中华放射肿瘤学杂志
中華放射腫瘤學雜誌
중화방사종류학잡지
CHINESE JOURNAL OF RADIATION ONCOLOGY
2014年
6期
527-529
,共3页
岳麒%段继梅%王志伟%谷丹%杨秀美%李荣清
嶽麒%段繼梅%王誌偉%穀丹%楊秀美%李榮清
악기%단계매%왕지위%곡단%양수미%리영청
螺旋断层疗法%体层摄影术,X线计算机%密度-CT值曲线%Cheesephantom模体
螺鏇斷層療法%體層攝影術,X線計算機%密度-CT值麯線%Cheesephantom模體
라선단층요법%체층섭영술,X선계산궤%밀도-CT치곡선%Cheesephantom모체
Helical Tomotherapy%Tomography,X-ray computed%Image value to density table curve%Water Cheesephantom
目的 探讨HT的MVCT扫描剂量率波动和不同扫描精度选择对图像CT值及噪声的影响.方法 MVCT在不同剂量率和扫描层厚下扫描带各种密度棒的水模体,对应各密度棒的CT值得出各自条件下的IVDT曲线进行对比.MVCT在不同剂量率和扫描层厚下扫描带有均匀密度棒的水模体,测量不同扫描条件下均匀水模体中的CT图像噪声.对结果进行方差及相关性分析.结果 不同扫描剂量率下对CT值影响显著(P=0.000),各密度棒对应CT值的变化大小与密度呈正相关(R2 =0.846),既高密度物质的CT值受的影响较大.不同剂量率对噪声也有显著影响,扫描剂量率越低图像噪声越大(P=0.000).2、4、6 mm扫描层厚对CT值(P=1.000)及图像噪声(P=0.667)无显著影响.结论 MVCT剂量率改变会引起CT值及图像噪声改变,在临床中应定期对MVCT剂量率及IVDT曲线稳定性进行监测,以保证ART计划中MVCT图像剂量计算的准确.
目的 探討HT的MVCT掃描劑量率波動和不同掃描精度選擇對圖像CT值及譟聲的影響.方法 MVCT在不同劑量率和掃描層厚下掃描帶各種密度棒的水模體,對應各密度棒的CT值得齣各自條件下的IVDT麯線進行對比.MVCT在不同劑量率和掃描層厚下掃描帶有均勻密度棒的水模體,測量不同掃描條件下均勻水模體中的CT圖像譟聲.對結果進行方差及相關性分析.結果 不同掃描劑量率下對CT值影響顯著(P=0.000),各密度棒對應CT值的變化大小與密度呈正相關(R2 =0.846),既高密度物質的CT值受的影響較大.不同劑量率對譟聲也有顯著影響,掃描劑量率越低圖像譟聲越大(P=0.000).2、4、6 mm掃描層厚對CT值(P=1.000)及圖像譟聲(P=0.667)無顯著影響.結論 MVCT劑量率改變會引起CT值及圖像譟聲改變,在臨床中應定期對MVCT劑量率及IVDT麯線穩定性進行鑑測,以保證ART計劃中MVCT圖像劑量計算的準確.
목적 탐토HT적MVCT소묘제량솔파동화불동소묘정도선택대도상CT치급조성적영향.방법 MVCT재불동제량솔화소묘층후하소묘대각충밀도봉적수모체,대응각밀도봉적CT치득출각자조건하적IVDT곡선진행대비.MVCT재불동제량솔화소묘층후하소묘대유균균밀도봉적수모체,측량불동소묘조건하균균수모체중적CT도상조성.대결과진행방차급상관성분석.결과 불동소묘제량솔하대CT치영향현저(P=0.000),각밀도봉대응CT치적변화대소여밀도정정상관(R2 =0.846),기고밀도물질적CT치수적영향교대.불동제량솔대조성야유현저영향,소묘제량솔월저도상조성월대(P=0.000).2、4、6 mm소묘층후대CT치(P=1.000)급도상조성(P=0.667)무현저영향.결론 MVCT제량솔개변회인기CT치급도상조성개변,재림상중응정기대MVCT제량솔급IVDT곡선은정성진행감측,이보증ART계화중MVCT도상제량계산적준학.
Objective To study the CT number and noise of HT MVCT on different dose rate and scanning thickness.Methods The CT number of different relative electron density were measured in the MVCT image of Cheesephantom with tissue substitute plugs scanned with different dose rate and slice thickness.The physical density corresponding to the CT number was plotted as the image value to density table (IVDT).The noise was measured in the MVCT image of Cheesephantom with solid water plugs scanned with different dose rate and slice thickness.Results There was a significant different of the CT number of the plugs with different dose rates (P =0.000),it shows a positive correlation between the varied CT number and density (R2 =0.846),there is larger impact on the high density number.There was still a significant effect on the noise with different dose rate (P =0.000 density),the noise increase as the dose rate decrease.There was no significant effect on IVDT (P =1.000) and noise (P =0.667) with different slice thickness.Conclusions The CT number and the noise vary with the dose rate,the QA of MVCT should be performed regularly to assure the quality of image and the accuracy of dose calculating on MVCT in adaptive.