手术器械%清洗%多酶清洗液%碱性清洗液
手術器械%清洗%多酶清洗液%堿性清洗液
수술기계%청세%다매청세액%감성청세액
Surgical instruments%Cleaning%Multiple enzyme cleaning fluid%Alkaline cleaning fluid
目的:探讨多酶清洗液与碱性清洗液保湿对使用后手术器械的清洗效果。方法选取2013年6-9月手术室剖宫产使用后的手术器械1200件,立即回收,采用随机数字表法将器械分为Ⅰ、Ⅱ、Ⅲ、Ⅳ组,每组300件。其中Ⅰ组采用多酶清洗液保湿2 h后清洗;Ⅱ组采用多酶清洗液保湿12 h后清洗;Ⅲ组采用碱性清洗液保湿2 h后清洗;Ⅳ组采用碱性清洗液保湿12 h后清洗,比较各组器械的清洗合格率。结果保湿2 h,多酶清洗液(Ⅰ组)、碱性清洗液(Ⅲ组)目测、镜下及试纸检测器检测器械清洗合格率比较,差异无统计学意义(P>0.05)。保湿12 h,碱性清洗液(Ⅳ组)目测合格292件(97.33%),镜下合格289件(96.33%),试纸检测合格291件(97.00%),分别高于多酶清洗液(Ⅱ组)的261件(87.00%),259件(86.33%),259件(86.33%),差异有统计学意义(χ2值分别为18.690,18.950,14.592;P<0.01)。采用多酶清洗液2 h(Ⅰ组)目测合格296件(98.67%),镜下合格294件(98.00%),试纸检测合格295件(98.33%),分别明显高于12 h (Ⅱ组),差异有统计学意义(χ2值分别为30.688,22.017,30.513;P<0.01);采用碱性清洗液2 h(Ⅲ组)与12 h(Ⅳ组)保湿处理后检测器械清洗合格率比较,差异均无统计学意义( P>0.05)。结论对正常使用后的手术器械宜在2 h内及时清洗,采用多酶清洗液或碱性清洗液保湿均可以确保清洗质量;不能在2h内回收清洗的器械,宜先采用碱性清洗液保湿处理,再使用机洗的清洗方法,以有效提高器械清洗质量。
目的:探討多酶清洗液與堿性清洗液保濕對使用後手術器械的清洗效果。方法選取2013年6-9月手術室剖宮產使用後的手術器械1200件,立即迴收,採用隨機數字錶法將器械分為Ⅰ、Ⅱ、Ⅲ、Ⅳ組,每組300件。其中Ⅰ組採用多酶清洗液保濕2 h後清洗;Ⅱ組採用多酶清洗液保濕12 h後清洗;Ⅲ組採用堿性清洗液保濕2 h後清洗;Ⅳ組採用堿性清洗液保濕12 h後清洗,比較各組器械的清洗閤格率。結果保濕2 h,多酶清洗液(Ⅰ組)、堿性清洗液(Ⅲ組)目測、鏡下及試紙檢測器檢測器械清洗閤格率比較,差異無統計學意義(P>0.05)。保濕12 h,堿性清洗液(Ⅳ組)目測閤格292件(97.33%),鏡下閤格289件(96.33%),試紙檢測閤格291件(97.00%),分彆高于多酶清洗液(Ⅱ組)的261件(87.00%),259件(86.33%),259件(86.33%),差異有統計學意義(χ2值分彆為18.690,18.950,14.592;P<0.01)。採用多酶清洗液2 h(Ⅰ組)目測閤格296件(98.67%),鏡下閤格294件(98.00%),試紙檢測閤格295件(98.33%),分彆明顯高于12 h (Ⅱ組),差異有統計學意義(χ2值分彆為30.688,22.017,30.513;P<0.01);採用堿性清洗液2 h(Ⅲ組)與12 h(Ⅳ組)保濕處理後檢測器械清洗閤格率比較,差異均無統計學意義( P>0.05)。結論對正常使用後的手術器械宜在2 h內及時清洗,採用多酶清洗液或堿性清洗液保濕均可以確保清洗質量;不能在2h內迴收清洗的器械,宜先採用堿性清洗液保濕處理,再使用機洗的清洗方法,以有效提高器械清洗質量。
목적:탐토다매청세액여감성청세액보습대사용후수술기계적청세효과。방법선취2013년6-9월수술실부궁산사용후적수술기계1200건,립즉회수,채용수궤수자표법장기계분위Ⅰ、Ⅱ、Ⅲ、Ⅳ조,매조300건。기중Ⅰ조채용다매청세액보습2 h후청세;Ⅱ조채용다매청세액보습12 h후청세;Ⅲ조채용감성청세액보습2 h후청세;Ⅳ조채용감성청세액보습12 h후청세,비교각조기계적청세합격솔。결과보습2 h,다매청세액(Ⅰ조)、감성청세액(Ⅲ조)목측、경하급시지검측기검측기계청세합격솔비교,차이무통계학의의(P>0.05)。보습12 h,감성청세액(Ⅳ조)목측합격292건(97.33%),경하합격289건(96.33%),시지검측합격291건(97.00%),분별고우다매청세액(Ⅱ조)적261건(87.00%),259건(86.33%),259건(86.33%),차이유통계학의의(χ2치분별위18.690,18.950,14.592;P<0.01)。채용다매청세액2 h(Ⅰ조)목측합격296건(98.67%),경하합격294건(98.00%),시지검측합격295건(98.33%),분별명현고우12 h (Ⅱ조),차이유통계학의의(χ2치분별위30.688,22.017,30.513;P<0.01);채용감성청세액2 h(Ⅲ조)여12 h(Ⅳ조)보습처리후검측기계청세합격솔비교,차이균무통계학의의( P>0.05)。결론대정상사용후적수술기계의재2 h내급시청세,채용다매청세액혹감성청세액보습균가이학보청세질량;불능재2h내회수청세적기계,의선채용감성청세액보습처리,재사용궤세적청세방법,이유효제고기계청세질량。
Objective To explore the cleaning effect of moisturizing through multiple enzymes cleaning fluid or alkaline cleaning fluid in the cleaning of medical equipment .Methods One thousand and two hundred surgical instruments used by cesarean delivery and recycled immediately from June 2013 to September 2013 were chosen and divided into the Ⅰ,Ⅱ,Ⅲ,Ⅳgroup according to the random number table , each with 300 .the ⅠandⅡgroup received the cleaning 2, 12 h after the moisturizing through the multiple enzymes cleaning fluid , respectively;and Ⅲ and Ⅳ group received the cleaning 2, 12 h after the moisturizing through the alkaline cleaning fluid , respectively .The qualification rates of cleaning were compared among the groups .Results No differences were found in the qualification rates of cleaning through the visual , microscopically and test paper detection between Ⅰ and Ⅲ groups ( P >0.05 ).The qualification rates of cleaning through the visual , microscopically and test paper detection were respectively 97.33% (292/300),96.33% (289/300) and 97.00%(291/300) in Ⅳ group, and were higher than 87.00% (261/300),86.33% (259/300) and 86.33%(259/300) in Ⅱ group, and the differences were statistically significant (χ2 =18.690,18.950, 14.592, respectively;P<0.01).The qualification rates of cleaning through the visual , microscopically and test paper detection were respectively 98.67% (296/300),98.00% (294/300) and 98.33% (295/300) in Ⅰgroup, and were higher than those of Ⅱgroup, and the differences were statistically significant (χ2 =30.688, 22.017,30.513, respectively;P <0.01).No differences were found in the qualification rates of cleaning betweenⅢandⅣgroups (P>0.05).Conclusions The surgical instruments after the normal use should be cleaned within 2 hours, and the moisturizing through multiple enzymes cleaning fluid or alkaline cleaning fluid can all ensure the cleaning quality .The surgical instruments after the use can not be cleaned within 2 hours, and should be firstly moisturized through the alkaline cleaning fluid and secondly cleaned through the machine so as to improve the cleaning quality of instruments .