中华放射肿瘤学杂志
中華放射腫瘤學雜誌
중화방사종류학잡지
CHINESE JOURNAL OF RADIATION ONCOLOGY
2014年
6期
535-539
,共5页
李胜元%李骁杨%严峥%高立%周宣%惠璐璐
李勝元%李驍楊%嚴崢%高立%週宣%惠璐璐
리성원%리효양%엄쟁%고립%주선%혜로로
磁共振成像定位%伽马刀治疗精度%多点检测
磁共振成像定位%伽馬刀治療精度%多點檢測
자공진성상정위%가마도치료정도%다점검측
Magnetic resonance imaging position%Treatment accuracy of Gamma knife%Multi-dot detection
目的 探讨MRI定位伽马刀治疗精度的检测与修正方法.方法 利用25点矩阵水箱对伽马刀治疗精度进行多点检测,将MRI定位坐标与经过CT定位曝光验证、治疗精度偏差<0.5 mm的坐标进行比对,发现MRI定位偏差的大小及图像有无变形,然后按偏差均值±图像几何畸变率给予修正,与标准坐标间无显著差异后再行胶片曝光验证.结果 所测3家医院MRI定位25点矩阵水箱的x、z坐标轴均无明显偏移,在y轴上存在的偏差分别为A院(-1.94±0.45)mm、B院(-2.22±0.29) mm和C院(-1.25±0.21)mm,A院在)轴上尚有1%的几何畸变率.以M表示偏差均值、GD为几何畸变率、Y0为矩阵中心点坐标,将25点坐标Y分别代入公式Y修=(Y-M) +GD (Y0-Y)修正后,3家医院与标准坐标间均相近,验证片焦斑偏移也在0.5mm内.结论 利用25点矩阵水箱进行伽马刀治疗精度多点检测和偏差的函数修正,可以明确MRI能否直接用于头部伽马刀定位.
目的 探討MRI定位伽馬刀治療精度的檢測與脩正方法.方法 利用25點矩陣水箱對伽馬刀治療精度進行多點檢測,將MRI定位坐標與經過CT定位曝光驗證、治療精度偏差<0.5 mm的坐標進行比對,髮現MRI定位偏差的大小及圖像有無變形,然後按偏差均值±圖像幾何畸變率給予脩正,與標準坐標間無顯著差異後再行膠片曝光驗證.結果 所測3傢醫院MRI定位25點矩陣水箱的x、z坐標軸均無明顯偏移,在y軸上存在的偏差分彆為A院(-1.94±0.45)mm、B院(-2.22±0.29) mm和C院(-1.25±0.21)mm,A院在)軸上尚有1%的幾何畸變率.以M錶示偏差均值、GD為幾何畸變率、Y0為矩陣中心點坐標,將25點坐標Y分彆代入公式Y脩=(Y-M) +GD (Y0-Y)脩正後,3傢醫院與標準坐標間均相近,驗證片焦斑偏移也在0.5mm內.結論 利用25點矩陣水箱進行伽馬刀治療精度多點檢測和偏差的函數脩正,可以明確MRI能否直接用于頭部伽馬刀定位.
목적 탐토MRI정위가마도치료정도적검측여수정방법.방법 이용25점구진수상대가마도치료정도진행다점검측,장MRI정위좌표여경과CT정위폭광험증、치료정도편차<0.5 mm적좌표진행비대,발현MRI정위편차적대소급도상유무변형,연후안편차균치±도상궤하기변솔급여수정,여표준좌표간무현저차이후재행효편폭광험증.결과 소측3가의원MRI정위25점구진수상적x、z좌표축균무명현편이,재y축상존재적편차분별위A원(-1.94±0.45)mm、B원(-2.22±0.29) mm화C원(-1.25±0.21)mm,A원재)축상상유1%적궤하기변솔.이M표시편차균치、GD위궤하기변솔、Y0위구진중심점좌표,장25점좌표Y분별대입공식Y수=(Y-M) +GD (Y0-Y)수정후,3가의원여표준좌표간균상근,험증편초반편이야재0.5mm내.결론 이용25점구진수상진행가마도치료정도다점검측화편차적함수수정,가이명학MRI능부직접용우두부가마도정위.
Objective To investigate MRI localization the accuracy and correct the deviation for gamma knife treatment.Methods With 25-point-matrix tank,the deviation of MRI localization and its regularities could be identified after the comparison between the coordinates of MRI localization and the ones which have already been verified by CT within the deviation of 0.5 mm.Then the original MRI coordinates will be corrected by the acquired mean deviation and the geometric distortion of images.Afterwards the corrected coordinates will be compared with the standard ones and finally validated by exposure film.Results There are no significant deviations on x-and z-axis after measurement in three hospitals,y-axis,however,bears deviation of (1.94 ±0.45) mm for hospital A,(-2.22 ±0.29) mm for hospital B,(-1.25 ±0.21) mm for hospital C,respectively.Furthermore there also exists geometric distortion of 1% on y-axis in hospital A.The corrected coordinates on y-axis (Yc) will be calculated from the formula:Yc =(Y-M) + GD (Y0-Y) (Y:the original coordinates on y-axis,M:the mean of deviation on y-axis,GD:the geometric distortion,Y0:the coordinate on y-axis of the central point among the 25-point matrix).Once completed,the corrected coordinates of MRI localization is of no significant difference with the standard coordinates verified by CT.Even the deviation of focal spot on validation film is within 0.5 mm.Conclusions The 25-point-matrix tank in the multi-point measurement of the accuracy and the correction of deviation for gamma knife treatment is feasible to determine whether MRI can be utilized in the localization for head gamma knife treatment.