化学研究与应用
化學研究與應用
화학연구여응용
CHEMICAL RESEARCH AND APPLICATION
2014年
6期
933-937
,共5页
肖新峰%渠美云%邵恬恬%徐辉
肖新峰%渠美雲%邵恬恬%徐輝
초신봉%거미운%소념념%서휘
流动注射光度法%化学需氧量COD%高锰酸盐指数CODMn%KMnO4%海水
流動註射光度法%化學需氧量COD%高錳痠鹽指數CODMn%KMnO4%海水
류동주사광도법%화학수양량COD%고맹산염지수CODMn%KMnO4%해수
flow injection spectrophotometry%chemical oxygen demand( COD)%permanganate index CODMn%KMnO4%seawater
采用流动注射光度法对海水化学需氧量( CODMn ,高锰酸盐指数)进行快速测定。实验以KMnO4作氧化剂,淀粉-KI作显色剂,葡萄糖作标准物质;对流动注射分析流路进行了设计,优化了反应圈长度、反应温度、反应物浓度、进样量、泵速等实验条件;并通过自动加热装置对反应器进行在线加热;在最优条件下,测定方法的线性范围为0~6mg·L-1(实际需氧量),最低检出限为0.13mg·L-1,实际海水测定时加标回收率在96%~104%范围内,对1mg·L-1的COD标准溶液进行了11次连续测定,相对标准偏差( RSD)为1.88%。该方法操作简单、方便、灵敏度高,用于海水实际样品分析时,其测定值与国标法测定值一致。
採用流動註射光度法對海水化學需氧量( CODMn ,高錳痠鹽指數)進行快速測定。實驗以KMnO4作氧化劑,澱粉-KI作顯色劑,葡萄糖作標準物質;對流動註射分析流路進行瞭設計,優化瞭反應圈長度、反應溫度、反應物濃度、進樣量、泵速等實驗條件;併通過自動加熱裝置對反應器進行在線加熱;在最優條件下,測定方法的線性範圍為0~6mg·L-1(實際需氧量),最低檢齣限為0.13mg·L-1,實際海水測定時加標迴收率在96%~104%範圍內,對1mg·L-1的COD標準溶液進行瞭11次連續測定,相對標準偏差( RSD)為1.88%。該方法操作簡單、方便、靈敏度高,用于海水實際樣品分析時,其測定值與國標法測定值一緻。
채용류동주사광도법대해수화학수양량( CODMn ,고맹산염지수)진행쾌속측정。실험이KMnO4작양화제,정분-KI작현색제,포도당작표준물질;대류동주사분석류로진행료설계,우화료반응권장도、반응온도、반응물농도、진양량、빙속등실험조건;병통과자동가열장치대반응기진행재선가열;재최우조건하,측정방법적선성범위위0~6mg·L-1(실제수양량),최저검출한위0.13mg·L-1,실제해수측정시가표회수솔재96%~104%범위내,대1mg·L-1적COD표준용액진행료11차련속측정,상대표준편차( RSD)위1.88%。해방법조작간단、방편、령민도고,용우해수실제양품분석시,기측정치여국표법측정치일치。
The method of flow injection spectrophotometry is described for rapidly determining chemical oxygen demand(CODMn, permanganate index) in seawater. The method uses potassium permanganate as oxidant reagent,starch-KI as chromogenic agent and glucose as standard. The flow injection analysis flow is designed and is optimized by optimizing the experimental conditions which include the length of the reaction coil,reaction temperature,reactant concentration,sample size,pump speed,etc. And the reactor is on-line heating through an automatic heating device. Under the selected conditions,the calibration graph is linear within 0-6mg·L-1 (the actual oxygen demand)and the detection limit is 0. 13 mg·L-1. The recovery is between 96%and 104%when it is applied to the determination of the actual seawater and the relative standard deviation of the method is 1. 88%RSD(n=11)for the sample with 1mg·L-1 of COD. The method is simple and convenient operation,and high sensitivity,etc. and the measurement value of analyzer is the same as the result of standard method when used in seawater analysis of actual samples.