microwave digestion%PTFE filter%PM2. 5%heavy metal elements%ICP-MS
采用4种微波消解体系对聚四氟乙烯( PTFE)材质滤膜采集的PM2.5样品进行消解,利用电感耦合等离子体质谱( ICP-MS)测定样品中Cr、Mn、Cu、As、Cd 5种重金属元素含量,并计算其空白加标和样品加标回收率,建立了测定PTFE材质滤膜采集的PM2.5样品中重金属元素的最优微波消解体系。结果表明,4种消解体系下,空白加标回收率和样品加标回收率均处于合理范围,分别为95?3%~116?3%和93?3%~118?5%。其中,HNO3放置过夜+H2 O2消解体系具有空白值低、测定稳定性强、操作简便、绿色环保、对人体健康危害小等优点,能够满足大批量PM2.5样品中重金属总量的快速、高效、准确分析的要求。
採用4種微波消解體繫對聚四氟乙烯( PTFE)材質濾膜採集的PM2.5樣品進行消解,利用電感耦閤等離子體質譜( ICP-MS)測定樣品中Cr、Mn、Cu、As、Cd 5種重金屬元素含量,併計算其空白加標和樣品加標迴收率,建立瞭測定PTFE材質濾膜採集的PM2.5樣品中重金屬元素的最優微波消解體繫。結果錶明,4種消解體繫下,空白加標迴收率和樣品加標迴收率均處于閤理範圍,分彆為95?3%~116?3%和93?3%~118?5%。其中,HNO3放置過夜+H2 O2消解體繫具有空白值低、測定穩定性彊、操作簡便、綠色環保、對人體健康危害小等優點,能夠滿足大批量PM2.5樣品中重金屬總量的快速、高效、準確分析的要求。
채용4충미파소해체계대취사불을희( PTFE)재질려막채집적PM2.5양품진행소해,이용전감우합등리자체질보( ICP-MS)측정양품중Cr、Mn、Cu、As、Cd 5충중금속원소함량,병계산기공백가표화양품가표회수솔,건립료측정PTFE재질려막채집적PM2.5양품중중금속원소적최우미파소해체계。결과표명,4충소해체계하,공백가표회수솔화양품가표회수솔균처우합리범위,분별위95?3%~116?3%화93?3%~118?5%。기중,HNO3방치과야+H2 O2소해체계구유공백치저、측정은정성강、조작간편、록색배보、대인체건강위해소등우점,능구만족대비량PM2.5양품중중금속총량적쾌속、고효、준학분석적요구。
Four kinds of microwave digestion system were studied to digest the PM2. 5 samples collected with PTFE filter. The content of Cr、Mn、Cu、As、Cd in the samples were determined by ICP-MS under optimized conditions. Meanwhile, the recovery rates of blank and samples in standard addition were calculated. Finally, the optimal digestion system to digest PM2. 5 sample collected with PTFE filter was established. The results showed that both of the recovery rates of blank and samples in standard addition under the four kinds of digestion system were basically in a reasonable range, which were between 95. 3%-116. 3% and 93. 3%-118. 5%, respectively. However, the digestion system of HNO3 ( placed overnight after added) + H2 O2 could be best satisfied with the demand of rapid, efficient and accurate analysis of the total quantities of heavy metals on a large scale of PM2. 5 samples because of the advantages of low blank value, better stability of determination, easy to operate, green environmental protection, less harmful to health and so on.