IOL Master%ALLEGRO Oculyzer%Anterior chamber%Cataract
目的 比较IOL Master与ALLEGRO Oculyzer测量前房深度的差异,并对两者的一致性进行评价.方法 前瞻性研究.对第三军医大学大坪医院眼科拟行白内障手术的患者53例(87眼),分别用IOL Master和ALLEGRO Oculyzer进行前房深度测量.采用配对t检验比较两者结果的差异,采用Pearson相关分析与Bland-Altman分析评价两者的一致性.结果 IOL Master与ALLEGRO Oculyzer测量的前房深度值分别为(3.17±0.44) mm和(3.21±0.46)mm,差异有统计学意义(t=3.26,P<0.05).两种方法所得测量值有高度相关性(r=0.96,P<0.01);Bland-Altman分析95%一致性界限为-0.20~0.28 mm,两者有较好的一致性.结论 IOL Master与ALLEGRO Oculyzer测量的前房深度值可以在人工晶状体的计算中交互使用.
目的 比較IOL Master與ALLEGRO Oculyzer測量前房深度的差異,併對兩者的一緻性進行評價.方法 前瞻性研究.對第三軍醫大學大坪醫院眼科擬行白內障手術的患者53例(87眼),分彆用IOL Master和ALLEGRO Oculyzer進行前房深度測量.採用配對t檢驗比較兩者結果的差異,採用Pearson相關分析與Bland-Altman分析評價兩者的一緻性.結果 IOL Master與ALLEGRO Oculyzer測量的前房深度值分彆為(3.17±0.44) mm和(3.21±0.46)mm,差異有統計學意義(t=3.26,P<0.05).兩種方法所得測量值有高度相關性(r=0.96,P<0.01);Bland-Altman分析95%一緻性界限為-0.20~0.28 mm,兩者有較好的一緻性.結論 IOL Master與ALLEGRO Oculyzer測量的前房深度值可以在人工晶狀體的計算中交互使用.
목적 비교IOL Master여ALLEGRO Oculyzer측량전방심도적차이,병대량자적일치성진행평개.방법 전첨성연구.대제삼군의대학대평의원안과의행백내장수술적환자53례(87안),분별용IOL Master화ALLEGRO Oculyzer진행전방심도측량.채용배대t검험비교량자결과적차이,채용Pearson상관분석여Bland-Altman분석평개량자적일치성.결과 IOL Master여ALLEGRO Oculyzer측량적전방심도치분별위(3.17±0.44) mm화(3.21±0.46)mm,차이유통계학의의(t=3.26,P<0.05).량충방법소득측량치유고도상관성(r=0.96,P<0.01);Bland-Altman분석95%일치성계한위-0.20~0.28 mm,량자유교호적일치성.결론 IOL Master여ALLEGRO Oculyzer측량적전방심도치가이재인공정상체적계산중교호사용.
Objective To compare the difference and assess the consistency between IOL Master and ALLEGRO Oculyzer for measuring anterior chamber depth (ACD).Methods It was a prospective study.Fifty-three patients (87 eyes) were recruited and ACD was measured by IOL Master and ALLEGRO Oculyzer.The statistical significance of the difference between the measurements was evaluated by a paired t test and consistency was assessed by Pearson's correlation and Bland-Altman analysis.Results The mean ACD was 3.17±0.44 mm and 3.21±0.46 mm with IOL Master and ALLEGRO Oculyzer,respectively.The difference between the two measurements was statistically significant (t=3.26,P<0.01) and a high correlation was found between the two measurements (r=0.96,P<0.01).Bland-Altmann analysis showed that the two measurements had comparable results for ACD with 95% limits of consistency,-0.20 to 0.28 mm.Conclusion IOL Master and ALLEGRO Oculyzer can be used interchangeably for ACD measurement to calculate intraocular lens power.