中华临床医师杂志(电子版)
中華臨床醫師雜誌(電子版)
중화림상의사잡지(전자판)
CHINESE JOURNAL OF CLINICIANS(ELECTRONIC VERSION)
2014年
21期
3822-3826
,共5页
林俊东%王祖琛%洪嘉志%郑亚明
林俊東%王祖琛%洪嘉誌%鄭亞明
림준동%왕조침%홍가지%정아명
椎间盘%体层摄影术,X线计算机%轴位扫描%低剂量
椎間盤%體層攝影術,X線計算機%軸位掃描%低劑量
추간반%체층섭영술,X선계산궤%축위소묘%저제량
Intervertebral disk%Tomography,X-ray computed%Axial scan%Low-dose
目的:探讨CT轴位扫描应用于腰椎间盘扫描以降低辐射剂量的可行性及其临床应用价值。方法随机抽取正常BMI的60例志愿者行CT腰椎间盘检查,在其他参数均相同情况下分别进行常规剂量(270 mA)和低剂量(180 mA、135 mA)扫描,比较两种不同低剂量CT扫描所得图像与常规剂量的图像质量差异,以及发现椎间盘病变的差异。结果在扫描图像质量方面,两种不同低剂量与常规剂量扫描有统计学差异(P<0.05)。但在诊断疾病方面,两种不同低剂量扫描与常规剂量扫描无统计学差异(P>0.05)。在满足诊断需要的同时,患者接受的辐射剂量分别降低36.5%及54.0%。结论轴位扫描应用于腰椎间盘CT扫描具有一定的可行性,并可大幅度降低患者辐射剂量。
目的:探討CT軸位掃描應用于腰椎間盤掃描以降低輻射劑量的可行性及其臨床應用價值。方法隨機抽取正常BMI的60例誌願者行CT腰椎間盤檢查,在其他參數均相同情況下分彆進行常規劑量(270 mA)和低劑量(180 mA、135 mA)掃描,比較兩種不同低劑量CT掃描所得圖像與常規劑量的圖像質量差異,以及髮現椎間盤病變的差異。結果在掃描圖像質量方麵,兩種不同低劑量與常規劑量掃描有統計學差異(P<0.05)。但在診斷疾病方麵,兩種不同低劑量掃描與常規劑量掃描無統計學差異(P>0.05)。在滿足診斷需要的同時,患者接受的輻射劑量分彆降低36.5%及54.0%。結論軸位掃描應用于腰椎間盤CT掃描具有一定的可行性,併可大幅度降低患者輻射劑量。
목적:탐토CT축위소묘응용우요추간반소묘이강저복사제량적가행성급기림상응용개치。방법수궤추취정상BMI적60례지원자행CT요추간반검사,재기타삼수균상동정황하분별진행상규제량(270 mA)화저제량(180 mA、135 mA)소묘,비교량충불동저제량CT소묘소득도상여상규제량적도상질량차이,이급발현추간반병변적차이。결과재소묘도상질량방면,량충불동저제량여상규제량소묘유통계학차이(P<0.05)。단재진단질병방면,량충불동저제량소묘여상규제량소묘무통계학차이(P>0.05)。재만족진단수요적동시,환자접수적복사제량분별강저36.5%급54.0%。결론축위소묘응용우요추간반CT소묘구유일정적가행성,병가대폭도강저환자복사제량。
Objective To discuss the clinical significance and the possibilities of low dose CT scanning with axial CT technology in lumbar intervertebral disc. Methods The sample was randomly selected of 60 volunteers from healthy body mass index (BMI) in patients with lumbar intervertebral disc diseases, with the other scanning parameters kept unchanged for all the cases in each group, were examined by MSCT respectively in standard dose (270 mA) and different low-dose (180 mA, 135 mA). Then, a comparison was made to evaluate the differences in image quality and detection of the lesion between the low-dose CT scanning and the standard dose CT scanning. Results The imaging was evaluated by the image quality score, there was statistical difference between the low-dose CT scanning and the standard dose CT scanning (P<0.05). However, there was not statistical difference in detection of the lesion between the low-dose CT scanning and the standard dose CT scanning (P>0.05), which did not influence doctor to view and make diagnosis.Comparing to 270 mA as the standard, the dose length product (DLP) of 180 mA and 135 mA was decreased by 36.5%and 54.0%respectively. Conclusion Low-dose CT scanning with axial technology in lumbar intervertebral disc is practicable and the radiation dose can be greatly reduced.