中华眼科杂志
中華眼科雜誌
중화안과잡지
Chinese Journal of Ophthalmology
2009年
12期
1084-1088
,共5页
陈伟蓉%孟倩丽%叶海昀%刘奕志
陳偉蓉%孟倩麗%葉海昀%劉奕誌
진위용%맹천려%협해윤%류혁지
晶体%人工%像差%对比敏感度%深度知觉
晶體%人工%像差%對比敏感度%深度知覺
정체%인공%상차%대비민감도%심도지각
Lens%intraocular%Aberration%Contrast sensitivity%Depth perception
目的 探讨超声乳化白内障吸除联合双眼分别植入折射型与衍射型多焦点人工晶状体(MIOL)术后的波前像差、对比敏感度(CS)及立体视功能.方法 前瞻性研究.将年龄相关性白内障患者分为MIOL组与单焦点人工晶状体(SIOL)组,每组各15例(30只眼),行超声乳化白内障吸除联合人工晶状体(IOL)植入术.MIOL组为主导眼植入ReZoom NXG1 IOL(即NXG1组),对侧眼植入Tecnis ZM900 IOL(即ZM900组);SIOL组为同期双眼植入Sensar AR40e IOL(即AR40e组).观察患者术后波前像差、CS、眩光敏感度(GS)及近立体视锐度.两组中波前像差和近立体视锐度比较分别采用单因素方差分析,CS和GS比较采用析因设计方差分析,两两比较采用ISD检验.结果 术后1个月观察发现,ReZoom NXG1 IOL眼、Tecnis ZM900 IOL眼的球差明显低于Sensar AR40e IOL眼的球差(F=11.734;P=0.001,0.000),三者之间的总体像差和高阶像差间的差异无统计学意义(F=0.724,0.173;P=0.493.0.842).ReZoom NXG1 IOL眼、Tecnis ZM900 IOL眼和Sensar R40e IOL眼在各视角的CS及CS比较,差异均无统计学意义(CS:F=0.959,0.978,2.779,0.590,0.485,0175;
P=0.398,0.388,0.079,0.564,0.625,0.847;GS:F=0.117,0.479,0.806,1.235,0.531,1.144;P=0.902,0.627,0.458,0.309,0.597,0.383).MIOL组患者术后近立体视锐度可达45.6弧秒,与SIOL组比较差异有统计学意义(F=2.923;P=0.010,0.014).结论 双眼联合植入折射型与衍射型IOL,在一定程度上可以使患者获得良好的视功能.
目的 探討超聲乳化白內障吸除聯閤雙眼分彆植入摺射型與衍射型多焦點人工晶狀體(MIOL)術後的波前像差、對比敏感度(CS)及立體視功能.方法 前瞻性研究.將年齡相關性白內障患者分為MIOL組與單焦點人工晶狀體(SIOL)組,每組各15例(30隻眼),行超聲乳化白內障吸除聯閤人工晶狀體(IOL)植入術.MIOL組為主導眼植入ReZoom NXG1 IOL(即NXG1組),對側眼植入Tecnis ZM900 IOL(即ZM900組);SIOL組為同期雙眼植入Sensar AR40e IOL(即AR40e組).觀察患者術後波前像差、CS、眩光敏感度(GS)及近立體視銳度.兩組中波前像差和近立體視銳度比較分彆採用單因素方差分析,CS和GS比較採用析因設計方差分析,兩兩比較採用ISD檢驗.結果 術後1箇月觀察髮現,ReZoom NXG1 IOL眼、Tecnis ZM900 IOL眼的毬差明顯低于Sensar AR40e IOL眼的毬差(F=11.734;P=0.001,0.000),三者之間的總體像差和高階像差間的差異無統計學意義(F=0.724,0.173;P=0.493.0.842).ReZoom NXG1 IOL眼、Tecnis ZM900 IOL眼和Sensar R40e IOL眼在各視角的CS及CS比較,差異均無統計學意義(CS:F=0.959,0.978,2.779,0.590,0.485,0175;
P=0.398,0.388,0.079,0.564,0.625,0.847;GS:F=0.117,0.479,0.806,1.235,0.531,1.144;P=0.902,0.627,0.458,0.309,0.597,0.383).MIOL組患者術後近立體視銳度可達45.6弧秒,與SIOL組比較差異有統計學意義(F=2.923;P=0.010,0.014).結論 雙眼聯閤植入摺射型與衍射型IOL,在一定程度上可以使患者穫得良好的視功能.
목적 탐토초성유화백내장흡제연합쌍안분별식입절사형여연사형다초점인공정상체(MIOL)술후적파전상차、대비민감도(CS)급입체시공능.방법 전첨성연구.장년령상관성백내장환자분위MIOL조여단초점인공정상체(SIOL)조,매조각15례(30지안),행초성유화백내장흡제연합인공정상체(IOL)식입술.MIOL조위주도안식입ReZoom NXG1 IOL(즉NXG1조),대측안식입Tecnis ZM900 IOL(즉ZM900조);SIOL조위동기쌍안식입Sensar AR40e IOL(즉AR40e조).관찰환자술후파전상차、CS、현광민감도(GS)급근입체시예도.량조중파전상차화근입체시예도비교분별채용단인소방차분석,CS화GS비교채용석인설계방차분석,량량비교채용ISD검험.결과 술후1개월관찰발현,ReZoom NXG1 IOL안、Tecnis ZM900 IOL안적구차명현저우Sensar AR40e IOL안적구차(F=11.734;P=0.001,0.000),삼자지간적총체상차화고계상차간적차이무통계학의의(F=0.724,0.173;P=0.493.0.842).ReZoom NXG1 IOL안、Tecnis ZM900 IOL안화Sensar R40e IOL안재각시각적CS급CS비교,차이균무통계학의의(CS:F=0.959,0.978,2.779,0.590,0.485,0175;
P=0.398,0.388,0.079,0.564,0.625,0.847;GS:F=0.117,0.479,0.806,1.235,0.531,1.144;P=0.902,0.627,0.458,0.309,0.597,0.383).MIOL조환자술후근입체시예도가체45.6호초,여SIOL조비교차이유통계학의의(F=2.923;P=0.010,0.014).결론 쌍안연합식입절사형여연사형IOL,재일정정도상가이사환자획득량호적시공능.
Objective To evaluate the visual quality after combined implantation of refractive and diffractive multifocal intraocular lenses (IOL) in both eyes.Methods Prospective,nonrandomized study.Age-related cataract patients were divided into multifocal IOL group (MIOL group) and monofocal IOL group (SIOL group) and received phacoemulsification.In MIOL group,a ReZoom NXG1 IOL and a Tecnis Z M900 IOL were implanted into the dominant eye and the fellow eye,respectively.In SIOL group,Sensar AR40e IOLs were implanted into both eyes.Monocular or binocular test was performed one month after surgery,including wavefront aberration,contrast sensitivity with or without glare and near stereoacuity.Results The root mean square of spherical aberration at pupil diameter of 5 mm was 0.058±0.159 μm in NXG1 eyes and 0.005±0.169 μm in ZM900 eyes,both of them were significantly lower than 0.408±0.160 μm achieved in AR40e eyes (F=11.734;P=0.001,0.000).There was no significant difference in the total ocular aberration,contrast sensitivity and high-order aberration among eyes with these different IOLs.The uncorrected near stereoacuity in MIOL group was (45.60±35.04) seconds of arc,which was significantly better than (110.00±41.23) seconds of arc achieved in SIOL group (F=2.923;P=0.010).Conclusion Combined implantation of muhifocal ReZoom NXG1 IOL and Tecnis ZM900 IOL provides a good visual quality with reduced spherical aberration and satisfactory near stereoacuity.