Chinese Journal of Laboratory Pathologist
2014年
2期
109-112
,共4页
动脉血%静脉血%HGB含量%全血细胞分析仪%血气分析仪
動脈血%靜脈血%HGB含量%全血細胞分析儀%血氣分析儀
동맥혈%정맥혈%HGB함량%전혈세포분석의%혈기분석의
Arterial blood%Venous blood%Hemoglobin level%Hematology analyzer%Blood-gas ana-lyzer
目的:分析动脉血气分析与静脉全血细胞分析测定HGB(hemoglobin, HGB)浓度的相关性。方法随机抽取77例动脉血标本与84例静脉血标本分别于ABL835血气分析仪和Sysmex XE-5000全血细胞分析仪检测并比对其测量结果;随机抽取同时采集动、静脉血并对应进行动脉血气分析和全血细胞分析的394例患者,分析其检测HGB结果的相关性。结果 Sysmex XE-5000全血细胞分析仪精密度高于ABL835血气分析仪,而血气分析仪检测动、静脉血HGB结果均略高于全血细胞分析仪,且差异均有统计学意义(P均<0.05);动、静脉血两种方法检测结果均呈明显正相关(r=0.997,r=0999,P均<0.05)。同一时间动、静脉血HGB全血细胞分析测量结果高于血气分析结果,二者差异有统计学意义(P<0.05),且两种方法检测结果呈高度正相关(r=0.992,P<0.05)。结论准确的动脉血气分析HGB结果可暂时替代静脉全血细胞分析HGB结果,并推断动脉中HGB含量低于静脉血约2 g/L。
目的:分析動脈血氣分析與靜脈全血細胞分析測定HGB(hemoglobin, HGB)濃度的相關性。方法隨機抽取77例動脈血標本與84例靜脈血標本分彆于ABL835血氣分析儀和Sysmex XE-5000全血細胞分析儀檢測併比對其測量結果;隨機抽取同時採集動、靜脈血併對應進行動脈血氣分析和全血細胞分析的394例患者,分析其檢測HGB結果的相關性。結果 Sysmex XE-5000全血細胞分析儀精密度高于ABL835血氣分析儀,而血氣分析儀檢測動、靜脈血HGB結果均略高于全血細胞分析儀,且差異均有統計學意義(P均<0.05);動、靜脈血兩種方法檢測結果均呈明顯正相關(r=0.997,r=0999,P均<0.05)。同一時間動、靜脈血HGB全血細胞分析測量結果高于血氣分析結果,二者差異有統計學意義(P<0.05),且兩種方法檢測結果呈高度正相關(r=0.992,P<0.05)。結論準確的動脈血氣分析HGB結果可暫時替代靜脈全血細胞分析HGB結果,併推斷動脈中HGB含量低于靜脈血約2 g/L。
목적:분석동맥혈기분석여정맥전혈세포분석측정HGB(hemoglobin, HGB)농도적상관성。방법수궤추취77례동맥혈표본여84례정맥혈표본분별우ABL835혈기분석의화Sysmex XE-5000전혈세포분석의검측병비대기측량결과;수궤추취동시채집동、정맥혈병대응진행동맥혈기분석화전혈세포분석적394례환자,분석기검측HGB결과적상관성。결과 Sysmex XE-5000전혈세포분석의정밀도고우ABL835혈기분석의,이혈기분석의검측동、정맥혈HGB결과균략고우전혈세포분석의,차차이균유통계학의의(P균<0.05);동、정맥혈량충방법검측결과균정명현정상관(r=0.997,r=0999,P균<0.05)。동일시간동、정맥혈HGB전혈세포분석측량결과고우혈기분석결과,이자차이유통계학의의(P<0.05),차량충방법검측결과정고도정상관(r=0.992,P<0.05)。결론준학적동맥혈기분석HGB결과가잠시체대정맥전혈세포분석HGB결과,병추단동맥중HGB함량저우정맥혈약2 g/L。
Objective To analyze the correlation of hemoglobin (HGB) concentration detected by the arterial blood gas analysis and venous whole blood cell analysis. Methods 77 cases of arterial blood samples and 84 cases of venous blood samples were randomly selected and then detected respectively with ABL835 blood-gas analyzer and the Sysmex XE-5000 hematology analyzer, and the results were compared. The arterial and venous blood samples of 394 cases were drawn randomly at the same time. Then the HGB results correla-tion of these samples who undergone the blood-gas analysis and the whole blood cell analysis were analyzed. Results The precision of Sysmex XE-5000 hematology analyzer was higher than ABL835 blood-gas analyzer. The HGB results from blood-gas analyzer were all slightly higher than the hematology analyzer , and the differ-ences all had statistical significance(Pall<0.05). The arterial and venous blood HGB results of two methods all shown positive correlation (r=0.997, r=0.999, Pall<0.05). The HGB result of whole blood cell analysis was higher than blood-gas analysis at the same time, and the difference had statistical significance(P<0.05), The results of two methods shown positive correlation (r=0.992, P<0.05). Conclusion The precise results of HGB by arterial blood-gas analyzer can temporarily replace the results of hematology analyzer , and arterial blood HGB content is less about 2 g/L than venous blood according the extrapolation.