INTERNATIONAL JOURNAL OF LABORATORY MEDICINE
2014年
12期
1628-1628,1665
,共2页
吕艳关%范广来%魏红%张凤兰%陆业霞%张慧玲%武荣
呂豔關%範廣來%魏紅%張鳳蘭%陸業霞%張慧玲%武榮
려염관%범엄래%위홍%장봉란%륙업하%장혜령%무영
黄体酮%准确度%精密度%变异系数%全自动微粒子化学发光分析仪
黃體酮%準確度%精密度%變異繫數%全自動微粒子化學髮光分析儀
황체동%준학도%정밀도%변이계수%전자동미입자화학발광분석의
progesterone%accuracy%precision%coefficient of variation%automated chemiluminescent analyzer
目的:对UniCel DXI 800型全自动微粒子化学发光分析仪定量检测血清黄体酮(Prog)进行性能评价。方法运用该系统检测血清Prog,并对其灵敏度、精密度、准确度及特异性等进行方法学评价。同时与Access 2免疫分析系统进行相关性比较。结果该系统检测血清 Prog 的精密度较高,变异系数小于10%,最低检测限为0.1 ng/mL,回收率范围为95.3%~104.7%,抗干扰能力各项指标均达到较高水准。与Access 2化学发光免疫系统结果比较具有很好的相关性,得到回归方程:Y=0.84X-0.63,r=0.978。结论该系统对定量检测血清Prog 有较高的灵敏度和高特异性,且精密度符合临床要求,与Access 2检测系统有很好的相关性,值得推广。
目的:對UniCel DXI 800型全自動微粒子化學髮光分析儀定量檢測血清黃體酮(Prog)進行性能評價。方法運用該繫統檢測血清Prog,併對其靈敏度、精密度、準確度及特異性等進行方法學評價。同時與Access 2免疫分析繫統進行相關性比較。結果該繫統檢測血清 Prog 的精密度較高,變異繫數小于10%,最低檢測限為0.1 ng/mL,迴收率範圍為95.3%~104.7%,抗榦擾能力各項指標均達到較高水準。與Access 2化學髮光免疫繫統結果比較具有很好的相關性,得到迴歸方程:Y=0.84X-0.63,r=0.978。結論該繫統對定量檢測血清Prog 有較高的靈敏度和高特異性,且精密度符閤臨床要求,與Access 2檢測繫統有很好的相關性,值得推廣。
목적:대UniCel DXI 800형전자동미입자화학발광분석의정량검측혈청황체동(Prog)진행성능평개。방법운용해계통검측혈청Prog,병대기령민도、정밀도、준학도급특이성등진행방법학평개。동시여Access 2면역분석계통진행상관성비교。결과해계통검측혈청 Prog 적정밀도교고,변이계수소우10%,최저검측한위0.1 ng/mL,회수솔범위위95.3%~104.7%,항간우능력각항지표균체도교고수준。여Access 2화학발광면역계통결과비교구유흔호적상관성,득도회귀방정:Y=0.84X-0.63,r=0.978。결론해계통대정량검측혈청Prog 유교고적령민도화고특이성,차정밀도부합림상요구,여Access 2검측계통유흔호적상관성,치득추엄。
Objective To evaluate the quantitative detection of serum progesterone by UnicellDXI-800 automated chemilumines-cent analyzer.Methods To detect the serum progesterone,and evaluate the sensitivity,precision,accuracy and specificity.At the same time,the correlation with Access 2 immune system was analyzed as well.Results The precision of this system of serum pro-gesterone detection was high,with the variation coefficient less than 10%,the sensitivity of 0.1 ng/mL,and the recovery range of 95.3% to 104.7%.And the anti-jamming capability indexes reached higher level.It had good correlation compareing with the pre-vious Access 2 chemiluminescence immune system,and the regression equation obtained wasY= 0.84X-0.63,r= 0.978.Conclu-sion The system for quantitative detection of serum progesterone had high sensitivity and specificity,and its precision could consistent with clinical requirements.It has good correlation with Access 2 chemiluminescence immune system and worthy of promotion.