监测,手术中%麻醉药,吸入%脑电描记术
鑑測,手術中%痳醉藥,吸入%腦電描記術
감측,수술중%마취약,흡입%뇌전묘기술
Monitoring,intraoperative%Anesthetics,inhalation%Electroencephalography
目的 评价Narcotrend指数(NI)与七氟醚吸入麻醉患者麻醉深度的相关性,并与BIS值比较.方法 择期手术患者62例,拟于全身麻醉下进行,ASA分级Ⅰ级或Ⅱ级,性别不限,年龄25~60岁,体重44~ 100kg.气管插管后5 min时吸入七氟醚,于呼气末七氟醚浓度(ETsevo) 0.6%、0.9%、1.3%、1.6%、2.1%、2.4%、2.9%、3.2%、3.8%和4.2%时同步记录NI值和BIS值.结果 ETsevo<2.9%时,其与NI值和BIS值相关系数分别r1=-0.627和r2=-0.669(P<0.O1);ETsevo 2.9%~4.2%时,其与NI值和BIS值相关系数分别r3=-0.608和r4=-0.230(P<0.01).相关系数绝对值比较:r1、r2及r3之间差异无统计学意义(P>0.05),r1、r2、r3大于r4(P<0.01).结论 ETsevo4.2%时,NI值能准确反映患者的麻醉深度;ETsevo 2.9%~ 4.2%时,NI值监测麻醉深度的准确性显著优于BIS值.
目的 評價Narcotrend指數(NI)與七氟醚吸入痳醉患者痳醉深度的相關性,併與BIS值比較.方法 擇期手術患者62例,擬于全身痳醉下進行,ASA分級Ⅰ級或Ⅱ級,性彆不限,年齡25~60歲,體重44~ 100kg.氣管插管後5 min時吸入七氟醚,于呼氣末七氟醚濃度(ETsevo) 0.6%、0.9%、1.3%、1.6%、2.1%、2.4%、2.9%、3.2%、3.8%和4.2%時同步記錄NI值和BIS值.結果 ETsevo<2.9%時,其與NI值和BIS值相關繫數分彆r1=-0.627和r2=-0.669(P<0.O1);ETsevo 2.9%~4.2%時,其與NI值和BIS值相關繫數分彆r3=-0.608和r4=-0.230(P<0.01).相關繫數絕對值比較:r1、r2及r3之間差異無統計學意義(P>0.05),r1、r2、r3大于r4(P<0.01).結論 ETsevo4.2%時,NI值能準確反映患者的痳醉深度;ETsevo 2.9%~ 4.2%時,NI值鑑測痳醉深度的準確性顯著優于BIS值.
목적 평개Narcotrend지수(NI)여칠불미흡입마취환자마취심도적상관성,병여BIS치비교.방법 택기수술환자62례,의우전신마취하진행,ASA분급Ⅰ급혹Ⅱ급,성별불한,년령25~60세,체중44~ 100kg.기관삽관후5 min시흡입칠불미,우호기말칠불미농도(ETsevo) 0.6%、0.9%、1.3%、1.6%、2.1%、2.4%、2.9%、3.2%、3.8%화4.2%시동보기록NI치화BIS치.결과 ETsevo<2.9%시,기여NI치화BIS치상관계수분별r1=-0.627화r2=-0.669(P<0.O1);ETsevo 2.9%~4.2%시,기여NI치화BIS치상관계수분별r3=-0.608화r4=-0.230(P<0.01).상관계수절대치비교:r1、r2급r3지간차이무통계학의의(P>0.05),r1、r2、r3대우r4(P<0.01).결론 ETsevo4.2%시,NI치능준학반영환자적마취심도;ETsevo 2.9%~ 4.2%시,NI치감측마취심도적준학성현저우우BIS치.
Objective To evaluate the correlation between Narcotrend index (NI) and the depth of anesthesia during inhalation anesthesia with sevoflurane and compare NI with BIS values in patients.Methods Sixtytwo ASA physical status Ⅰ or Ⅱ patients,aged 25-60 yr,weighing 44-100kg,scheduled for elective surgeries under general anesthesia,were enrolled in the study.Sevoflurane inhalation was started at 5 min after endotracheal intubation.When the end-tidal concentration of sevoflurane (ETsevo) reached 0.6%,0.9%,1.3%,1.6%,2.1%,2.4%,2.9%,3.2%,3.8% and 4.2%,NI and BIS values were recorded simultaneously.Results The correlation coefficients between ETsevo and NI and BIS values were-0.627 (r1) and-0.669 (r2) respectively when ETsevo < 2.9%,and-0.608 (r3) and-0.230 (r4) respectively when 2.9% ≤ ETsevo ≤4.2% (P < 0.01).Comparison of the absolute value of correlation coefficients:no significant difference was found between r1,r2 and r3 (P > 0.05) ; r1,r2 and r3 were significantly greater than r4 (P < 0.01).Conclusion When ETsevo 4.2%,NI values can accurately reflect the depth of anesthesia; when 2.9% ≤ ETsevo ≤4.2%,the accuracy of NI values in monitoring the depth of anesthesia is superior to that of BIS values.