dry-type blood cell analyzer%blood cell analysis%performance evaluation
目的:通过与希思美康(SysmexXE-2100D)全自动血液分析仪的比对,分析干式血细胞计数仪(QBCRD)各项指标的临床诊断性能。方法采用双盲的方法对2台仪器检测的白细胞( white blood cell , WBC )、血红蛋白( hemoglobin , HB )、血小板(platelets,PLT)、红细胞比容(hematocrit,HCT)、粒细胞绝对值(granulocyte absolutely count,GRAN)、淋巴细胞及单核细胞绝对值( lymphocyte/monocyte absolutely count ,LY/MO)等6个参数进行比对,对仪器的精密度进行检测,计算线性回归方程及配对t检验。结果 QBCRD干式血细胞计数仪6项参数指标的CV均小于5%,表明该仪器精密度良好。2台仪器相关性显示WBC、HB、PLT、HCT、GRAN及LY/MO的结果相关性较好(r2分别为0.98、0.98、0.98、0.95、0.99和0.96),配对t检验,P值大部分均大于0.05,说明比对仪器与参考仪器6项检测结果之间差异无统计学意义。结论 QBCRD干式血细胞计数仪与SysmexXE-2100D全自动血液分析仪各参数之间测定结果一致,具有可比性。该仪器适合于部队野战医院、现场急救等情况下使用。
目的:通過與希思美康(SysmexXE-2100D)全自動血液分析儀的比對,分析榦式血細胞計數儀(QBCRD)各項指標的臨床診斷性能。方法採用雙盲的方法對2檯儀器檢測的白細胞( white blood cell , WBC )、血紅蛋白( hemoglobin , HB )、血小闆(platelets,PLT)、紅細胞比容(hematocrit,HCT)、粒細胞絕對值(granulocyte absolutely count,GRAN)、淋巴細胞及單覈細胞絕對值( lymphocyte/monocyte absolutely count ,LY/MO)等6箇參數進行比對,對儀器的精密度進行檢測,計算線性迴歸方程及配對t檢驗。結果 QBCRD榦式血細胞計數儀6項參數指標的CV均小于5%,錶明該儀器精密度良好。2檯儀器相關性顯示WBC、HB、PLT、HCT、GRAN及LY/MO的結果相關性較好(r2分彆為0.98、0.98、0.98、0.95、0.99和0.96),配對t檢驗,P值大部分均大于0.05,說明比對儀器與參攷儀器6項檢測結果之間差異無統計學意義。結論 QBCRD榦式血細胞計數儀與SysmexXE-2100D全自動血液分析儀各參數之間測定結果一緻,具有可比性。該儀器適閤于部隊野戰醫院、現場急救等情況下使用。
목적:통과여희사미강(SysmexXE-2100D)전자동혈액분석의적비대,분석간식혈세포계수의(QBCRD)각항지표적림상진단성능。방법채용쌍맹적방법대2태의기검측적백세포( white blood cell , WBC )、혈홍단백( hemoglobin , HB )、혈소판(platelets,PLT)、홍세포비용(hematocrit,HCT)、립세포절대치(granulocyte absolutely count,GRAN)、림파세포급단핵세포절대치( lymphocyte/monocyte absolutely count ,LY/MO)등6개삼수진행비대,대의기적정밀도진행검측,계산선성회귀방정급배대t검험。결과 QBCRD간식혈세포계수의6항삼수지표적CV균소우5%,표명해의기정밀도량호。2태의기상관성현시WBC、HB、PLT、HCT、GRAN급LY/MO적결과상관성교호(r2분별위0.98、0.98、0.98、0.95、0.99화0.96),배대t검험,P치대부분균대우0.05,설명비대의기여삼고의기6항검측결과지간차이무통계학의의。결론 QBCRD간식혈세포계수의여SysmexXE-2100D전자동혈액분석의각삼수지간측정결과일치,구유가비성。해의기괄합우부대야전의원、현장급구등정황하사용。
Objective A comprehensive performance evaluation was done for dry-type blood cell analyzer ( QBCRD ) based on comparison with Sysmex XE-2100D Blood cell analysis system.Methods White blood cell, hemoglobin, platelet, hematocrit, granulocyte absolute count , lymphocyte absolute count , and monocyte absolute count were tested by QBCRD and Sysmex XE-2100 D blood cell analysis system in double-blind trail.Precision of QBCRD was calculated , and also the linearity regression equation , paired T-test. Results CVs of the above 6 parameters were all less than 5%, showing a good precision .Good correlation with Sysmex XE-2100 D blood cell analysis system in the above 6 parameters(r2:0.99, 0.98, 0.98, 0.95, 0.99 and 0.96).There was no significant difference between QBCRD and Sysmex XE-2100 D blood cell analysis system in the above 6 parameters based on P value of paired T-test are all above 0.05.Conclusion All the items evaluated in QBCRD were highly correlated with System XE-2100D automated blood cell analysis system.QBCRD dry method blood cell analyzer is suitable for field hospital of army or on-the-spot first aid.