磁共振成像%回波平面成像%牙冠(假体)
磁共振成像%迴波平麵成像%牙冠(假體)
자공진성상%회파평면성상%아관(가체)
Magnetic resonance imaging%Echo-planar imaging%Crowns
目的:探讨在有金属冠存在的情况下,1.5 T磁共振成像T2*/梯度回波序列( GRE)、T2W/反转恢复自旋回波(IR)、T2W/快速自旋回波序列(FSE)、T1W/自旋回波序列(SE)4种序列产生伪影的程度。方法给实验犬依次佩戴钴铬合金、镍铬合金、低钛合金、纯钛金属4种金属冠,按序进行4种序列常规1.5 T磁共振成像的颅脑扫描,分别测量4种序列在4种金属冠存在下所产生伪影的最大面积并计数金属伪影涉及图像的层数,对测量结果进行统计分析。结果各金属冠在1.5 T磁共振成像4种序列均产生了伪影。4种序列钴铬合金、镍铬合金、低钛合金、纯钛金属伪影最大面积比较,差异有统计学意义( P<0.05);其中T2W/IR组、T2W/FSE组、T1W/SE组较T2*/GRE组均降低(P<0.05)。结论金属冠引起T2*/GRE序列伪影最大,其他3种序列产生伪影程度相近。
目的:探討在有金屬冠存在的情況下,1.5 T磁共振成像T2*/梯度迴波序列( GRE)、T2W/反轉恢複自鏇迴波(IR)、T2W/快速自鏇迴波序列(FSE)、T1W/自鏇迴波序列(SE)4種序列產生偽影的程度。方法給實驗犬依次珮戴鈷鉻閤金、鎳鉻閤金、低鈦閤金、純鈦金屬4種金屬冠,按序進行4種序列常規1.5 T磁共振成像的顱腦掃描,分彆測量4種序列在4種金屬冠存在下所產生偽影的最大麵積併計數金屬偽影涉及圖像的層數,對測量結果進行統計分析。結果各金屬冠在1.5 T磁共振成像4種序列均產生瞭偽影。4種序列鈷鉻閤金、鎳鉻閤金、低鈦閤金、純鈦金屬偽影最大麵積比較,差異有統計學意義( P<0.05);其中T2W/IR組、T2W/FSE組、T1W/SE組較T2*/GRE組均降低(P<0.05)。結論金屬冠引起T2*/GRE序列偽影最大,其他3種序列產生偽影程度相近。
목적:탐토재유금속관존재적정황하,1.5 T자공진성상T2*/제도회파서렬( GRE)、T2W/반전회복자선회파(IR)、T2W/쾌속자선회파서렬(FSE)、T1W/자선회파서렬(SE)4충서렬산생위영적정도。방법급실험견의차패대고락합금、얼락합금、저태합금、순태금속4충금속관,안서진행4충서렬상규1.5 T자공진성상적로뇌소묘,분별측량4충서렬재4충금속관존재하소산생위영적최대면적병계수금속위영섭급도상적층수,대측량결과진행통계분석。결과각금속관재1.5 T자공진성상4충서렬균산생료위영。4충서렬고락합금、얼락합금、저태합금、순태금속위영최대면적비교,차이유통계학의의( P<0.05);기중T2W/IR조、T2W/FSE조、T1W/SE조교T2*/GRE조균강저(P<0.05)。결론금속관인기T2*/GRE서렬위영최대,기타3충서렬산생위영정도상근。
Objective To observe and evaluate the metallic artifacts caused by dental metal crowns with four se-quences of magnetic resonance imaging in T2*/GRE,T2W/IR,T2W/FSE and T1W/SE. Methods Crowns made from four different materials( Co-Gr,Ni-Gr,Ti alloy and Pure-Ti)were included in this study. The experimental animal was a ripe crossbreed dog and the crowns were fabricated for its upper right second premolar. Each mental crown was examined by a head MRI(1. 5 T)scan with four sequences of T2*/GRE,T2W/IR,T2W/FSE and T1W/SE,respectively. The existence and ex-tent of the artifacts were assessed and compared. Results All metal crowns produced artifacts. The sequence T2*/GRE exhibited significantly wider artifact than the others(P<0. 05). For any of the test alloy crown,artifacts were seen on more MRI slices in the sequence T2*/GRE than in other three sequences which produced similar artifact presentation(P<0. 05). Conclusion A-nalysis of the artifact size and artifact-related imaging slices demonstrated that among all the four sequences,T2*/GRE has the strongest influence while the other three sequences contributed equally to artifact generation.