to images obtained by a custom-built ultra-high resolution OCT ( UHR-OCT ) system. <br> METHODS:Eighteen left eyes of 18 subjects (23.2±0.9) years;mean age±standard deviation) were imaged twice on non-consecutive days by a custom-built ultra-high resolution OCT ( UHR -OCT ) system. A hierarchic algorithm based on the axial gradient information and a shortest path search was developed to measure the corneal epithelial thickness profiles from the SD-OCT images. <br> RESULTS: There was good correlation between theautomatic and manual hierarchy positions of the epithelium ( p >0.05, paired t -test ). The average epithelial thickness of the 6mm central corneal region determined by the automatic hierarchic algorithm was 52.1±2.1μm, ranging from 50.3 ±1.8μm to 53.2 ±2.4μm along the horizontal meridian. The epithelial thickness differences between automatic and manual hierarchy by the UHR-OCT systems were 0.4 ±0.3μm.The intraclass correlation coefficients ( ICC ) and coefficients of repeatability ( CoR ) with repeated UHR - OCT measurements for epithelial thickness were 0.90±0.05 and 1.44±0.01μm respectively. <br> CONCLUSION: Regarding UHR -OCT system, the custom -developed automatic hierarchy has good repeatability and accuracy for measuring the thickness of corneal epithelium.