激光与红外
激光與紅外
격광여홍외
LASER & INFRARED
2014年
5期
528-532
,共5页
王亚红%周萍%董大明%郑文刚%叶松%王文重
王亞紅%週萍%董大明%鄭文剛%葉鬆%王文重
왕아홍%주평%동대명%정문강%협송%왕문중
近红外光谱%透反射%蔗糖%偏最小二乘%主成分回归
近紅外光譜%透反射%蔗糖%偏最小二乘%主成分迴歸
근홍외광보%투반사%자당%편최소이승%주성분회귀
near infrared spectroscopy%transmission and reflection%sucrose%partial least squares%principal component regression
利用近红外透反射光谱技术,研究短波近红外光谱(780~1100nm)无损检测蔗糖溶液的可行性,并通过主成分回归(PCR)和偏最小二乘(PLS )方法建立了蔗糖溶液的近红外定量分析模型。采用Savitzky-Golay卷积平滑(5点)和多元散射校正(MSC)进行预处理,并且对预处理后的数据进行建模分析。PCR定量分析的结果:主成分数 PC =7,交互验证相关系数RCV =0.957335,交互验证的校正标准偏差RMSECV=0.015859;PLS定量分析结果:主成分数PC=4,交互验证相关系数RCV =0.975789,交互验证的校正标准偏差RMSECV=0.012251。分别用PCR和PLS的校正模型对预测集样本进行预测,两种模型的预测标准偏差RMSEP分别为0.0127,0.0118。二者均对高浓度蔗糖溶液的预测结果比较理想,而且在PLS模型下,77%以上的样本相对误差在10%以下,较PCR模型的高。综合结果,PLS所建立的模型简单,而且精度很高,所以,基于短波近红外光谱的蔗糖浓度的快速无损检测是可行的。
利用近紅外透反射光譜技術,研究短波近紅外光譜(780~1100nm)無損檢測蔗糖溶液的可行性,併通過主成分迴歸(PCR)和偏最小二乘(PLS )方法建立瞭蔗糖溶液的近紅外定量分析模型。採用Savitzky-Golay捲積平滑(5點)和多元散射校正(MSC)進行預處理,併且對預處理後的數據進行建模分析。PCR定量分析的結果:主成分數 PC =7,交互驗證相關繫數RCV =0.957335,交互驗證的校正標準偏差RMSECV=0.015859;PLS定量分析結果:主成分數PC=4,交互驗證相關繫數RCV =0.975789,交互驗證的校正標準偏差RMSECV=0.012251。分彆用PCR和PLS的校正模型對預測集樣本進行預測,兩種模型的預測標準偏差RMSEP分彆為0.0127,0.0118。二者均對高濃度蔗糖溶液的預測結果比較理想,而且在PLS模型下,77%以上的樣本相對誤差在10%以下,較PCR模型的高。綜閤結果,PLS所建立的模型簡單,而且精度很高,所以,基于短波近紅外光譜的蔗糖濃度的快速無損檢測是可行的。
이용근홍외투반사광보기술,연구단파근홍외광보(780~1100nm)무손검측자당용액적가행성,병통과주성분회귀(PCR)화편최소이승(PLS )방법건립료자당용액적근홍외정량분석모형。채용Savitzky-Golay권적평활(5점)화다원산사교정(MSC)진행예처리,병차대예처리후적수거진행건모분석。PCR정량분석적결과:주성분수 PC =7,교호험증상관계수RCV =0.957335,교호험증적교정표준편차RMSECV=0.015859;PLS정량분석결과:주성분수PC=4,교호험증상관계수RCV =0.975789,교호험증적교정표준편차RMSECV=0.012251。분별용PCR화PLS적교정모형대예측집양본진행예측,량충모형적예측표준편차RMSEP분별위0.0127,0.0118。이자균대고농도자당용액적예측결과비교이상,이차재PLS모형하,77%이상적양본상대오차재10%이하,교PCR모형적고。종합결과,PLS소건립적모형간단,이차정도흔고,소이,기우단파근홍외광보적자당농도적쾌속무손검측시가행적。
Near-infrared spectra of trans-reflective is applied to determine the feasibility of nondestructive testing of the sucrose solution in the near-infrared wavelength range of 780~1 1 00nm,and principal component regression (PCR) and partial least squares(PLS)are used to establish the near-infrared quantitative analysis model of sucrose solu-tion.After the pretreatment of (Savitzky-Golay)5 points smoothing and multiplicative scatter correction (MSC),the (PCR and PLS)calibration model is made.The results of quantitative analysis of PCR are the number of principal component PC=7,the correlation coefficient RCV =0.957335,root mean square error of cross validation RMSECV=0.01 5859;The results of quantitative analysis of PLS are the number of principal component PC=4,the correlation coefficient RCV =0.975789,root mean square error of cross-validation RMSECV=0.01 2251.The prediction samples are predicted respectively by the calibration model of PCR and PLS.Root mean square errors of prediction are 0.01 27,0.01 1 8 respectively.Both of the methods are satisfactory for the predicted results of high concentration sam-ples.The relative errors of more than 77% of validation samples are below 1 0% using PLS.All of results show that PLS model is simpler and its prediction accuracy is higher,and short-wave near-infrared spectrometry is a valuable, rapid and nondestructive tool for the quantitative analysis of sucrose solution.