本体感觉%位置觉%康复评定%信度%膝关节
本體感覺%位置覺%康複評定%信度%膝關節
본체감각%위치각%강복평정%신도%슬관절
Proprioception%Joint position sense%Rehabilitation evaluation%Reliability%Knee joint
目的 研究等速测试系统评定膝关节位置觉的重复测试信度和不同测试者间测试信度.方法 应用Biodex System 3型等速系统测试16例健康志愿者双侧膝关节位置觉(采用被动角度重建测试法),选择屈膝30°、45°和60°为目标角度,记录受试者实际角度与目标角度间的差值绝对值(误差角).由2名测试者负责重复测试和不同测试者间重复测试,前、后2次测试中间间隔3d.结果 受试者屈膝30°、45°和60°时,其误差角的重复测试信度组内相关系数(ICC)分别为0.728、0.780和0.807,不同测试者间信度Pearson相关系数分别为0.676、0.610和0.705,提示均具有显著相关性(P<0.05).通过单因素方差分析发现受试者双膝、前后2次测试时各目标角度误差角间差异均无统计学意义(P>0.05),所有受试者角度重建时误差角的95%可信区间上限均<3°.结论 采用等速系统评定受试者膝关节位置觉具有较好的重复测试信度,在屈膝30 ~60°范围内选择1个目标角度测试即可,当受试者膝关节角度重建误差角>3 a时应考虑其临床意义.
目的 研究等速測試繫統評定膝關節位置覺的重複測試信度和不同測試者間測試信度.方法 應用Biodex System 3型等速繫統測試16例健康誌願者雙側膝關節位置覺(採用被動角度重建測試法),選擇屈膝30°、45°和60°為目標角度,記錄受試者實際角度與目標角度間的差值絕對值(誤差角).由2名測試者負責重複測試和不同測試者間重複測試,前、後2次測試中間間隔3d.結果 受試者屈膝30°、45°和60°時,其誤差角的重複測試信度組內相關繫數(ICC)分彆為0.728、0.780和0.807,不同測試者間信度Pearson相關繫數分彆為0.676、0.610和0.705,提示均具有顯著相關性(P<0.05).通過單因素方差分析髮現受試者雙膝、前後2次測試時各目標角度誤差角間差異均無統計學意義(P>0.05),所有受試者角度重建時誤差角的95%可信區間上限均<3°.結論 採用等速繫統評定受試者膝關節位置覺具有較好的重複測試信度,在屈膝30 ~60°範圍內選擇1箇目標角度測試即可,噹受試者膝關節角度重建誤差角>3 a時應攷慮其臨床意義.
목적 연구등속측시계통평정슬관절위치각적중복측시신도화불동측시자간측시신도.방법 응용Biodex System 3형등속계통측시16례건강지원자쌍측슬관절위치각(채용피동각도중건측시법),선택굴슬30°、45°화60°위목표각도,기록수시자실제각도여목표각도간적차치절대치(오차각).유2명측시자부책중복측시화불동측시자간중복측시,전、후2차측시중간간격3d.결과 수시자굴슬30°、45°화60°시,기오차각적중복측시신도조내상관계수(ICC)분별위0.728、0.780화0.807,불동측시자간신도Pearson상관계수분별위0.676、0.610화0.705,제시균구유현저상관성(P<0.05).통과단인소방차분석발현수시자쌍슬、전후2차측시시각목표각도오차각간차이균무통계학의의(P>0.05),소유수시자각도중건시오차각적95%가신구간상한균<3°.결론 채용등속계통평정수시자슬관절위치각구유교호적중복측시신도,재굴슬30 ~60°범위내선택1개목표각도측시즉가,당수시자슬관절각도중건오차각>3 a시응고필기림상의의.
Objective To determine the intra-and inter-tester reliability of a dynamometer in assessment of knee joint position senses (JPS).Methods Sixteen healthy volunteers (intra-rater N =6,inter-rater N =10) were examined with regard to intra-and inter-tester reliability of knee JPS by using Biodex System 3 dynamometer,measured by passive replication test,with subjects passively returning to the initial positions at 30°,45° and 60° knee flexion.The absolute error (AE) angle was recorded.Both intra-and inter-tester reliability studies involved two testers.Results The intra-tester reliability of the AE were moderate reliable at 30°,45°and 60° knee flexion,and the values of intra-class correlation coefficients (ICC) were 0.728,0.780 and 0.807,respectively.The values of Pearson's correlation coefficients for inter-tester reliability were 0.676,0.610 and 0.705,respectively (P < 0.05).There was no significant difference between the AE values of dominance and non-dominant knees,test and retest,and in the AE values of 30°,45°and 60° knee flexion (P > 0.05).The 95% confidence upper limits of all subjects' AE values were less than 3°.Conclusion The major findings of this study suggest that the test-retest reliability of the knee JPS were moderate using Biodex System 3 dynamometer,and selecting a target angle of passive replication test can achieve the same purpose in 30° to 60°knee flexion.The clinical significance should be taken into consideration when the AE changes were greater than 3 degrees.