目的:探讨Techno TwinStation全自动血型仪在ABO、Rh(D)血型鉴定及不规则抗体筛查中的应用价值。方法 Techno TwinStation全自动血型仪对41577例标本进行ABO正反定型、Rh(D)血型鉴定;对1959例标本进行不规则抗体的筛查,并与手工法(血型鉴定采用试管法,不规则抗体筛查采用微柱凝胶法)进行比较。结果41577例标本中全自动血型仪与试管法ABO血型鉴定的准确率分别为99.80%和99.75%,无统计学意义(﹥0.05),Rh(D)血型鉴定准确率均为100%,不规则抗体筛查仪器法与手工微柱凝胶法的阳性率均为0.31%(5/1959),假阳性0例。结论 Techno全自动血型仪在ABO,Rh(D)血型鉴定及不规则抗体筛查中可避免人为差错,提高工作效率,其结果更安全,可靠。
目的:探討Techno TwinStation全自動血型儀在ABO、Rh(D)血型鑒定及不規則抗體篩查中的應用價值。方法 Techno TwinStation全自動血型儀對41577例標本進行ABO正反定型、Rh(D)血型鑒定;對1959例標本進行不規則抗體的篩查,併與手工法(血型鑒定採用試管法,不規則抗體篩查採用微柱凝膠法)進行比較。結果41577例標本中全自動血型儀與試管法ABO血型鑒定的準確率分彆為99.80%和99.75%,無統計學意義(﹥0.05),Rh(D)血型鑒定準確率均為100%,不規則抗體篩查儀器法與手工微柱凝膠法的暘性率均為0.31%(5/1959),假暘性0例。結論 Techno全自動血型儀在ABO,Rh(D)血型鑒定及不規則抗體篩查中可避免人為差錯,提高工作效率,其結果更安全,可靠。
목적:탐토Techno TwinStation전자동혈형의재ABO、Rh(D)혈형감정급불규칙항체사사중적응용개치。방법 Techno TwinStation전자동혈형의대41577례표본진행ABO정반정형、Rh(D)혈형감정;대1959례표본진행불규칙항체적사사,병여수공법(혈형감정채용시관법,불규칙항체사사채용미주응효법)진행비교。결과41577례표본중전자동혈형의여시관법ABO혈형감정적준학솔분별위99.80%화99.75%,무통계학의의(﹥0.05),Rh(D)혈형감정준학솔균위100%,불규칙항체사사의기법여수공미주응효법적양성솔균위0.31%(5/1959),가양성0례。결론 Techno전자동혈형의재ABO,Rh(D)혈형감정급불규칙항체사사중가피면인위차착,제고공작효솔,기결과경안전,가고。
Objective To study the applied value of Techno TwinStation Automate Analyzing Unit in ABO and Rh(D)blood type identification and screen the irregular antibodies.Methods ABO and Rh (D)blood group were studied in 41577 samples and the irregular antibodies were screened in 1959 samples using Techno TwinStation Automate Analyzing Unit.Results Compared with manual method (we used test tube method to identify the ABO blood typing and micro-column gel technique to screen ir egular antibodies).Results The accuracy of Techno TwinStation Automate Analyzing Unit and test tube method in ABO blood type identification were 99.80%and 99.75%respectively,there was no significant dif erence.The accuracy of Rh(D)detection was 100%using two methods.The positive rate of the irregular antibodies was 0.31%(5/1959)no mat er using Automate Analyzing Unit or micro-column gel technique,and there were no false positive results.Conclusion The usage of Techno TwinStation Automate Analyzing Unit in ABO and Rh (D)blood type identification and the ir egular antibodies detection can avoid mistakes to occur and raise the work ef iciency,the results are safe and reliable.