Sirius系统%Lenstar 900%眼前节参数%一致性
Sirius繫統%Lenstar 900%眼前節參數%一緻性
Sirius계통%Lenstar 900%안전절삼수%일치성
Sirius system%Lenstar 900%anterior segment parameters%agreement
目的:比较两种非接触眼生物测量仪Lenstar 900( LS900)和Sirius眼前节分析系统测量眼前节有关参数的异同。<br> 方法:对近视患者54例87眼分别采用LS900及Sirius系统测量角膜中央厚度( central corneal thickness, CCT)、前房深度(aqueous depth, ACD)、角膜平K(flat keratometry, FK)、角膜陡 K(steep keratometry,SK)、白到白(white to white, WTW)。采用配对 t 检验、Pearson 相关和 Bland-Altman图分析和评估。<br> 结果:Sirius系统检测的CCT和ACD大于LS900,而FK、SK和 WTW 小于 LS900,差值的均值分别为-6.11±6.32μm;-0.09±0.07 mm;0.18±0.25 D;0.21±0.36 D 和0.25±0.39mm,差异有统计学意义(P<0.01)。 Pearson相关性分析两种检测结果均显著相关( P<0.01)。 Bland-Altman 一致性比较:95%一致性界限分别为6.26~18.49μm,0.04~0.22mm,0.68~0.32D,0.92~0.50D和1.00~0.5mm。<br> 结论:LS900和Sirius系统对眼前节参数的检测具有良好的相关性和一致性,但也存在一定差异。这可能与Lenstar和Sirius测量方式并不一致有关,因此还有待进一步研究。
目的:比較兩種非接觸眼生物測量儀Lenstar 900( LS900)和Sirius眼前節分析繫統測量眼前節有關參數的異同。<br> 方法:對近視患者54例87眼分彆採用LS900及Sirius繫統測量角膜中央厚度( central corneal thickness, CCT)、前房深度(aqueous depth, ACD)、角膜平K(flat keratometry, FK)、角膜陡 K(steep keratometry,SK)、白到白(white to white, WTW)。採用配對 t 檢驗、Pearson 相關和 Bland-Altman圖分析和評估。<br> 結果:Sirius繫統檢測的CCT和ACD大于LS900,而FK、SK和 WTW 小于 LS900,差值的均值分彆為-6.11±6.32μm;-0.09±0.07 mm;0.18±0.25 D;0.21±0.36 D 和0.25±0.39mm,差異有統計學意義(P<0.01)。 Pearson相關性分析兩種檢測結果均顯著相關( P<0.01)。 Bland-Altman 一緻性比較:95%一緻性界限分彆為6.26~18.49μm,0.04~0.22mm,0.68~0.32D,0.92~0.50D和1.00~0.5mm。<br> 結論:LS900和Sirius繫統對眼前節參數的檢測具有良好的相關性和一緻性,但也存在一定差異。這可能與Lenstar和Sirius測量方式併不一緻有關,因此還有待進一步研究。
목적:비교량충비접촉안생물측량의Lenstar 900( LS900)화Sirius안전절분석계통측량안전절유관삼수적이동。<br> 방법:대근시환자54례87안분별채용LS900급Sirius계통측량각막중앙후도( central corneal thickness, CCT)、전방심도(aqueous depth, ACD)、각막평K(flat keratometry, FK)、각막두 K(steep keratometry,SK)、백도백(white to white, WTW)。채용배대 t 검험、Pearson 상관화 Bland-Altman도분석화평고。<br> 결과:Sirius계통검측적CCT화ACD대우LS900,이FK、SK화 WTW 소우 LS900,차치적균치분별위-6.11±6.32μm;-0.09±0.07 mm;0.18±0.25 D;0.21±0.36 D 화0.25±0.39mm,차이유통계학의의(P<0.01)。 Pearson상관성분석량충검측결과균현저상관( P<0.01)。 Bland-Altman 일치성비교:95%일치성계한분별위6.26~18.49μm,0.04~0.22mm,0.68~0.32D,0.92~0.50D화1.00~0.5mm。<br> 결론:LS900화Sirius계통대안전절삼수적검측구유량호적상관성화일치성,단야존재일정차이。저가능여Lenstar화Sirius측량방식병불일치유관,인차환유대진일보연구。
AlM: To compare the anterior segment measurements derived from optical low coherence reflectometer ( Lenstar LS900 ) and combined Scheimpflug - Placido disk topographer ( Sirius) . <br> METHODS: ln this study, we enrolled healthy myopic subjects 54 ( 87 eyes ) . The central corneal thickness (CCT), anterior chamber depth (ACD), flat keratometry ( FK) readings , steep keratometry ( SK ) readings and white to white ( WTW ) were measured by LS900 and Sirius. Evaluation and analysis were performed using paired t tests, the Pearson correlation, and Bland-Altman analyses. <br> RESULTS: The CCT and ACD measurements were significantly lower whereas FK, SK and WTW measurements were higher with LS900 (P<0. 001). Mean differences were -6. 11 ± 6. 32μm,-0. 09 ± 0. 07mm, 0. 18 ± 0. 25D, 0. 21 ± 0. 36D and 0. 25 ± 0. 39mm with statistical significant (P<0. 01). Pearson correlation analysis showed high correlation between the 2 devices for all measurements (P<0. 001). On Bland-Altman analysis, 95% limits of agreement for all measurements were 6. 26 to -18. 49μm; 0. 04 to 0. 22mm; 0. 68 to 0. 32D; 0. 92 to 0. 50D and 1. 00 to 0. 5mm. <br> CONCLUSlON: Anterior segment parameters evaluated with LS900 and Sirius systemare correlated well and achieve good agreement. However, there are significantly statistical differences which may be caused by the different measurement modes, so they may not be interchangeable use under certain clinical circumstances.