blood typing analyzer%ABO blood type%antibody titer
目的:探索使用PK7300全自动血型分析仪测定ABO血型抗体效价的方法,探讨献血者血型检测中与ABO血型抗体效价相关的影响因素及注意事项。方法仪器微板法测定抽样标本的ABO血型抗体效价;仪器微板法及手工试管法测定各血型混合血浆的ABO血型抗体效价;统计分析该中心2013年5~7月检测出的ABO血型抗体效价降低的情况。结果A型献血者与B型献血者的ABO血型抗体效价比较,差异无统计学意义(P>0.05),O型献血者的ABO血型抗体效价明显高于A型或B型献血者(P<0.05);手工试管法的灵敏度明显优于仪器微板法,而手工试管法二的灵敏度又明显优于手工试管法一(P<0.05);ABO血型抗体效价降低的A型献血者明显多于B型或O型(均P<0.01),ABO血型抗体效价降低的男性献血者明显多于女性(P<0.01)。结论仪器微板法测定ABO血型抗体效价操作简便、结果判读客观、重复性好,1∶1~1∶80的稀释比例范围能满足绝大多数标本的测定需要,采用梯度稀释有助于减小测定误差;血型、性别、年龄、亚型等因素可能与献血者ABO血型抗体效价降低相关。
目的:探索使用PK7300全自動血型分析儀測定ABO血型抗體效價的方法,探討獻血者血型檢測中與ABO血型抗體效價相關的影響因素及註意事項。方法儀器微闆法測定抽樣標本的ABO血型抗體效價;儀器微闆法及手工試管法測定各血型混閤血漿的ABO血型抗體效價;統計分析該中心2013年5~7月檢測齣的ABO血型抗體效價降低的情況。結果A型獻血者與B型獻血者的ABO血型抗體效價比較,差異無統計學意義(P>0.05),O型獻血者的ABO血型抗體效價明顯高于A型或B型獻血者(P<0.05);手工試管法的靈敏度明顯優于儀器微闆法,而手工試管法二的靈敏度又明顯優于手工試管法一(P<0.05);ABO血型抗體效價降低的A型獻血者明顯多于B型或O型(均P<0.01),ABO血型抗體效價降低的男性獻血者明顯多于女性(P<0.01)。結論儀器微闆法測定ABO血型抗體效價操作簡便、結果判讀客觀、重複性好,1∶1~1∶80的稀釋比例範圍能滿足絕大多數標本的測定需要,採用梯度稀釋有助于減小測定誤差;血型、性彆、年齡、亞型等因素可能與獻血者ABO血型抗體效價降低相關。
목적:탐색사용PK7300전자동혈형분석의측정ABO혈형항체효개적방법,탐토헌혈자혈형검측중여ABO혈형항체효개상관적영향인소급주의사항。방법의기미판법측정추양표본적ABO혈형항체효개;의기미판법급수공시관법측정각혈형혼합혈장적ABO혈형항체효개;통계분석해중심2013년5~7월검측출적ABO혈형항체효개강저적정황。결과A형헌혈자여B형헌혈자적ABO혈형항체효개비교,차이무통계학의의(P>0.05),O형헌혈자적ABO혈형항체효개명현고우A형혹B형헌혈자(P<0.05);수공시관법적령민도명현우우의기미판법,이수공시관법이적령민도우명현우우수공시관법일(P<0.05);ABO혈형항체효개강저적A형헌혈자명현다우B형혹O형(균P<0.01),ABO혈형항체효개강저적남성헌혈자명현다우녀성(P<0.01)。결론의기미판법측정ABO혈형항체효개조작간편、결과판독객관、중복성호,1∶1~1∶80적희석비례범위능만족절대다수표본적측정수요,채용제도희석유조우감소측정오차;혈형、성별、년령、아형등인소가능여헌혈자ABO혈형항체효개강저상관。
Objective To explore the detection method of ABO blood type antibody titration by the PK7300 automated blood typing analyzer and to investigate the ABO blood type antibody titer‐related influencing factors and the matter needing attentions in blood typing test for blood donors .Methods The ABO blood type antibody titers of the samples were measured by the analyzer microplate method;the ABO blood type antibody titer of each mixed plas‐ma was respectively measured by the analyzer microplate method and by the manual tube method;the decrease situa‐tion of ABO blood type antibody titers detected in our center from May to July 2013 were statistically analyzed .Re‐sults There was no statistically significant difference in the ABO blood type antibody titers between the blood type A donors and blood type B donors(P>0 .05) ,the ABO blood type antibody titers in the blood type O donors were significantly higher than those in the blood type A or blood type B donors(P<0 .05);the sensitivity of the manual tube method was significantly superior to that of the analyzer microplate method ,and the sensitivity of the manual tube method 2 was significantly superior to that of the manual tube method 1(P<0 .05);the blood type A donors with decreased ABO blood type antibody titers were significantly more than the blood type B donors or blood type O donors(P<0 .01) ,and the male donors with decreased ABO antibody titers were significantly more than the female donors(P<0 .01) .Conclusion The analyzer microplate method for detecting the ABO blood type antibody titration is simple to operate ,objective in the result judgment ,has good repeatability ,the dilution range of 1∶1 -1∶80 can meet the measurement requirements for most of samples and gradient dilution may contribute to reduce the measure‐ment errors;the factors of blood type ,gender ,age and subtype may be related to the decreased ABO antibody titers of blood donors .