采取对照试验,对各色谱条件下氯酸根-次氯酸根离子混标实验进行分析,得出其优化条件:低比例淋洗液及较高柱温条件下,对常见阴离子的测定优化有益;而大体积进样较稍大体积进样,对其测定的改善并不明显。
採取對照試驗,對各色譜條件下氯痠根-次氯痠根離子混標實驗進行分析,得齣其優化條件:低比例淋洗液及較高柱溫條件下,對常見陰離子的測定優化有益;而大體積進樣較稍大體積進樣,對其測定的改善併不明顯。
채취대조시험,대각색보조건하록산근-차록산근리자혼표실험진행분석,득출기우화조건:저비례림세액급교고주온조건하,대상견음리자적측정우화유익;이대체적진양교초대체적진양,대기측정적개선병불명현。
By taking controlled trials, the mixed standard experiment of ClO3 --ClO - is analyzed under the chromatographic conditions, the optimal conditions are obtained: it is beneficial for common anions to measure and optimize under low proportion of eluent and higher column temperature conditions; and the improvement of determination is not obvious between large volume sample injection and larger volume sample injection.