目的 研究常规冠状动脉造影(CCA)与采用XperSwing技术冠状动脉造影的辐射剂量差异.方法 应用中国人仿真胸部体模,采用XperSwing和CCA2种检查技术,将热释光个人剂量计(TLD)布放于仿真胸部体模的主要组织器官,测量各组织器官的吸收剂量并读取设备的剂量面积乘积(DAP)值.实际测量有效剂量等于组织器官吸收剂量乘以相应组织权重因子后的总和;设备读取的有效剂量等于DAP乘以转换因子(0.185 mSv/ Gy·cm2).结果 采用XperSwing技术造影的组织器官平均吸收剂量低于CCA检查;CCA和XperSwing技术造影的设备读取平均有效剂量分别为3.88和3.13 mSv,低于实际测量的平均有效剂量(5.57和4.94 mSv).结论 采用XperSwing技术的辐射剂量低于CCA,设备读取的有效剂量有可能低估了患者实际接受的有效剂量水平.
目的 研究常規冠狀動脈造影(CCA)與採用XperSwing技術冠狀動脈造影的輻射劑量差異.方法 應用中國人倣真胸部體模,採用XperSwing和CCA2種檢查技術,將熱釋光箇人劑量計(TLD)佈放于倣真胸部體模的主要組織器官,測量各組織器官的吸收劑量併讀取設備的劑量麵積乘積(DAP)值.實際測量有效劑量等于組織器官吸收劑量乘以相應組織權重因子後的總和;設備讀取的有效劑量等于DAP乘以轉換因子(0.185 mSv/ Gy·cm2).結果 採用XperSwing技術造影的組織器官平均吸收劑量低于CCA檢查;CCA和XperSwing技術造影的設備讀取平均有效劑量分彆為3.88和3.13 mSv,低于實際測量的平均有效劑量(5.57和4.94 mSv).結論 採用XperSwing技術的輻射劑量低于CCA,設備讀取的有效劑量有可能低估瞭患者實際接受的有效劑量水平.
목적 연구상규관상동맥조영(CCA)여채용XperSwing기술관상동맥조영적복사제량차이.방법 응용중국인방진흉부체모,채용XperSwing화CCA2충검사기술,장열석광개인제량계(TLD)포방우방진흉부체모적주요조직기관,측량각조직기관적흡수제량병독취설비적제량면적승적(DAP)치.실제측량유효제량등우조직기관흡수제량승이상응조직권중인자후적총화;설비독취적유효제량등우DAP승이전환인자(0.185 mSv/ Gy·cm2).결과 채용XperSwing기술조영적조직기관평균흡수제량저우CCA검사;CCA화XperSwing기술조영적설비독취평균유효제량분별위3.88화3.13 mSv,저우실제측량적평균유효제량(5.57화4.94 mSv).결론 채용XperSwing기술적복사제량저우CCA,설비독취적유효제량유가능저고료환자실제접수적유효제량수평.
Objective To evaluate the effective organ radiation dose in patients with conventional coronary angiography(CCA) and XperSwing coronary angiography.Methods The operating environment of CCA and XperSwing coronary angiography was simulated using the Chinese anthropomorphic chest phantom.The thermoluminescent dosimeters (TLDs) were placed on major organs of the phantomm.The organs' absorbed doses were measured and the dose-area products (DAPs) were recorded.The real effective dose was the summation of products of every organs'absorbed dose × a factor.The calculated effective doses was the product of the DAP × a factor.Results The average absorbed effective doses of XperSwing coronary angiography were lower than that of the CCA.The calculated effective doses of CCA and XperSwing coronary angiography were 3.88 mSv and 3.13 mSv,respectively,and less than the real ones (5.57 mSv,4.94 mSv).Conclusions XperSwing coronary angiography would have less radiation dose than CCA does.The calculated effective dose would be less than the real one.