目的:建立简便、快速地测定人全血中西罗莫司血药浓度的HPLC-MS/MS方法。方法:全血样品经硫酸锌及含有子囊霉素(内标)的乙腈溶液沉淀后,取上清液20μL进样分析。采用Symmetry-C18色谱柱(2.1 mm×50 mm,3.5μm),流动相梯度洗脱,流速为0.5 mL·min-1,质谱采用电喷雾离子化电离源,经多反应监测模式检测。结果:西罗莫司和子囊霉素的保留时间分别为2.24 min和2.21 min,空白血浆中内源性物质不干扰测定;西罗莫司的血药浓度在1~30 ng· mL-1时,线性关系良好(r=0.9995);日内、日间RSD值均小于10%,提取回收率为86.4%~90.4%。结论:本方法灵敏、准确、简单、快速,可用于西罗莫司的血药浓度监测和早期肝移植患者药代动力学研究。
目的:建立簡便、快速地測定人全血中西囉莫司血藥濃度的HPLC-MS/MS方法。方法:全血樣品經硫痠鋅及含有子囊黴素(內標)的乙腈溶液沉澱後,取上清液20μL進樣分析。採用Symmetry-C18色譜柱(2.1 mm×50 mm,3.5μm),流動相梯度洗脫,流速為0.5 mL·min-1,質譜採用電噴霧離子化電離源,經多反應鑑測模式檢測。結果:西囉莫司和子囊黴素的保留時間分彆為2.24 min和2.21 min,空白血漿中內源性物質不榦擾測定;西囉莫司的血藥濃度在1~30 ng· mL-1時,線性關繫良好(r=0.9995);日內、日間RSD值均小于10%,提取迴收率為86.4%~90.4%。結論:本方法靈敏、準確、簡單、快速,可用于西囉莫司的血藥濃度鑑測和早期肝移植患者藥代動力學研究。
목적:건립간편、쾌속지측정인전혈중서라막사혈약농도적HPLC-MS/MS방법。방법:전혈양품경류산자급함유자낭매소(내표)적을정용액침정후,취상청액20μL진양분석。채용Symmetry-C18색보주(2.1 mm×50 mm,3.5μm),류동상제도세탈,류속위0.5 mL·min-1,질보채용전분무리자화전리원,경다반응감측모식검측。결과:서라막사화자낭매소적보류시간분별위2.24 min화2.21 min,공백혈장중내원성물질불간우측정;서라막사적혈약농도재1~30 ng· mL-1시,선성관계량호(r=0.9995);일내、일간RSD치균소우10%,제취회수솔위86.4%~90.4%。결론:본방법령민、준학、간단、쾌속,가용우서라막사적혈약농도감측화조기간이식환자약대동역학연구。
Objective:To establish a simple and rapid HPLC-MS/MS method for determination of sirolimus in human whole blood. Methods:The samples of whole blood were deproteinated by zinc sulfate and internal standard reconstituted in acetonitrile and separated by a Symmetry-C18 column (2.1 mm×50 mm, 3.5 μm) at the flow rate of 0.5 mL·min-1 with gradient elution. Detection was performed on a tandem mass spectrometry with electrospray ionization source in positive ionmode. Multiple reactions monitoring mode was used to quantify sirolimus and internal standard respectively. Results: The retention time of sirolimus and internal standard were 2.24 min and 2.21 min, respectively. The linear range of sirolimus was 1–30 ng·mL-1( r=0.999 5). The intra-day and inter-day RSD were both less than 10%. The extraction recoveries were 86.4%–90.4%. Conclusion:The method is sensitive, accurate, simple and rapid for monitoring of whole blood concentration of sirolimus and pharmacokinetic study in early liver tansplant recipients.