CHINA MEDICAL EQUIPMENT
2014年
11期
131-133,120
,共4页
全自动血液体液分析仪%精密度%携带污染率%线性范围%准确度
全自動血液體液分析儀%精密度%攜帶汙染率%線性範圍%準確度
전자동혈액체액분석의%정밀도%휴대오염솔%선성범위%준학도
fully-automatic blood and body lfuid analyzer%precision%contamination rate%linearity range%accuracy
目的:对Sysmex XN-9000全自动血液体液分析仪的主要性能进行初步评价。方法根据美国临床和实验室标准化协会(CLSI)指南文件EP15-A2、EP6-A2的要求,对XN-9000仪进行精密度、携带污染率、线性范围和准确度验证,以Beckman LH750全自动血液分析仪的检测结果作为评估依据,将验证结果与厂商声明的性能指标或公认的质量标准进行比较。结果 XN-9000仪测定白细胞(WBC)、红细胞(RBC)、血红蛋白(HGB)、红细胞压积(HCT)和血小板(PLT)主要参数的批内变异系数(CV%)为0.11%~2.42%,批间CV%为0.12%~2.50%,线性相关性分析r为0.99991~0.99999,携带污染率为0.00%~0.38%。对比相关性分析,各项指标r均>0.999。准确度验证各项指标的偏倚为-0.67%~0.96%。结论 XN-9000仪的精密度、携带污染率、线性范围和准确度等主要评估指标均符合要求。
目的:對Sysmex XN-9000全自動血液體液分析儀的主要性能進行初步評價。方法根據美國臨床和實驗室標準化協會(CLSI)指南文件EP15-A2、EP6-A2的要求,對XN-9000儀進行精密度、攜帶汙染率、線性範圍和準確度驗證,以Beckman LH750全自動血液分析儀的檢測結果作為評估依據,將驗證結果與廠商聲明的性能指標或公認的質量標準進行比較。結果 XN-9000儀測定白細胞(WBC)、紅細胞(RBC)、血紅蛋白(HGB)、紅細胞壓積(HCT)和血小闆(PLT)主要參數的批內變異繫數(CV%)為0.11%~2.42%,批間CV%為0.12%~2.50%,線性相關性分析r為0.99991~0.99999,攜帶汙染率為0.00%~0.38%。對比相關性分析,各項指標r均>0.999。準確度驗證各項指標的偏倚為-0.67%~0.96%。結論 XN-9000儀的精密度、攜帶汙染率、線性範圍和準確度等主要評估指標均符閤要求。
목적:대Sysmex XN-9000전자동혈액체액분석의적주요성능진행초보평개。방법근거미국림상화실험실표준화협회(CLSI)지남문건EP15-A2、EP6-A2적요구,대XN-9000의진행정밀도、휴대오염솔、선성범위화준학도험증,이Beckman LH750전자동혈액분석의적검측결과작위평고의거,장험증결과여엄상성명적성능지표혹공인적질량표준진행비교。결과 XN-9000의측정백세포(WBC)、홍세포(RBC)、혈홍단백(HGB)、홍세포압적(HCT)화혈소판(PLT)주요삼수적비내변이계수(CV%)위0.11%~2.42%,비간CV%위0.12%~2.50%,선성상관성분석r위0.99991~0.99999,휴대오염솔위0.00%~0.38%。대비상관성분석,각항지표r균>0.999。준학도험증각항지표적편의위-0.67%~0.96%。결론 XN-9000의적정밀도、휴대오염솔、선성범위화준학도등주요평고지표균부합요구。
Objective To evaluate the main performance of Sysmex XN-9000 fully-automatic blood and body fluid analyzer. Methods According to the guidelines EP15-A2 and EP6-A2 of clinical and laboratory standards institute (CLSI), the precision, contamination rate, linearity range and accuracy of XN-9000 analyzer were verified based on the detection results of Beckman LH750 full-automatic blood analyzer which were choosed as assessment basis. Then the verification results of XN-9000 analyzer were compared with performance indicators declared by its manufacturer or recognized quality standards. Results The intra-assay and inter-assay coefficient variations (CV%) of major parameters including WBC, RBC, HGB, HCT and PLT measured by XN-9000 analyzer were 0.11%~2.42% and 0.12%~2.50%, respectively. The correlation coefficient (r ) of linear correlation analysis between observational values and theoretical values of the above parameters was 0.99991~0.99999 while the contamination rate of the above parameters was 0.00%~0.38%. The correlation coefficient (r ) of comparative correlation analysis between the detection indicators of XN-9000 analyzer and those of Backman LH750 analyzer was greater than 0.999. The Bias of accuracy verification of the above parameters of XN-9000 analyzer was -0.67%~0.96%. Conclusion The main parameters including precision, contamination rate, linearity range and accuracy of XN-9000 analyzer can meet the requirements of relevant standards.