曹吉星%刘苏冰%聂晓丽%买志彬%唐秀侠%马宁%孙红燕
曹吉星%劉囌冰%聶曉麗%買誌彬%唐秀俠%馬寧%孫紅燕
조길성%류소빙%섭효려%매지빈%당수협%마저%손홍연
Lenstar LS900%IOL Master%Orbscan-Ⅱ%角膜直径%前房深度
Lenstar LS900%IOL Master%Orbscan-Ⅱ%角膜直徑%前房深度
Lenstar LS900%IOL Master%Orbscan-Ⅱ%각막직경%전방심도
Lenstar LS900%IOL Master%Orbscan-Ⅱ%Keratometry%Corneal diameter%Anterior chamber depth
目的 探讨Lenstar LS900、IOL Master、Orbscan-Ⅱ三种不同仪器测角膜直径和前房深度有无差异.方法 对2013年2~3月在武警河南总队医院屈光手术中心就诊的近视患者63例(126只眼),分别采用Lenstar LS900、IOL Master、Orbscan-Ⅱ测量角膜直径和前房深度,三种仪器均测量3次及以上,取其平均值,并对所得数据进行统计学分析.结果 (1)差异性:三种仪器测量角膜直径值分别为(11.95±0.38) mm、(11.84±0.33) mm、(11.52±0.31) mm,差异均有统计学意义(F=193.15,P=0.004),用SNK-q多个均数间全面比较差异均有统计学意义(Pall≤0.05);测得前房深度值分别为(3.68±0.21) mm、(3.67±0.21) mm、(3.60±0.20) mm,差异均有统计学意义(F =3351.40,P =0.001),用SNK-q多个均数间全面比较,Lenstar LS900与IOL Master测量差异均无统计学意义(q=64.21,P=0.135),Orbscan-Ⅱ测量与Lenstar LS900、IOL Master相比差异均有统计学意义(Pall≤0.05).(2)相关性:角膜直径测量中,相关分析显示三种仪器的测量结果均有显著相关关系,相关系数分别为0.795(P=0.008)、0.754 (P =0.014)、0.637(P=0.019);前房深度测量中,相关分析显示三种仪器测量均有显著相关关系,相关系数分别为0.923 (P =0.010)、0.895 (P =0.013)、0.905 (P =0.023).(3)一致性:Bland-Ahman分析显示三种设备测角膜直径和前房深度一致性较好.结论 角膜直径和前房深度测量,Lenstar LS900与IOL Master测量值相近,相对Orbscan-Ⅱ较大;三种设备之间均有较高的正相关性;三种设备有较好的一致性,其中Orbscan-Ⅱ重复性最好.
目的 探討Lenstar LS900、IOL Master、Orbscan-Ⅱ三種不同儀器測角膜直徑和前房深度有無差異.方法 對2013年2~3月在武警河南總隊醫院屈光手術中心就診的近視患者63例(126隻眼),分彆採用Lenstar LS900、IOL Master、Orbscan-Ⅱ測量角膜直徑和前房深度,三種儀器均測量3次及以上,取其平均值,併對所得數據進行統計學分析.結果 (1)差異性:三種儀器測量角膜直徑值分彆為(11.95±0.38) mm、(11.84±0.33) mm、(11.52±0.31) mm,差異均有統計學意義(F=193.15,P=0.004),用SNK-q多箇均數間全麵比較差異均有統計學意義(Pall≤0.05);測得前房深度值分彆為(3.68±0.21) mm、(3.67±0.21) mm、(3.60±0.20) mm,差異均有統計學意義(F =3351.40,P =0.001),用SNK-q多箇均數間全麵比較,Lenstar LS900與IOL Master測量差異均無統計學意義(q=64.21,P=0.135),Orbscan-Ⅱ測量與Lenstar LS900、IOL Master相比差異均有統計學意義(Pall≤0.05).(2)相關性:角膜直徑測量中,相關分析顯示三種儀器的測量結果均有顯著相關關繫,相關繫數分彆為0.795(P=0.008)、0.754 (P =0.014)、0.637(P=0.019);前房深度測量中,相關分析顯示三種儀器測量均有顯著相關關繫,相關繫數分彆為0.923 (P =0.010)、0.895 (P =0.013)、0.905 (P =0.023).(3)一緻性:Bland-Ahman分析顯示三種設備測角膜直徑和前房深度一緻性較好.結論 角膜直徑和前房深度測量,Lenstar LS900與IOL Master測量值相近,相對Orbscan-Ⅱ較大;三種設備之間均有較高的正相關性;三種設備有較好的一緻性,其中Orbscan-Ⅱ重複性最好.
목적 탐토Lenstar LS900、IOL Master、Orbscan-Ⅱ삼충불동의기측각막직경화전방심도유무차이.방법 대2013년2~3월재무경하남총대의원굴광수술중심취진적근시환자63례(126지안),분별채용Lenstar LS900、IOL Master、Orbscan-Ⅱ측량각막직경화전방심도,삼충의기균측량3차급이상,취기평균치,병대소득수거진행통계학분석.결과 (1)차이성:삼충의기측량각막직경치분별위(11.95±0.38) mm、(11.84±0.33) mm、(11.52±0.31) mm,차이균유통계학의의(F=193.15,P=0.004),용SNK-q다개균수간전면비교차이균유통계학의의(Pall≤0.05);측득전방심도치분별위(3.68±0.21) mm、(3.67±0.21) mm、(3.60±0.20) mm,차이균유통계학의의(F =3351.40,P =0.001),용SNK-q다개균수간전면비교,Lenstar LS900여IOL Master측량차이균무통계학의의(q=64.21,P=0.135),Orbscan-Ⅱ측량여Lenstar LS900、IOL Master상비차이균유통계학의의(Pall≤0.05).(2)상관성:각막직경측량중,상관분석현시삼충의기적측량결과균유현저상관관계,상관계수분별위0.795(P=0.008)、0.754 (P =0.014)、0.637(P=0.019);전방심도측량중,상관분석현시삼충의기측량균유현저상관관계,상관계수분별위0.923 (P =0.010)、0.895 (P =0.013)、0.905 (P =0.023).(3)일치성:Bland-Ahman분석현시삼충설비측각막직경화전방심도일치성교호.결론 각막직경화전방심도측량,Lenstar LS900여IOL Master측량치상근,상대Orbscan-Ⅱ교대;삼충설비지간균유교고적정상관성;삼충설비유교호적일치성,기중Orbscan-Ⅱ중복성최호.
Objective To compare the differences of corneal diameter and anterior chamber depth measured by the Lenstar LS900,IOL Master and Orbscan-Ⅱ.Methods Sixty-three patients with myopia (126 eyes) were involved,and they were the ones who came to Henan Crops Hospital of Armed Police Force in February to March in 2013.Sequential measurements of corneal diameter and anterior chamber depth were performed by the same operator with Lenstar LS900,IOL Master and Orbscan-Ⅱ.Three instruments were measured for 3 times or more.The average value and the data were obtained and statistically analyzed.Results 1.difference:The averages of WTW measured by three instruments were (11.95±0.38)mm,(11.84±0.33)mm and (11.52±0.31)mm.The difference of corneal diameter measured by three different instruments was statistically significant (F =193.15,P =0.004),and the differences between each two instruments were statistically significant,too (Pall≤ 0.05); The averages of ACD measured by Lenstar LS900,IOL Master and Orbscan-Ⅱ were (3.68± 0.21)mm,(3.67±0.21)mm and (3.60±0.20)mm.The difference of anterior chamber depth measured by three consistence instruments was statistically significant (F =3351.40,P =0.001).The difference between Lenstar LS900 and IOL Master was not statistically significant (q=64.21,P =0.135),but there were statistical significance in difference between Orbscan-Ⅱ and Lenstar LS900 (Pall≤0.05).The comparison between IOL Master and Orbscan-Ⅱ was the same result.2.correlation:Positive correlationship existed each two WTW results measured by Lenstar LS900,IOL Master and Orbscan-ii,and the correlation coefficients were 0.795 (P =0.008),0.754 (P =0.014),0.637 (P =0.019); There were also positive correlationship between ACD results measured by Lenstar LS900,IOL Master and Orbscan-Ⅱ,and the correlation coefficients were 0.923 (P =0.010),0.895 (P =0.013),0.905 (P =0.023).3.consistence:Bland-Altman analysis showed that 95% consistency limit of WTW and ACD measured by three different instruments were both good.Conclusions The results of Lenstar LS900 and IOL Master are the most similar in the measurement of WTW and ACD,and they are larger than that of Orbscan-Ⅱ.The three instruments have high positive correlation and consistency.The instrument which has the best repeatability is Orbscan-Ⅱ.