刘卫勇%隋秀芳%康冰飞%刘力%单立奎%叶磊
劉衛勇%隋秀芳%康冰飛%劉力%單立奎%葉磊
류위용%수수방%강빙비%류력%단립규%협뢰
超声检查%探头%声触诊组织定量%一致性
超聲檢查%探頭%聲觸診組織定量%一緻性
초성검사%탐두%성촉진조직정량%일치성
ultrasonography%transducer%virtual touch tissue quantification%agreement
对31例慢性肝炎患者和45例健康志愿者行4C1和9L4探头检测肝实质剪切波速度(SWV)。对照组和慢性肝炎组4C1缺失值出现率均低于9L4,4C1对照组和慢性肝炎组缺失值出现率、9L4对照组和慢性肝炎组缺失值出现率差异均无统计学意义;对照组和慢性肝炎组4C1检测SWV均小于9L4;对照组4C1和9L4组内相关系数(ICC)较低,慢性肝炎组4C1和9L4的 ICC 一般;Bland-Altman 分析4C1和9L4探头 SWV 差值的均数对照组和慢性肝炎组分别为-0.85和-0.64。目前声触诊组织定量(VTQ)技术4C1和9L4探头测量肝实质弹性一致性较差。
對31例慢性肝炎患者和45例健康誌願者行4C1和9L4探頭檢測肝實質剪切波速度(SWV)。對照組和慢性肝炎組4C1缺失值齣現率均低于9L4,4C1對照組和慢性肝炎組缺失值齣現率、9L4對照組和慢性肝炎組缺失值齣現率差異均無統計學意義;對照組和慢性肝炎組4C1檢測SWV均小于9L4;對照組4C1和9L4組內相關繫數(ICC)較低,慢性肝炎組4C1和9L4的 ICC 一般;Bland-Altman 分析4C1和9L4探頭 SWV 差值的均數對照組和慢性肝炎組分彆為-0.85和-0.64。目前聲觸診組織定量(VTQ)技術4C1和9L4探頭測量肝實質彈性一緻性較差。
대31례만성간염환자화45례건강지원자행4C1화9L4탐두검측간실질전절파속도(SWV)。대조조화만성간염조4C1결실치출현솔균저우9L4,4C1대조조화만성간염조결실치출현솔、9L4대조조화만성간염조결실치출현솔차이균무통계학의의;대조조화만성간염조4C1검측SWV균소우9L4;대조조4C1화9L4조내상관계수(ICC)교저,만성간염조4C1화9L4적 ICC 일반;Bland-Altman 분석4C1화9L4탐두 SWV 차치적균수대조조화만성간염조분별위-0.85화-0.64。목전성촉진조직정량(VTQ)기술4C1화9L4탐두측량간실질탄성일치성교차。
Shear wave velocity (SWV) was performed in 31 patients with chronic hepatitis B (CHBs) and 45 healthy volunteers (control group). The occurrence rate of absence data with 4C1 was lower significantly than with 9L4 transducer both in CHBs and control group. The occurrence rate of absence data in the CHBs was not different from that in healthy controls with 4C1 transducer, which also could be seen when using 9L4 transducer. The SWV were significantly higher with 9L4 than 4C1 transducer both in control group and in CHBs. For SWV measurements performed by 4C1 and 9L4 transducer, the intraclass correlation coefficient (ICC) in the CHBs and control group were classified as poor and fair agreement. Bland-Altman analysis compared the differences against mean of SWV measured by 4C1 and 9L4 transducer for the CHBs and control group were at -0.85 and -0.64, respectively. VTQ in its present stage of development has low agreement between 4C1 and 9L4 transducer in the assessment of liver stiffness.