幼稚粒细胞%Sysmex XE - 5000 血液分析仪%显微镜检查
幼稚粒細胞%Sysmex XE - 5000 血液分析儀%顯微鏡檢查
유치립세포%Sysmex XE - 5000 혈액분석의%현미경검사
immature granulocytes%blood Sysmex XE - 5000 analyzer%microscopy
目的:Sysmex XE -5000血液分析仪对外周血幼稚粒细胞(IG)检出与显微镜分类计数结果对比,进行评价分析及正常临界值的设定。方法:Sysmex XE -5000血液分析仪对438份血标本采用白细胞分类检测通道(CBC +DIFF 通道)检测幼稚粒细胞,并用显微镜人工计数 IG。结果:Sysmex XE -5000血液分析仪和显微镜镜检法对 IG 检出呈正相关( r =0.76),SysmexXE -5000血液分析仪对IG%检测的敏感性为62.9%,特异性为83.4%,临界值﹥0.6。正常组的 IG 检测人工镜检均未见幼稚粒细胞,仪器检测显示 IG%范围为0~0.6(x ±3s)。结论:当 Sysmex XE-5000血液分析仪检测结果 IG%≤0.6[或绝对值(IG#)≤0.1×109/ L]可认为是正常标本,当 IG%﹥0.6[或绝对值(IG#)﹥0.1×109/ L)则为异常标本,应进行血涂片复查以明确诊断。
目的:Sysmex XE -5000血液分析儀對外週血幼稚粒細胞(IG)檢齣與顯微鏡分類計數結果對比,進行評價分析及正常臨界值的設定。方法:Sysmex XE -5000血液分析儀對438份血標本採用白細胞分類檢測通道(CBC +DIFF 通道)檢測幼稚粒細胞,併用顯微鏡人工計數 IG。結果:Sysmex XE -5000血液分析儀和顯微鏡鏡檢法對 IG 檢齣呈正相關( r =0.76),SysmexXE -5000血液分析儀對IG%檢測的敏感性為62.9%,特異性為83.4%,臨界值﹥0.6。正常組的 IG 檢測人工鏡檢均未見幼稚粒細胞,儀器檢測顯示 IG%範圍為0~0.6(x ±3s)。結論:噹 Sysmex XE-5000血液分析儀檢測結果 IG%≤0.6[或絕對值(IG#)≤0.1×109/ L]可認為是正常標本,噹 IG%﹥0.6[或絕對值(IG#)﹥0.1×109/ L)則為異常標本,應進行血塗片複查以明確診斷。
목적:Sysmex XE -5000혈액분석의대외주혈유치립세포(IG)검출여현미경분류계수결과대비,진행평개분석급정상림계치적설정。방법:Sysmex XE -5000혈액분석의대438빈혈표본채용백세포분류검측통도(CBC +DIFF 통도)검측유치립세포,병용현미경인공계수 IG。결과:Sysmex XE -5000혈액분석의화현미경경검법대 IG 검출정정상관( r =0.76),SysmexXE -5000혈액분석의대IG%검측적민감성위62.9%,특이성위83.4%,림계치﹥0.6。정상조적 IG 검측인공경검균미견유치립세포,의기검측현시 IG%범위위0~0.6(x ±3s)。결론:당 Sysmex XE-5000혈액분석의검측결과 IG%≤0.6[혹절대치(IG#)≤0.1×109/ L]가인위시정상표본,당 IG%﹥0.6[혹절대치(IG#)﹥0.1×109/ L)칙위이상표본,응진행혈도편복사이명학진단。
Objective:To compare the results of immature granulocyte by sysmex XE - 5000 hematology analyzer and mi-croscopic count,to evaluation analysis and set critical value. Methods:All 438 blood samples using the white blood cell detec-tion channels(CBC + DIFF channel)to detect with sysmex XE - 5000 hematology analyzer,to count IG manual using micro-scope. Results:The corelation of sysmex XE - 5000 hematology analyzer and microscopic examination method for the detection of IG correlation(r = 0. 76),the detection sensitivity of IG% in sysmex XE - 5000 hematology analyzer was 62. 9% ,the spe-cificity was 83. 4% ,critical value ﹥ 0. 6. The detectionof IG in normal group showed no immature granulocyte,the range of IG% by instrument was 0 - 0. 6(x ± 3 s). Conclusion:When test results of IG% by sysmex XE - 5000 hematology analyzer was≤0 . 6( or IG# ≤0 . 1 × 109 / L)canbe considerednormal specimens,whenIG% ﹥ 0 . 6[ or absolute value( IG# )﹥ 0 . 1 × 109 / L]compared with abnormal specimens,blood smears should be reviewed to confirm the diagnosis.