国际检验医学杂志
國際檢驗醫學雜誌
국제검험의학잡지
INTERNATIONAL JOURNAL OF LABORATORY MEDICINE
2014年
11期
1478-1479
,共2页
测量范围%偏倚可接受距离%偏倚%方法学比对
測量範圍%偏倚可接受距離%偏倚%方法學比對
측량범위%편의가접수거리%편의%방법학비대
measurement range%distance from deviation to tolerable error%bias%method comparison
目的:改进 EP9-A2指南使之可用于全线性范围内多个定量检测系统间的偏倚评估。方法在4套检测系统(A、B、C、D)上检测40份患者标本的血清总胆固醇浓度,以 A 系统为参考系统,按 EP9-A2指南评估 B、C、D 系统的偏倚;在此基础上,计算两两系统间偏倚可接受距离 DD(偏倚与分析质量目标之间的距离),以置信区间包含0为标准来识别系统间不可接受的偏倚。结果与 A 系统相比,B、C 两系统呈负向偏倚,D 系统呈正向偏倚,但 B、C、D 与 A 系统的偏倚均可接受;各系统间 DD 呈正态分布;除DD BD 置信区间包含0,B、D 两系统间偏倚不可接受外,其余系统间偏倚均可被接受;将 D 系统结果经 A 系统校正后,各系统间偏倚均可接受;作图可提供全线性范围内的偏倚评估。结论DD 均值的置信区间,可在全线性范围内有效评估3个或更多系统间的偏倚,可用于评估系统间的结果可比性。
目的:改進 EP9-A2指南使之可用于全線性範圍內多箇定量檢測繫統間的偏倚評估。方法在4套檢測繫統(A、B、C、D)上檢測40份患者標本的血清總膽固醇濃度,以 A 繫統為參攷繫統,按 EP9-A2指南評估 B、C、D 繫統的偏倚;在此基礎上,計算兩兩繫統間偏倚可接受距離 DD(偏倚與分析質量目標之間的距離),以置信區間包含0為標準來識彆繫統間不可接受的偏倚。結果與 A 繫統相比,B、C 兩繫統呈負嚮偏倚,D 繫統呈正嚮偏倚,但 B、C、D 與 A 繫統的偏倚均可接受;各繫統間 DD 呈正態分佈;除DD BD 置信區間包含0,B、D 兩繫統間偏倚不可接受外,其餘繫統間偏倚均可被接受;將 D 繫統結果經 A 繫統校正後,各繫統間偏倚均可接受;作圖可提供全線性範圍內的偏倚評估。結論DD 均值的置信區間,可在全線性範圍內有效評估3箇或更多繫統間的偏倚,可用于評估繫統間的結果可比性。
목적:개진 EP9-A2지남사지가용우전선성범위내다개정량검측계통간적편의평고。방법재4투검측계통(A、B、C、D)상검측40빈환자표본적혈청총담고순농도,이 A 계통위삼고계통,안 EP9-A2지남평고 B、C、D 계통적편의;재차기출상,계산량량계통간편의가접수거리 DD(편의여분석질량목표지간적거리),이치신구간포함0위표준래식별계통간불가접수적편의。결과여 A 계통상비,B、C 량계통정부향편의,D 계통정정향편의,단 B、C、D 여 A 계통적편의균가접수;각계통간 DD 정정태분포;제DD BD 치신구간포함0,B、D 량계통간편의불가접수외,기여계통간편의균가피접수;장 D 계통결과경 A 계통교정후,각계통간편의균가접수;작도가제공전선성범위내적편의평고。결론DD 균치적치신구간,가재전선성범위내유효평고3개혹경다계통간적편의,가용우평고계통간적결과가비성。
Objective To improve EP9-A2 for Bias estimation among multiple quantitative detection systems within full range of AMR.Methods 40 patients specimens were determined twice for serum total cholesterol by four detection systems(A,B,C and D).With system A served as comparative method,Bias between A and the other three was evaluated according to CLSI EP9-A2 separately.Furthermore,DD(distance from deviation to tolerable error)and its average confidence intervals between every two sys-tem were calculated and compared with zero.The confidence interval of greater than zero was served as criteria for accepting bias between every two system.Results Bias between A and the other three meet the analytical quality specification according to EP9-A2,although that of D system was positive,and those of B and C system were negative.DD between every two system obeyed nor-mality distribution.All biases between every two system wes acceptable except that between B and D,causing of their interval con-taining zero.After correcting of results from system D,Biases between every two system were all acceptable.Plots of confidence in-terval could provide a full range bias assessment within AMR.Conclusion Comparability and Bias estimation in full range of AMR for results between every two system among 3 or more systems could be evaluated by confidence intervals.