陈韵%武哲明%陈海松%何曼莎
陳韻%武哲明%陳海鬆%何曼莎
진운%무철명%진해송%하만사
眼轴%非接触式光学相干生物测量仪%高度近视%人工晶状体
眼軸%非接觸式光學相榦生物測量儀%高度近視%人工晶狀體
안축%비접촉식광학상간생물측량의%고도근시%인공정상체
Axial length%IOL-Master%Cataract%Myopia
目的:比较非接触式光学相干生物测量仪(intraocularlen-master,IOL-Master)与传统超声生物测量方法在测量高度近视白内障眼人工晶状体度数的准确性,评价IOL-Master的特性及其临床应用价值。方法对48例62只高度近视眼并发白内障患者施行超声乳化白内障吸出术。术前分别用IOL-Master、传统超声生物测量仪测量眼轴长度,用IOL-Master测量角膜曲率,根据测量结果使用SRK/T公式计算人工晶状体度数,术中植入可折叠人工晶状体。术后1个月随诊检查视力及眼屈光度。结果IOL-Master和传统超声生物测量法,测得的眼轴长度分别为(29.32±1.48)mm和(28.78±1.54)mm,二者对比差异有显著性(P<0.05);术后1个月的平均绝对屈光误差两组对比差异有显著性(P<0.05)。结论IOL-Master是一种高精确性、安全可靠、操作简单的人工晶状体度数测量工具。尤其对高度近视白内障手术人工晶状体的选择有重要临床价值。
目的:比較非接觸式光學相榦生物測量儀(intraocularlen-master,IOL-Master)與傳統超聲生物測量方法在測量高度近視白內障眼人工晶狀體度數的準確性,評價IOL-Master的特性及其臨床應用價值。方法對48例62隻高度近視眼併髮白內障患者施行超聲乳化白內障吸齣術。術前分彆用IOL-Master、傳統超聲生物測量儀測量眼軸長度,用IOL-Master測量角膜麯率,根據測量結果使用SRK/T公式計算人工晶狀體度數,術中植入可摺疊人工晶狀體。術後1箇月隨診檢查視力及眼屈光度。結果IOL-Master和傳統超聲生物測量法,測得的眼軸長度分彆為(29.32±1.48)mm和(28.78±1.54)mm,二者對比差異有顯著性(P<0.05);術後1箇月的平均絕對屈光誤差兩組對比差異有顯著性(P<0.05)。結論IOL-Master是一種高精確性、安全可靠、操作簡單的人工晶狀體度數測量工具。尤其對高度近視白內障手術人工晶狀體的選擇有重要臨床價值。
목적:비교비접촉식광학상간생물측량의(intraocularlen-master,IOL-Master)여전통초성생물측량방법재측량고도근시백내장안인공정상체도수적준학성,평개IOL-Master적특성급기림상응용개치。방법대48례62지고도근시안병발백내장환자시행초성유화백내장흡출술。술전분별용IOL-Master、전통초성생물측량의측량안축장도,용IOL-Master측량각막곡솔,근거측량결과사용SRK/T공식계산인공정상체도수,술중식입가절첩인공정상체。술후1개월수진검사시력급안굴광도。결과IOL-Master화전통초성생물측량법,측득적안축장도분별위(29.32±1.48)mm화(28.78±1.54)mm,이자대비차이유현저성(P<0.05);술후1개월적평균절대굴광오차량조대비차이유현저성(P<0.05)。결론IOL-Master시일충고정학성、안전가고、조작간단적인공정상체도수측량공구。우기대고도근시백내장수술인공정상체적선택유중요림상개치。
Objective To compare the measurement of axial length(AL)by IOL-Master and contact ultrasonic(US) axial scan(A-scan)in high myopia with cataract. Methods 48 patients(62 eyes)of high myopia with cataract were comprised in this study. Before the phacoemulsiifcation,the axial length and corneal curvature were measured with IOL-Master and contact A-scan and respectively. Keratometric power was measured by IOL-Master.IOL power calculation was carried out according to SRK/T formula in the basis of IOL-Master and contact A-scan and respectively. Visual acuity and ocular diopter were examined during the followed up time for one month. Results The axial length was(28.78±1.54)mm by contact A-scan,(29.32±1.48)mm IOL-Master respectively. There were signiifcant difference between contact A-scan and IOL-Master(P<0.05).Mean absolute refractive error (MAFE) after one month was (0.431±0. 168)D(0~1.00)D by IOL-Master, and(0.682±0.27)D(0~1.15)D by contact A-scan. There was a signiifcant difference between contact A-scan and IOL-Master(P<0.05). Conclusions The IOL-Master is a non-contact, accurate, safe tool for calculating IOL power, and it is more accurate than contact A-scan,especially for high myopia.