CHINESE JOURNAL OF DENTAL MATERIALS AND DEVICES
2014年
3期
137-140
,共4页
阮征%黄远亮%黄兰%吴倩雯%吴艳玲%郭华艳%嵇忠梅%章斯慎
阮徵%黃遠亮%黃蘭%吳倩雯%吳豔玲%郭華豔%嵇忠梅%章斯慎
원정%황원량%황란%오천문%오염령%곽화염%혜충매%장사신
骨结合%种植体稳定系数%共振频率分析%锥束CT%骨密度
骨結閤%種植體穩定繫數%共振頻率分析%錐束CT%骨密度
골결합%충식체은정계수%공진빈솔분석%추속CT%골밀도
Osseointegration%Implant stability quotient%Resonance frequency analysis%CBCT%Bone density
目的:应用锥束CT三维重建影像技术对后牙种植区骨密度进行定量测量,同时结合OS-STEL种植体稳定系数(ISQ),分析表示骨密度的豪森菲尔德单位(HU)值及种植体稳定系数(ISQ)值对植入种植体稳定性的影响。方法对32名(男14人,女18人)后牙种植修复患者的49枚种植体进行术前测量种受植部位HU值,记录植入最大扭矩(Ncm),在种植初期及5个月后进行共振频率分析(RFA)。结果49枚后牙区种植体全部存留;植入区HU值:477.76±129.88;植入时最大扭矩:35.82±10.275;初期ISQ:77.55±6.84;骨结合后ISQ:78.78±6.25。植入最大扭矩与初期ISQ(P=0.851)、植入区HU值与初期ISQ(P=0.721)未检出相关性,而植入区HU值与骨结合后ISQ值则有显著相关性(P<0.01)。结论骨密度与种植体骨结合后稳定性密切相关,骨密度HU值越高,预后的种植体稳定性越高。
目的:應用錐束CT三維重建影像技術對後牙種植區骨密度進行定量測量,同時結閤OS-STEL種植體穩定繫數(ISQ),分析錶示骨密度的豪森菲爾德單位(HU)值及種植體穩定繫數(ISQ)值對植入種植體穩定性的影響。方法對32名(男14人,女18人)後牙種植脩複患者的49枚種植體進行術前測量種受植部位HU值,記錄植入最大扭矩(Ncm),在種植初期及5箇月後進行共振頻率分析(RFA)。結果49枚後牙區種植體全部存留;植入區HU值:477.76±129.88;植入時最大扭矩:35.82±10.275;初期ISQ:77.55±6.84;骨結閤後ISQ:78.78±6.25。植入最大扭矩與初期ISQ(P=0.851)、植入區HU值與初期ISQ(P=0.721)未檢齣相關性,而植入區HU值與骨結閤後ISQ值則有顯著相關性(P<0.01)。結論骨密度與種植體骨結閤後穩定性密切相關,骨密度HU值越高,預後的種植體穩定性越高。
목적:응용추속CT삼유중건영상기술대후아충식구골밀도진행정량측량,동시결합OS-STEL충식체은정계수(ISQ),분석표시골밀도적호삼비이덕단위(HU)치급충식체은정계수(ISQ)치대식입충식체은정성적영향。방법대32명(남14인,녀18인)후아충식수복환자적49매충식체진행술전측량충수식부위HU치,기록식입최대뉴구(Ncm),재충식초기급5개월후진행공진빈솔분석(RFA)。결과49매후아구충식체전부존류;식입구HU치:477.76±129.88;식입시최대뉴구:35.82±10.275;초기ISQ:77.55±6.84;골결합후ISQ:78.78±6.25。식입최대뉴구여초기ISQ(P=0.851)、식입구HU치여초기ISQ(P=0.721)미검출상관성,이식입구HU치여골결합후ISQ치칙유현저상관성(P<0.01)。결론골밀도여충식체골결합후은정성밀절상관,골밀도HU치월고,예후적충식체은정성월고。
Objective To analysis of the influence of bone mineral density in Hounsfield units values (HU) and implant stability factor value (ISQ ) on implant stability in posterior region using Cone beam computed tomography(CBCT) combined with OSSTEL implant stability factor (ISQ). Methods Hounsfield units( HU) value and the maximum torque of 49 implants of 32 patients (male 14, female 18, mean age 49 years) with poste-rior implants were measured before implantation and then the resonance frequency analysis (RFA)were recorded at 0, 5 months after implantation. Results All of 49 posterior teeth implants survived;HU value in the implanted region:477.76 ± 129.88;the implant maximum torque:35.82 ± 10.275;the initial ISQ:77.55 ± 6.84; ISQ at 5 months:78.78+6.25. There was no significant difference between the maximum torque and initial ISQ (P=0.851), as well as the HU value and initial ISQ (P=0.721);There were significant differences between HU value and ISQ value at 5 months (P<0.01). Conclusion The bone mineral density is closely related to the implant stability in posterior region. Higher the HU values of bone mineral density, higher the implant stability.