中华神经医学杂志
中華神經醫學雜誌
중화신경의학잡지
CHINESE JOURNAL OF NEUROMEDICINE
2013年
11期
1133-1135
,共3页
刘彦超%李西锋%段传志%张炘%何旭英%顾大群%李辉%苏世星%赖凌峰
劉彥超%李西鋒%段傳誌%張炘%何旭英%顧大群%李輝%囌世星%賴凌峰
류언초%리서봉%단전지%장흔%하욱영%고대군%리휘%소세성%뢰릉봉
微导管%塑形%Jasper弹簧圈
微導管%塑形%Jasper彈簧圈
미도관%소형%Jasper탄황권
Microcather%Steam shaping%Jasper Guglielmi detachable coil
目的 探讨临床常用的Prowler 14(P14)、Enchelon 10(E10)和SL 10 3种微导管在蒸汽塑形前后其末端长度的变化,为临床上判断弹簧圈的最佳解脱位置提供参考. 方法 将Jasper 弹簧圈(1D、2D和3D)分别推入P 14、E 10和SL 10 45°塑形、单弯塑形和双弯塑形前后的三种微导管中,当Jasper弹簧圈的显影点推出微导管近端显影点远端1mm时,测量微导管末端到弹簧圈解脱区近端推送导丝的长度变化. 结果 3种微导管分别在3种不同角度塑形前后,微导管末端到弹簧圈解脱区近端推送导丝的长度变化差异均有统计学意义(P<0.05).3种微导管在塑形后其末段长度均缩短,且微导管末端采用双弯塑形时,其末端平均缩短长度大于单弯塑形和45°塑形时的长度,其中45°塑形时其末端缩短长度最小. 结论 蒸气塑形会引起微导管末端长度变短.在一定范围内,末端塑形弯曲度越大,微导管末端缩短越明显.
目的 探討臨床常用的Prowler 14(P14)、Enchelon 10(E10)和SL 10 3種微導管在蒸汽塑形前後其末耑長度的變化,為臨床上判斷彈簧圈的最佳解脫位置提供參攷. 方法 將Jasper 彈簧圈(1D、2D和3D)分彆推入P 14、E 10和SL 10 45°塑形、單彎塑形和雙彎塑形前後的三種微導管中,噹Jasper彈簧圈的顯影點推齣微導管近耑顯影點遠耑1mm時,測量微導管末耑到彈簧圈解脫區近耑推送導絲的長度變化. 結果 3種微導管分彆在3種不同角度塑形前後,微導管末耑到彈簧圈解脫區近耑推送導絲的長度變化差異均有統計學意義(P<0.05).3種微導管在塑形後其末段長度均縮短,且微導管末耑採用雙彎塑形時,其末耑平均縮短長度大于單彎塑形和45°塑形時的長度,其中45°塑形時其末耑縮短長度最小. 結論 蒸氣塑形會引起微導管末耑長度變短.在一定範圍內,末耑塑形彎麯度越大,微導管末耑縮短越明顯.
목적 탐토림상상용적Prowler 14(P14)、Enchelon 10(E10)화SL 10 3충미도관재증기소형전후기말단장도적변화,위림상상판단탄황권적최가해탈위치제공삼고. 방법 장Jasper 탄황권(1D、2D화3D)분별추입P 14、E 10화SL 10 45°소형、단만소형화쌍만소형전후적삼충미도관중,당Jasper탄황권적현영점추출미도관근단현영점원단1mm시,측량미도관말단도탄황권해탈구근단추송도사적장도변화. 결과 3충미도관분별재3충불동각도소형전후,미도관말단도탄황권해탈구근단추송도사적장도변화차이균유통계학의의(P<0.05).3충미도관재소형후기말단장도균축단,차미도관말단채용쌍만소형시,기말단평균축단장도대우단만소형화45°소형시적장도,기중45°소형시기말단축단장도최소. 결론 증기소형회인기미도관말단장도변단.재일정범위내,말단소형만곡도월대,미도관말단축단월명현.
Objective To investigate the distal-tip changes in length of three different microcathers (Prowler 14,Enchelon 10 and SL10) after steam shaping and provide reference for clinician to make the optimum marker alignment for cerebral aneurysm embolization with Jasper Guglielmi detachable coils (JasperlD,Jasper2D and Jasper3D).Methods The distance from the catheter tip to the proximal of coil detachment zone was measured when the proximal coil marker was positioned 1 mm exactly distal to the distal end of the proximal catheter marker before and after the end of microcatheter modified by 45 degree,single curved and double curved shape with steaming.Results Steam shaping had a significant association with the distal-tip changes in length when each type of microcatheters was modified by 45 degree,single curved and double curved shape (P<0.05).In addition,the length shortening of three types of microcatheters changedmost under double curved shape and single curved,while changed least under 45 degree shape.Conclusions Microcather curving with steam shaping can shorten the length of distal-tip.The more the microcather curved,the more obvious the length of shorten within a certain range.