CT肺动脉成像%辐射剂量%管电压%X射线计算机体层摄影
CT肺動脈成像%輻射劑量%管電壓%X射線計算機體層攝影
CT폐동맥성상%복사제량%관전압%X사선계산궤체층섭영
CT pulmonary angiography%Radiation dose%Tube voltage%Tomography
目的 探讨80 kV低管电压对128层螺旋CT肺动脉成像(CTPA)图像质量及辐射剂量的影响.方法 对临床怀疑肺栓塞患者60例行128层螺旋CT肺动脉成像检查,并采用随机数字表法分为低管电压80 kV组和常规管电压120 kV组各30例,两组均开启自动管电流调制技术.记录两组的CT容积剂量指数(CTDIvol)、剂量长度乘积(DLP),计算有效剂量(E)值;测量图像背景噪声(BN)、肺动脉强化CT值(SI),计算信噪比(SNR)、对比噪声比(CNR).比较两组的有效剂量E、SNR、CNR以及图像质量.结果 80 kV组的E为(0.99 ±0.27) mSv,显著低于120 kV组的(3.02±0.87)mSv(t=12.281,P<0.05).80 kV组SI、BN均显著高于120 kV组(t=-3.377、-5.855,P<0.05),SNR、CNR和图像质量评分差异均无统计学意义(P>0.05).结论 与常规管电压120 kV扫描方案相比,采用低管电压80 kV,结合自动管电流调制技术,可以在保证图像质量的同时,有效降低辐射剂量,在多层螺旋CT肺动脉成像中具有较高的临床应用价值.
目的 探討80 kV低管電壓對128層螺鏇CT肺動脈成像(CTPA)圖像質量及輻射劑量的影響.方法 對臨床懷疑肺栓塞患者60例行128層螺鏇CT肺動脈成像檢查,併採用隨機數字錶法分為低管電壓80 kV組和常規管電壓120 kV組各30例,兩組均開啟自動管電流調製技術.記錄兩組的CT容積劑量指數(CTDIvol)、劑量長度乘積(DLP),計算有效劑量(E)值;測量圖像揹景譟聲(BN)、肺動脈彊化CT值(SI),計算信譟比(SNR)、對比譟聲比(CNR).比較兩組的有效劑量E、SNR、CNR以及圖像質量.結果 80 kV組的E為(0.99 ±0.27) mSv,顯著低于120 kV組的(3.02±0.87)mSv(t=12.281,P<0.05).80 kV組SI、BN均顯著高于120 kV組(t=-3.377、-5.855,P<0.05),SNR、CNR和圖像質量評分差異均無統計學意義(P>0.05).結論 與常規管電壓120 kV掃描方案相比,採用低管電壓80 kV,結閤自動管電流調製技術,可以在保證圖像質量的同時,有效降低輻射劑量,在多層螺鏇CT肺動脈成像中具有較高的臨床應用價值.
목적 탐토80 kV저관전압대128층라선CT폐동맥성상(CTPA)도상질량급복사제량적영향.방법 대림상부의폐전새환자60례행128층라선CT폐동맥성상검사,병채용수궤수자표법분위저관전압80 kV조화상규관전압120 kV조각30례,량조균개계자동관전류조제기술.기록량조적CT용적제량지수(CTDIvol)、제량장도승적(DLP),계산유효제량(E)치;측량도상배경조성(BN)、폐동맥강화CT치(SI),계산신조비(SNR)、대비조성비(CNR).비교량조적유효제량E、SNR、CNR이급도상질량.결과 80 kV조적E위(0.99 ±0.27) mSv,현저저우120 kV조적(3.02±0.87)mSv(t=12.281,P<0.05).80 kV조SI、BN균현저고우120 kV조(t=-3.377、-5.855,P<0.05),SNR、CNR화도상질량평분차이균무통계학의의(P>0.05).결론 여상규관전압120 kV소묘방안상비,채용저관전압80 kV,결합자동관전류조제기술,가이재보증도상질량적동시,유효강저복사제량,재다층라선CT폐동맥성상중구유교고적림상응용개치.
Objective To investigate the influence of 80 kV on dose reduction and image quality in 128-slice CT pulmonary angiography (CTPA) compared with the conventional 120 kV.Methods Sixty patients suspected of pulmonary embolism (PE) who underwent CTPA were randomly divided into 80 kV and 120 kV groups by random digits table method.All the patients underwent CTPA with automatic tube current modulation.The values of volume CT dose index (CTDIvol),dose length product (DLP) were recorded and effective dose (E) was calculated.The attenuation was measured in central and peripheral pulmonary arteries as well as background noise (BN),then signal-to-noise-ratio (SNR) and contrast-tonoise-ratio (CNR) were calculated.The radiation exposure,pulmonary arterial enhancement,background noise,SNR,CNR and scores of image quality by a 5-point scale were compared between two groups.Results E value was significantly lower at 80 kV [(0.99 ± 0.27)mSv] compared with 120 kV [(3.02 ± 0.87) mSv,t =12.281,P < 0.05].The mean attenuation value of pulmonary arteries as well as BN in 80 kV group were significantly higher than in 120 kV group (P < 0.05).SNR and CNR did not differ significantly between two groups.No significant difference was detected on scores of image quality (P > 0.05).Conclusions The application of 80 kV protocol combined with automatic tube current modulation scanning in 128-slice MSCT pulmonary angiography could significantly reduce radiation dose compared with the conventional 120 kV protocol without deterioration in diagnostic image quality.