CHINESE JOURNAL OF RADIATION ONCOLOGY
2013年
2期
154-156
,共3页
董洋%李丰彤%王境生%袁智勇%宋勇春%庄洪卿
董洋%李豐彤%王境生%袁智勇%宋勇春%莊洪卿
동양%리봉동%왕경생%원지용%송용춘%장홍경
射波刀%图像引导放疗%吸收剂量
射波刀%圖像引導放療%吸收劑量
사파도%도상인도방료%흡수제량
Cyberknife%Image guided radiotherapy%Absorbed dose
目的 定量评价G3型射波刀图像引导系统千伏级X线球管照射产生的额外电离辐射吸收剂量.方法 将PTW30010电离室插入CIRS Model 002LFC和002PRA体模不同位置、不同密度各5处组织等效模体插件中,测量不同曝光条件下单次曝光所产生的吸收剂量.通过以上数据定量评估300例不同部位、不同引导方式患者的曝光条件和平均曝光次数以及吸收剂量.结果 平均分割治疗次数为3.94次,中位数4次.采用静态追踪模式患者平均每位每次分割治疗接受曝光次数为36.5次,采用动态追踪模式时为49.2次.单次曝光最小吸收剂量为0.5 μGy,最大为385μGy.结论 射波刀的千伏级图像引导系统在实现3种图像引导功能的同时,对患者产生额外吸收剂量保持在一个很低范围内.
目的 定量評價G3型射波刀圖像引導繫統韆伏級X線毬管照射產生的額外電離輻射吸收劑量.方法 將PTW30010電離室插入CIRS Model 002LFC和002PRA體模不同位置、不同密度各5處組織等效模體插件中,測量不同曝光條件下單次曝光所產生的吸收劑量.通過以上數據定量評估300例不同部位、不同引導方式患者的曝光條件和平均曝光次數以及吸收劑量.結果 平均分割治療次數為3.94次,中位數4次.採用靜態追蹤模式患者平均每位每次分割治療接受曝光次數為36.5次,採用動態追蹤模式時為49.2次.單次曝光最小吸收劑量為0.5 μGy,最大為385μGy.結論 射波刀的韆伏級圖像引導繫統在實現3種圖像引導功能的同時,對患者產生額外吸收劑量保持在一箇很低範圍內.
목적 정량평개G3형사파도도상인도계통천복급X선구관조사산생적액외전리복사흡수제량.방법 장PTW30010전리실삽입CIRS Model 002LFC화002PRA체모불동위치、불동밀도각5처조직등효모체삽건중,측량불동폭광조건하단차폭광소산생적흡수제량.통과이상수거정량평고300례불동부위、불동인도방식환자적폭광조건화평균폭광차수이급흡수제량.결과 평균분할치료차수위3.94차,중위수4차.채용정태추종모식환자평균매위매차분할치료접수폭광차수위36.5차,채용동태추종모식시위49.2차.단차폭광최소흡수제량위0.5 μGy,최대위385μGy.결론 사파도적천복급도상인도계통재실현3충도상인도공능적동시,대환자산생액외흡수제량보지재일개흔저범위내.
Objective To evaluate quantitatively the extra absorbed dose generated by image guided system of the Cyberknife.Methods The exposure parameters and the average frequency of images acquisition during the treatments with G3 Cyberknife in 300 cases with various tumor locations and tracking modes were collected for analysis ;The measurements of the absorbed doses in the phantom with 5 points after single exposure generated by image guided system in various exposure parameters were performed with equipment of phantom CIRS Model 002LFC and 002 PRA with ion chamber PTW30010 included.Based on the data we collected andmeasured,the conclusion of how much extra absorbed dose generated by image guided system of Cyberknife the patients received during Cyberknife treatments was drawn.Results With an average fractions of 3.94(the median was 4),the average exposure frequency was 36.5 times for static tracking per patient per fraction,while it was 49.2 times for dynamic tracking.The experimental results with various exposure parameters,positions and tissue densities showed that the minimum absorbed dose after single exposure was 0.5 μGy,while the maximum was 385 μGy.Conclusions The image guided system of the Cyberknife can induce quite low absorbed dose in patient,while providing all three types of image guided function.