微剂量输液器%气道湿化%输液报警装置%持续恒温
微劑量輸液器%氣道濕化%輸液報警裝置%持續恆溫
미제량수액기%기도습화%수액보경장치%지속항온
Microdose apparatus infusions%Airway humidification%Transfusion alarm device%Constant temperature
目的:探讨微剂量输液器在呼吸机湿化器注水过程中的应用,以期加强机械通气患者气道均匀温、湿化,提高呼吸机湿化器的安全管理。方法采用半随机抽取信封法将83例使用呼吸机机械通气的患者分为两组,试验组采用微剂量输液器及输液报警装置进行湿化器持续注水,对照组采用传统的普通输液器间断注入湿化水,观察两组的注水时间、注水后患者吸入端温度差异、水温回升时间及痰液黏稠度等情况。结果试验组注水时间为(2.40±0.40) min/d,明显低于对照组的(13.24±11.03)min/d,两组比较差异有统计学意义(t =6.3658;P<0.01),试验组注水后吸入气体温度为(36.79±1.56)℃,明显高于对照组的(24.29±1.60)℃,两组比较差异有统计学意义(t=36.0382;P<0.01),机械通气第3天痰液黏稠度明显减轻。结论采用微剂量输液器及输液报警装置进行湿化器持续注水提高了湿化水量控制的准确性,使患者气道得到持续恒温湿化,弥补了湿化器的报警缺陷,保证患者的湿化安全,减少了护士的工作量。
目的:探討微劑量輸液器在呼吸機濕化器註水過程中的應用,以期加彊機械通氣患者氣道均勻溫、濕化,提高呼吸機濕化器的安全管理。方法採用半隨機抽取信封法將83例使用呼吸機機械通氣的患者分為兩組,試驗組採用微劑量輸液器及輸液報警裝置進行濕化器持續註水,對照組採用傳統的普通輸液器間斷註入濕化水,觀察兩組的註水時間、註水後患者吸入耑溫度差異、水溫迴升時間及痰液黏稠度等情況。結果試驗組註水時間為(2.40±0.40) min/d,明顯低于對照組的(13.24±11.03)min/d,兩組比較差異有統計學意義(t =6.3658;P<0.01),試驗組註水後吸入氣體溫度為(36.79±1.56)℃,明顯高于對照組的(24.29±1.60)℃,兩組比較差異有統計學意義(t=36.0382;P<0.01),機械通氣第3天痰液黏稠度明顯減輕。結論採用微劑量輸液器及輸液報警裝置進行濕化器持續註水提高瞭濕化水量控製的準確性,使患者氣道得到持續恆溫濕化,瀰補瞭濕化器的報警缺陷,保證患者的濕化安全,減少瞭護士的工作量。
목적:탐토미제량수액기재호흡궤습화기주수과정중적응용,이기가강궤계통기환자기도균균온、습화,제고호흡궤습화기적안전관리。방법채용반수궤추취신봉법장83례사용호흡궤궤계통기적환자분위량조,시험조채용미제량수액기급수액보경장치진행습화기지속주수,대조조채용전통적보통수액기간단주입습화수,관찰량조적주수시간、주수후환자흡입단온도차이、수온회승시간급담액점주도등정황。결과시험조주수시간위(2.40±0.40) min/d,명현저우대조조적(13.24±11.03)min/d,량조비교차이유통계학의의(t =6.3658;P<0.01),시험조주수후흡입기체온도위(36.79±1.56)℃,명현고우대조조적(24.29±1.60)℃,량조비교차이유통계학의의(t=36.0382;P<0.01),궤계통기제3천담액점주도명현감경。결론채용미제량수액기급수액보경장치진행습화기지속주수제고료습화수량공제적준학성,사환자기도득도지속항온습화,미보료습화기적보경결함,보증환자적습화안전,감소료호사적공작량。
Objective To explore the application of microdose apparatus infusionis in the process of water injection in the ventilator humidification device in order to strengthen the uniform temperature and moisture in patients with mechanical ventilation and improve the safety management of ventilator humidification device . Methods Eighty-three patients with mechanical ventilation were divided into the experimental group and the control group according to the semi random envelope .The experimental group received the constant water injection through the microdose apparatus infusions and transfusion alarm device , and the control group received the intermittent water injection through the normal infusion .The time of water injection , the temperature of end inhalation in patients after water injection , the elevation time of water temperature and sticky sputum were compared between the two groups.Results The time of water injection was (2.40 ±0.40) min/d in the experimental group, and was significantly lower than (13.24 ±11.03) min/d in the control group, and the difference was statistically significant (t =6.3658, P <0.01).The temperature of inspired gas after water injection was (36.79 ±1.56)℃ in the experimental group, and was significantly higher than (24.29 ± 1.60) ℃in the control group, and the difference was statistically significant (t=36.038 2, P<0.01).The sticky sputum was significantly reduced the third day after mechanical ventilation .Conclusions The microdose apparatus infusions and transfusion alarm device can keep on water injection in the humidification device , improve the accuracy of the humidifying water control , make the airway constant temperature humidification in patients, make up the alarm defect of humidification device , ensure the safety of humidification in patients , and reduce the workload of nurses .