中华眼视光学与视觉科学杂志
中華眼視光學與視覺科學雜誌
중화안시광학여시각과학잡지
CHINESE JOURNAL OF OPTOMETRY OPHTHALMOLOGY AND VISUAL SCIENCE
2013年
9期
547-550
,共4页
潘超%黄锦海%林华尤%华焱军%杨凤%谭维娜%王勤美
潘超%黃錦海%林華尤%華焱軍%楊鳳%譚維娜%王勤美
반초%황금해%림화우%화염군%양봉%담유나%왕근미
前房深度%Galilei Ⅱ%Pentacam%IOLMaster
前房深度%Galilei Ⅱ%Pentacam%IOLMaster
전방심도%Galilei Ⅱ%Pentacam%IOLMaster
Anterior chamber depth%Galilei Ⅱ%Pentacam%IOLMaster
目的 评估Galilei Ⅱ、Pentacam和IOLMaster测量近视人群前房深度(ACD)的差异性、相关性和一致性.方法 前瞻性系列病例研究.40名近视志愿者(40眼),等效球镜度为-1.00~-8.00 D,平均(-4.40±1.80)D.分别应用Galilei Ⅱ、Pentacam和IOLMaster 3种仪器测量ACD,检查顺序以随机方式确定.采用Bonferroni矫正的重复测量方差分析比较3种仪器测得ACD的差异性,Pearson相关分析3种仪器测得ACD的相关性,Bland-Altman法评估3种仪器测得ACD的一致性.结果 Galilei Ⅱ、Pentacam和IOLMaster测得的ACD均值分别为(3.81±0.20)mm,(3.72±0.20)mm和(3.71±0.20)mm.3种仪器测得的ACD均有密切的相关性(r=0.984、0.970、0.968,P<0.01).Galilei Ⅱ和Pentacam测得的ACD差异为(0.10±0.04)mm,差异有统计学意义(P<0.01);Galilei Ⅱ和IOLMaster测得的ACD差异为(0.10±0.05)mm,差异有统计学意义(P<0.01);Pentacam和IOLMaster测得的ACD差异为(0.00±0.05)mm,差异无统计学意义(P>0.05).Galilei Ⅱ和Pentacam,Galilei Ⅱ和IOLMaster,Pentacam和IOLMaster测得的ACD的95%一致性界限分别为0.026~0.167 mm、0.004~0.195 mm和-0.096~0.102 mm.结论 测量近视人群时,Pentacam跟IOLMaster测得的ACD非常接近,Galilei Ⅱ比Pentaeam和IOLMaster测得的ACD稍深,差异在临床可以接受的范围内,并且具有高度相关性和一致性,因此3种仪器在测量近视人群ACD时可以相互替代使用.
目的 評估Galilei Ⅱ、Pentacam和IOLMaster測量近視人群前房深度(ACD)的差異性、相關性和一緻性.方法 前瞻性繫列病例研究.40名近視誌願者(40眼),等效毬鏡度為-1.00~-8.00 D,平均(-4.40±1.80)D.分彆應用Galilei Ⅱ、Pentacam和IOLMaster 3種儀器測量ACD,檢查順序以隨機方式確定.採用Bonferroni矯正的重複測量方差分析比較3種儀器測得ACD的差異性,Pearson相關分析3種儀器測得ACD的相關性,Bland-Altman法評估3種儀器測得ACD的一緻性.結果 Galilei Ⅱ、Pentacam和IOLMaster測得的ACD均值分彆為(3.81±0.20)mm,(3.72±0.20)mm和(3.71±0.20)mm.3種儀器測得的ACD均有密切的相關性(r=0.984、0.970、0.968,P<0.01).Galilei Ⅱ和Pentacam測得的ACD差異為(0.10±0.04)mm,差異有統計學意義(P<0.01);Galilei Ⅱ和IOLMaster測得的ACD差異為(0.10±0.05)mm,差異有統計學意義(P<0.01);Pentacam和IOLMaster測得的ACD差異為(0.00±0.05)mm,差異無統計學意義(P>0.05).Galilei Ⅱ和Pentacam,Galilei Ⅱ和IOLMaster,Pentacam和IOLMaster測得的ACD的95%一緻性界限分彆為0.026~0.167 mm、0.004~0.195 mm和-0.096~0.102 mm.結論 測量近視人群時,Pentacam跟IOLMaster測得的ACD非常接近,Galilei Ⅱ比Pentaeam和IOLMaster測得的ACD稍深,差異在臨床可以接受的範圍內,併且具有高度相關性和一緻性,因此3種儀器在測量近視人群ACD時可以相互替代使用.
목적 평고Galilei Ⅱ、Pentacam화IOLMaster측량근시인군전방심도(ACD)적차이성、상관성화일치성.방법 전첨성계렬병례연구.40명근시지원자(40안),등효구경도위-1.00~-8.00 D,평균(-4.40±1.80)D.분별응용Galilei Ⅱ、Pentacam화IOLMaster 3충의기측량ACD,검사순서이수궤방식학정.채용Bonferroni교정적중복측량방차분석비교3충의기측득ACD적차이성,Pearson상관분석3충의기측득ACD적상관성,Bland-Altman법평고3충의기측득ACD적일치성.결과 Galilei Ⅱ、Pentacam화IOLMaster측득적ACD균치분별위(3.81±0.20)mm,(3.72±0.20)mm화(3.71±0.20)mm.3충의기측득적ACD균유밀절적상관성(r=0.984、0.970、0.968,P<0.01).Galilei Ⅱ화Pentacam측득적ACD차이위(0.10±0.04)mm,차이유통계학의의(P<0.01);Galilei Ⅱ화IOLMaster측득적ACD차이위(0.10±0.05)mm,차이유통계학의의(P<0.01);Pentacam화IOLMaster측득적ACD차이위(0.00±0.05)mm,차이무통계학의의(P>0.05).Galilei Ⅱ화Pentacam,Galilei Ⅱ화IOLMaster,Pentacam화IOLMaster측득적ACD적95%일치성계한분별위0.026~0.167 mm、0.004~0.195 mm화-0.096~0.102 mm.결론 측량근시인군시,Pentacam근IOLMaster측득적ACD비상접근,Galilei Ⅱ비Pentaeam화IOLMaster측득적ACD초심,차이재림상가이접수적범위내,병차구유고도상관성화일치성,인차3충의기재측량근시인군ACD시가이상호체대사용.
Objective To assess the agreement of ACD measurements among Galilei Ⅱ,Pentacam and IOLMaster in myopic eyes.Methods Forty eyes of forty healthy myopia volunteers,whose spherical equivalent were-1.00~-8.00 D,with an average of-4.40±1.80 D,were enrolled in this prospective study.ACD were randomly measured by Galilei Ⅱ,Pentacam and IOLMaster.The differences between ACD measurements by the 3 devices were analyzed by Bonferroni-corrected repeated-measures analysis of variance.Pearson correlation and Bland-Altman analysis were used to assess the correlation and agreement among the 3 measurements.Results The ACD (mean±standard deviation) with Galilei Ⅱ,Pentacam and IOLMaster were 3.81±0.20 mm,3.72±0.20 mm and 3.71± 0.20 mm,respectively.Good correlation were found between any two of 3 devices (r=0.984,0.970,0.968,P<0.01).The difference between Galilei Ⅱ and Pentacam was 0.10±0.04 mm and statistically significant (P<0.01).The difference between Galilei Ⅱ and IOLMaster was 0.10±0.05 mm and statistically significant (P<0.01).The difference between Pentacam and IOLMaster was 0.00±0.05 mm without statistically significant difference (P>0.05).The 95% limits of agreement between Galilei Ⅱ and Pentacam,Galilei Ⅱ and IOLMaster,Pentacam and IOLMaster were 0.026~0.167 mm,0.004~ 0.195 mm and-0.096~0.102 mm,respectively.Conclusion In myopic eyes,Pentacam measured similar ACD values with IOLMaster.Galilei Ⅱ measured slightly deeper ACD than either Pentacam or IOLMaster,the differences were within clinically acceptable level.Therefore,they can be regarded clinically interchangeable in measuring myopic eyes.