Low dose%Tomography%X-ray computed%Rabbit brain perfusion imaging%Effective dose
目的 评价70 kV管电压低剂量扫描在兔脑CT灌注成像的可行性.方法 健康新西兰大白兔20只,每只兔分别使用70、80 kV管电压行兔脑CT灌注检查,两次检查间隔时间为24 h.比较两组方案有效剂量(E)、脑血流量(CBF)、脑血容量(CBV)、平均通过时间(MTT)及图像质量指标信噪比(SNR).结果 70与80 kV组兔脑CT灌注的有效剂量分别为1.91和2.93 mSv,70 kV组较80 kV组减少约34.8%.两组扫描所测量脑灌注参数CBF、CBV、MTT间的差异无统计学意义.两组测量的SNR分别为(23.15±5.98)和(21.23±9.18),差异无统计学意义.结论 使用70 kV管电压行兔脑CT灌注成像可降低有效剂量,而对灌注参数结果及图像质量无明显影响.
目的 評價70 kV管電壓低劑量掃描在兔腦CT灌註成像的可行性.方法 健康新西蘭大白兔20隻,每隻兔分彆使用70、80 kV管電壓行兔腦CT灌註檢查,兩次檢查間隔時間為24 h.比較兩組方案有效劑量(E)、腦血流量(CBF)、腦血容量(CBV)、平均通過時間(MTT)及圖像質量指標信譟比(SNR).結果 70與80 kV組兔腦CT灌註的有效劑量分彆為1.91和2.93 mSv,70 kV組較80 kV組減少約34.8%.兩組掃描所測量腦灌註參數CBF、CBV、MTT間的差異無統計學意義.兩組測量的SNR分彆為(23.15±5.98)和(21.23±9.18),差異無統計學意義.結論 使用70 kV管電壓行兔腦CT灌註成像可降低有效劑量,而對灌註參數結果及圖像質量無明顯影響.
목적 평개70 kV관전압저제량소묘재토뇌CT관주성상적가행성.방법 건강신서란대백토20지,매지토분별사용70、80 kV관전압행토뇌CT관주검사,량차검사간격시간위24 h.비교량조방안유효제량(E)、뇌혈류량(CBF)、뇌혈용량(CBV)、평균통과시간(MTT)급도상질량지표신조비(SNR).결과 70여80 kV조토뇌CT관주적유효제량분별위1.91화2.93 mSv,70 kV조교80 kV조감소약34.8%.량조소묘소측량뇌관주삼수CBF、CBV、MTT간적차이무통계학의의.량조측량적SNR분별위(23.15±5.98)화(21.23±9.18),차이무통계학의의.결론 사용70 kV관전압행토뇌CT관주성상가강저유효제량,이대관주삼수결과급도상질량무명현영향.
Objective To evaluate the feasibility of low dose CT perfusion imaging of brain using the tube voltage of 70 kV.Methods Twenty New Zealand white rabbits underwent cerebral CT perfusion imaging with tube voltage 70 and 80 kV separately,and the interval between two scans was 24 h.The effective dosage (E),cerebral perfusion parameters (CBF,CBV,MTT) and image quality indicator (SNR) of middle cerebral artery of each protocol were acquired and compared statistically.Results The effective dosage of70 kV group was 1.91 mSv,34.8% lower than that of 80 kV group with the dose of 2.93 mSv.There were no significant differences between the cerebral perfusion parameters acquired at different corresponding ROI.SNR of 70 kV group and 80 kV group were (23.15 ± 5.98) and (21.23 ±9.18) with no significantly statistical difference.Conclusions The effective dosage of 70 kV cerebral CT perfusion imaging could be dramatically reduced with no significant influence on the results of perfusion parameters and image quality.