目的:对BC-6600型全自动血细胞分析仪(简称BC-6600分析仪)进行性能评价。方法对BC-6600型分析仪精密度、准确度、携带污染率、线性范围、可比性、白细胞分类计数等性能进行评价。结果 BC-6600型分析仪白细胞计数、红细胞计数、血红蛋白、血小板计数、网织红细胞计数结果具有良好的精密度和准确度,携带污染率分别为0.38%、0.48%、0.00%、0.98%、0.72%,在允许范围内;线性范围基本能覆盖临床标本浓度范围;与BC-5800型全自动血细胞分析仪检测结果具有良好的相关性;与手工分类计数相比,中性粒细胞、淋巴细胞和嗜酸粒细胞计数结果相关性良好,单核细胞计数结果相关性一般,嗜碱粒细胞计数结果相关性较差。结论 BC-6600型分析仪分析性能良好,能满足临床血液常规检测要求。
目的:對BC-6600型全自動血細胞分析儀(簡稱BC-6600分析儀)進行性能評價。方法對BC-6600型分析儀精密度、準確度、攜帶汙染率、線性範圍、可比性、白細胞分類計數等性能進行評價。結果 BC-6600型分析儀白細胞計數、紅細胞計數、血紅蛋白、血小闆計數、網織紅細胞計數結果具有良好的精密度和準確度,攜帶汙染率分彆為0.38%、0.48%、0.00%、0.98%、0.72%,在允許範圍內;線性範圍基本能覆蓋臨床標本濃度範圍;與BC-5800型全自動血細胞分析儀檢測結果具有良好的相關性;與手工分類計數相比,中性粒細胞、淋巴細胞和嗜痠粒細胞計數結果相關性良好,單覈細胞計數結果相關性一般,嗜堿粒細胞計數結果相關性較差。結論 BC-6600型分析儀分析性能良好,能滿足臨床血液常規檢測要求。
목적:대BC-6600형전자동혈세포분석의(간칭BC-6600분석의)진행성능평개。방법대BC-6600형분석의정밀도、준학도、휴대오염솔、선성범위、가비성、백세포분류계수등성능진행평개。결과 BC-6600형분석의백세포계수、홍세포계수、혈홍단백、혈소판계수、망직홍세포계수결과구유량호적정밀도화준학도,휴대오염솔분별위0.38%、0.48%、0.00%、0.98%、0.72%,재윤허범위내;선성범위기본능복개림상표본농도범위;여BC-5800형전자동혈세포분석의검측결과구유량호적상관성;여수공분류계수상비,중성립세포、림파세포화기산립세포계수결과상관성량호,단핵세포계수결과상관성일반,기감립세포계수결과상관성교차。결론 BC-6600형분석의분석성능량호,능만족림상혈액상규검측요구。
Objective To evaluate the performance of BC-6600 automated hematology analyzer .Methods Per-formance of precision ,accuracy ,carryover rate ,linearity ,comparability and differential counting were analyzed . Results The detection of white blood cells ,red blood cells ,hemoglobin ,platelet and reticulocytes were with fine pre-cision and accuracy ,and the carryover rates were 0 .38% ,0 .48% ,0 .00% ,0 .98% and 0 .72% .The linearity range covered the concentration range of clinical specimens .The detected results of BC-6600 and BC-5800 hematology ana-lyzer were with fine correlation .Compared with manual methods ,the counting of neutrophile granulocyte ,lymphocyte and eosnophils were fine correlated ,but the counting of monocyte was with low correlation .Conclusion The analysis performance of BC-6600 hematology analyzer could be fine ,which might be able to meet requirement for blood routine tests .