分析人工读数对校准结果的波动性、均匀性影响。针对人工读数存在缺陷,提出适用于恒温槽校准自动采集软件方案,并对PDT-2A高精密数字温度计进行基于Visual Basic自动采集软件实现,对软件功能及准确性实验验证。
分析人工讀數對校準結果的波動性、均勻性影響。針對人工讀數存在缺陷,提齣適用于恆溫槽校準自動採集軟件方案,併對PDT-2A高精密數字溫度計進行基于Visual Basic自動採集軟件實現,對軟件功能及準確性實驗驗證。
분석인공독수대교준결과적파동성、균균성영향。침대인공독수존재결함,제출괄용우항온조교준자동채집연건방안,병대PDT-2A고정밀수자온도계진행기우Visual Basic자동채집연건실현,대연건공능급준학성실험험증。
An introduction to the current application of thermostatic bath calibration in measurement department is presented in the paper, and the impact of artificial readings on the fluctuation and uniformity of the correction result is analyzed. With a view of the inconvenience and defect of artificial readings, An automatic acquisition software designed scenario applicable to thermostatic bath calibration is put forward. An automatic acquisition software realization based on Visual Basic is designed for PDT-2A high precise digital thermometer, and its function and accuracy is further experimented.