原子能科学技术
原子能科學技術
원자능과학기술
ATOMIC ENERGY SCIENCE AND TECHNOLOGY
2014年
11期
2116-2122
,共7页
孔栋%顾思毅%倪婕%孙亮
孔棟%顧思毅%倪婕%孫亮
공동%고사의%예첩%손량
加速器%电子对效应%正电子%湮没辐射%污染电子
加速器%電子對效應%正電子%湮沒輻射%汙染電子
가속기%전자대효응%정전자%인몰복사%오염전자
accelerator%electron-pair effect%positron%annihilation radiation%contami-nation electron
以Geant4编程构建6 M V医用加速器机头和照射条件,并通过实测数据比对验证模型的正确性,模拟加速器粒子输运过程,获取不同射野和不同层面的粒子相空间文件,以正电子和湮没光子为指标研究射野及各机头部件对电子对效应产生的影响,模拟电子对效应对加速器射束及水模中剂量沉积的影响。结果表明:射野面积越小,电子对效应产生越多;在靶、均整器及次级准直器处均会有较明显的电子对效应发生,其中次级准直器是主要的发生部件;电子对效应使射束中污染电子增加了约4%,降低了射束的平均能量,使射束粒子分布更加分散,增加了模体表面剂量及射野外剂量。
以Geant4編程構建6 M V醫用加速器機頭和照射條件,併通過實測數據比對驗證模型的正確性,模擬加速器粒子輸運過程,穫取不同射野和不同層麵的粒子相空間文件,以正電子和湮沒光子為指標研究射野及各機頭部件對電子對效應產生的影響,模擬電子對效應對加速器射束及水模中劑量沉積的影響。結果錶明:射野麵積越小,電子對效應產生越多;在靶、均整器及次級準直器處均會有較明顯的電子對效應髮生,其中次級準直器是主要的髮生部件;電子對效應使射束中汙染電子增加瞭約4%,降低瞭射束的平均能量,使射束粒子分佈更加分散,增加瞭模體錶麵劑量及射野外劑量。
이Geant4편정구건6 M V의용가속기궤두화조사조건,병통과실측수거비대험증모형적정학성,모의가속기입자수운과정,획취불동사야화불동층면적입자상공간문건,이정전자화인몰광자위지표연구사야급각궤두부건대전자대효응산생적영향,모의전자대효응대가속기사속급수모중제량침적적영향。결과표명:사야면적월소,전자대효응산생월다;재파、균정기급차급준직기처균회유교명현적전자대효응발생,기중차급준직기시주요적발생부건;전자대효응사사속중오염전자증가료약4%,강저료사속적평균능량,사사속입자분포경가분산,증가료모체표면제량급사야외제량。
The Geant4 w as used to construct the head and irradiation conditions of 6MV medical accelerator . The measurement data in the water phantom were compared to ensure the constructed model correct .Phase space files of different fields and different levels were obtained by simulating the process of particle transporting of the accelera‐tor .The positron and annihilation photon were as indicator ,and the effect of fields and components of the accelerator head on the production of electron‐pair was studied by analyzing those files .The influence of electron‐pair effect on accelerator beam and ener‐gy deposited in the phantom was simulated .The results show that the smaller the field is ,the more the production of electron‐pair is .The production of electron‐pair is obvi‐ous when beam passes through target ,flatten filter and secondary collimator ,and it is especially obvious to secondary collimator .The electron‐pair effect makes contamination electron increase about 4% ,beam’s average energy lower ,beam’s particle angular dis‐tribution more dispersed and doses of surface of phantom and outside fields increase .