目的:探讨 Urit-1500尿液分析仪检测尿液红细胞(RBC)、白细胞(WBC)等细胞成分的临床准确性。方法选择该院2013年6~10月份收集的1085份尿液标本作为研究资料,分别采用 Urit-1500尿液分析仪测定与尿沉渣显微镜镜检,并对尿液分析仪检测尿细胞成分结果进行分析。结果 Urit-1500尿液分析仪测定1085份尿液标本中:其测定 RBC、WBC 的结果与显微镜法符合率分别为93.8%(1018/1085)、94.0%(1020/1085);RBC 阳性295例中经显微镜法镜检阴性49例,占16.6%;RBC 阴性790例中经显微镜法镜检阳性18例,占2.3%;WBC 阳性197例中经显微镜法镜检阴性28例,占14.2%;WBC 阴性888例中经显微镜法镜检阳性37例,占4.2%。结论 Urit-1500尿液分析仪法检测尿 RBC、WBC 时应结合显微镜镜检,并结合其他检验指标及临床资料进行综合分析,尽量减少误诊和漏诊。
目的:探討 Urit-1500尿液分析儀檢測尿液紅細胞(RBC)、白細胞(WBC)等細胞成分的臨床準確性。方法選擇該院2013年6~10月份收集的1085份尿液標本作為研究資料,分彆採用 Urit-1500尿液分析儀測定與尿沉渣顯微鏡鏡檢,併對尿液分析儀檢測尿細胞成分結果進行分析。結果 Urit-1500尿液分析儀測定1085份尿液標本中:其測定 RBC、WBC 的結果與顯微鏡法符閤率分彆為93.8%(1018/1085)、94.0%(1020/1085);RBC 暘性295例中經顯微鏡法鏡檢陰性49例,佔16.6%;RBC 陰性790例中經顯微鏡法鏡檢暘性18例,佔2.3%;WBC 暘性197例中經顯微鏡法鏡檢陰性28例,佔14.2%;WBC 陰性888例中經顯微鏡法鏡檢暘性37例,佔4.2%。結論 Urit-1500尿液分析儀法檢測尿 RBC、WBC 時應結閤顯微鏡鏡檢,併結閤其他檢驗指標及臨床資料進行綜閤分析,儘量減少誤診和漏診。
목적:탐토 Urit-1500뇨액분석의검측뇨액홍세포(RBC)、백세포(WBC)등세포성분적림상준학성。방법선택해원2013년6~10월빈수집적1085빈뇨액표본작위연구자료,분별채용 Urit-1500뇨액분석의측정여뇨침사현미경경검,병대뇨액분석의검측뇨세포성분결과진행분석。결과 Urit-1500뇨액분석의측정1085빈뇨액표본중:기측정 RBC、WBC 적결과여현미경법부합솔분별위93.8%(1018/1085)、94.0%(1020/1085);RBC 양성295례중경현미경법경검음성49례,점16.6%;RBC 음성790례중경현미경법경검양성18례,점2.3%;WBC 양성197례중경현미경법경검음성28례,점14.2%;WBC 음성888례중경현미경법경검양성37례,점4.2%。결론 Urit-1500뇨액분석의법검측뇨 RBC、WBC 시응결합현미경경검,병결합기타검험지표급림상자료진행종합분석,진량감소오진화루진。
Objective To explore the clinical accuracy of urine cellular components of RBC,WBC,etc.detected by the Urit-1500 urine analyzer.Methods 1 085 urine samples in this hospital from June to October 2013 were selected and detected by the Urit-1500 urine analyzer and the microscope detection respectively.The results of urine cellular components detected by the urine analy-zer were analyzed.Results In 1 085 urine samples,the coincidence rate of two detection methods in detecting RBC and WBC was 93.8%(1 018/1 085)and 94.0%(1 020/1 085)respectively.In 295 RBC positive urine samples tested by the urine analyzer,the number of negative results tested by microscopy was 49,accounting for 16.6%.In 790 RBC negative urine samples tested by the u-rine analyzer,the number of negative results tested by microscopy was 18,accounting for 2.3%.In 197 WBC positive urine samples tested by urine analyzer,the number of negative results tested by microscopy was 28,accounting for 14.2%.In 888 WBC negative urine samples by urine analyzer,the number of negative results tested by microscopy was 37,accounting for 4.2%.Conclusion De-tecting urine RBC and WBC by using the Urit-1500 urine analyzer should combined with the microscopy and other test indexes and even clinical data to conduct the comprehensive analysis for reducing the misdiagnosis and missed diagnosis as far as possible.