目的 评估Sirius、Orbscan Ⅱ及超声测厚仪测量中央角膜厚度(CCT)的差异和一致性.方法 对250例(500只眼)近视眼患者,依次采用Sirius、Orbscan Ⅱ及USP测量CCT.结果 Siri-US、OrbscanⅡ及USP测得的平均CCT分别为(525.7±36.7) μm、(515.5±44.1)μm、(529.1±35.8)μm.3种仪器测量CCT高度相关:Sirius与USP(r=0.964,P<0.001)、Orbscan Ⅱ与USP(r=0.932,P <0.001)、Sirius与Orbscan Ⅱ(r=0.940,P<0.001).CCT的差值均数(包括一致性界限的上线和下线)为Sirius与USP (3.4±9.8)μm、(-15.7,22.6)、Orbscan Ⅱ与USP (13.6±16.9) μm(-19.4,46.7)、Sirius与Orbscan Ⅱ(10.2土15.8)μm(-20.8,41.2).结论 与USP相比,Sirius与Orbscan Ⅱ低估了中央角膜厚度.与Orbscan Ⅱ相比,Sirius与USP有较好的相关性和一致性.3种仪器测量CCT不可互相替换.Sirius可能逐渐取代角膜厚度测量的金标准.
目的 評估Sirius、Orbscan Ⅱ及超聲測厚儀測量中央角膜厚度(CCT)的差異和一緻性.方法 對250例(500隻眼)近視眼患者,依次採用Sirius、Orbscan Ⅱ及USP測量CCT.結果 Siri-US、OrbscanⅡ及USP測得的平均CCT分彆為(525.7±36.7) μm、(515.5±44.1)μm、(529.1±35.8)μm.3種儀器測量CCT高度相關:Sirius與USP(r=0.964,P<0.001)、Orbscan Ⅱ與USP(r=0.932,P <0.001)、Sirius與Orbscan Ⅱ(r=0.940,P<0.001).CCT的差值均數(包括一緻性界限的上線和下線)為Sirius與USP (3.4±9.8)μm、(-15.7,22.6)、Orbscan Ⅱ與USP (13.6±16.9) μm(-19.4,46.7)、Sirius與Orbscan Ⅱ(10.2土15.8)μm(-20.8,41.2).結論 與USP相比,Sirius與Orbscan Ⅱ低估瞭中央角膜厚度.與Orbscan Ⅱ相比,Sirius與USP有較好的相關性和一緻性.3種儀器測量CCT不可互相替換.Sirius可能逐漸取代角膜厚度測量的金標準.
목적 평고Sirius、Orbscan Ⅱ급초성측후의측량중앙각막후도(CCT)적차이화일치성.방법 대250례(500지안)근시안환자,의차채용Sirius、Orbscan Ⅱ급USP측량CCT.결과 Siri-US、OrbscanⅡ급USP측득적평균CCT분별위(525.7±36.7) μm、(515.5±44.1)μm、(529.1±35.8)μm.3충의기측량CCT고도상관:Sirius여USP(r=0.964,P<0.001)、Orbscan Ⅱ여USP(r=0.932,P <0.001)、Sirius여Orbscan Ⅱ(r=0.940,P<0.001).CCT적차치균수(포괄일치성계한적상선화하선)위Sirius여USP (3.4±9.8)μm、(-15.7,22.6)、Orbscan Ⅱ여USP (13.6±16.9) μm(-19.4,46.7)、Sirius여Orbscan Ⅱ(10.2토15.8)μm(-20.8,41.2).결론 여USP상비,Sirius여Orbscan Ⅱ저고료중앙각막후도.여Orbscan Ⅱ상비,Sirius여USP유교호적상관성화일치성.3충의기측량CCT불가호상체환.Sirius가능축점취대각막후도측량적금표준.
Objective To assess the difference and agreement of central corneal thickness measurements between Sirius,Orbscan Ⅱ,and ultrasound pachymetry (USP).Methods On 250 myopic patients (500 eyes),the central corneal thickness was measured using Sirius,Orbscan Ⅱ and ultrasound pachymetry in turn.Results The mean measurements of average central corneal thickness by Sirius,Orbscan Ⅱ,and USP were 525.7±36.7μm,515.5±44.1μm and 529.1±35.8μm respectively.There was high correlation between the three instruments:Sirius with USP (r =0.964,P <0.001),Orbscan Ⅱ with USP (r =0.932,P <0.001) and Sirius with Orbscan Ⅱ (r =0.940,P <0.001).The mean differences (and upper/lower LOA) for central corneal thickness measurements were 3.4±9.8μm (-15.7,22.6) between Sirius and USP, 13.6±16.9μm (-19.4,46.7) between Orbscan Ⅱ and USP,and 10.2±15.8μm (-20.8,41.2) between Sirius and Orbscan Ⅱ.Conclusions Sirius and Orbscan Ⅱ underestimate corneal thickness compared with that measured with USP.Sirius has better correlation and agreement with the gold standard USP,as compared with Orbscan Ⅱ.The three instruments can not be used interchangeably for measuring central corneal thickness.Sirius eventually may replace USP as the gold standard for corneal pachymetry.