超声弹性成像%乳腺占位性病变%诊断
超聲彈性成像%乳腺佔位性病變%診斷
초성탄성성상%유선점위성병변%진단
Ultrasound elastography%Breast masses%Diagnosis
目的探讨超声弹性成像(UE)硬度评分法及应变率比值法在诊断乳腺占位性病变中的价值.方法对78例,共156个乳腺占位性病变术前行超声弹性成像检查.测量病变组织与周围组织弹性应变率比值,以3.08为界点判断占位性病变的良恶性.超声弹性成像硬度评分标准采用改良5分法,1-3分为良性,4-5分为恶性.以手术病理为金标准,对比分析超声弹性成像硬度评分法、应变率比值法以及二者联合应用在诊断乳腺占位性病变良恶性的准确性.结果156个占位性病变中,恶性15个,良性141个.超声弹性成像硬度评分法诊断乳腺良恶性病变的准确性91%,灵敏度99.3%,特异性86.7%.弹性应变率比值法诊断乳腺良恶性病变的准确性92.3%,灵敏度99.3%,特异性73.4%.联合应用超声弹性成像硬度评分法及弹性应变率比值法,诊断乳腺占位性病变良恶性的准确性98.1%,灵敏度98.6%,特异性93.3%.结论超声弹性成像硬度评分及弹性应变率比值在乳腺占位性病变良恶性诊断中均有较高价值,但两者结合可以提高其诊断准确性.
目的探討超聲彈性成像(UE)硬度評分法及應變率比值法在診斷乳腺佔位性病變中的價值.方法對78例,共156箇乳腺佔位性病變術前行超聲彈性成像檢查.測量病變組織與週圍組織彈性應變率比值,以3.08為界點判斷佔位性病變的良噁性.超聲彈性成像硬度評分標準採用改良5分法,1-3分為良性,4-5分為噁性.以手術病理為金標準,對比分析超聲彈性成像硬度評分法、應變率比值法以及二者聯閤應用在診斷乳腺佔位性病變良噁性的準確性.結果156箇佔位性病變中,噁性15箇,良性141箇.超聲彈性成像硬度評分法診斷乳腺良噁性病變的準確性91%,靈敏度99.3%,特異性86.7%.彈性應變率比值法診斷乳腺良噁性病變的準確性92.3%,靈敏度99.3%,特異性73.4%.聯閤應用超聲彈性成像硬度評分法及彈性應變率比值法,診斷乳腺佔位性病變良噁性的準確性98.1%,靈敏度98.6%,特異性93.3%.結論超聲彈性成像硬度評分及彈性應變率比值在乳腺佔位性病變良噁性診斷中均有較高價值,但兩者結閤可以提高其診斷準確性.
목적탐토초성탄성성상(UE)경도평분법급응변솔비치법재진단유선점위성병변중적개치.방법대78례,공156개유선점위성병변술전행초성탄성성상검사.측량병변조직여주위조직탄성응변솔비치,이3.08위계점판단점위성병변적량악성.초성탄성성상경도평분표준채용개량5분법,1-3분위량성,4-5분위악성.이수술병리위금표준,대비분석초성탄성성상경도평분법、응변솔비치법이급이자연합응용재진단유선점위성병변량악성적준학성.결과156개점위성병변중,악성15개,량성141개.초성탄성성상경도평분법진단유선량악성병변적준학성91%,령민도99.3%,특이성86.7%.탄성응변솔비치법진단유선량악성병변적준학성92.3%,령민도99.3%,특이성73.4%.연합응용초성탄성성상경도평분법급탄성응변솔비치법,진단유선점위성병변량악성적준학성98.1%,령민도98.6%,특이성93.3%.결론초성탄성성상경도평분급탄성응변솔비치재유선점위성병변량악성진단중균유교고개치,단량자결합가이제고기진단준학성.
Objective To investigate the clinical value of ultrasound elastography (UE) score and the elastic strain rate ratio in the diagnosis of breast masses .Methods 78 patients with breast masses(156 masses) were examined with ultrasound elastography. The masses’ and its surrounding tissues’ elastic strain rate ratio were measured . 3.08 was regard as the boundary points to differentiate the masses to be benign or malignant. The elastogram was graded using 5-score evaluating method. 1-3 points were graded to be benign, and 4-5 points were graded to be malignant. Using the postoperative pathological diagnosis as gold standard to comparative analyse the diagnosis accuracy of ultrasound elastography score and the elastic strain rate ratio in differenting benign and malignant breast masses . Results Of the 156 masses, 15 were malignant and 141 were benign. Accuracy of elastography score in differentiating breast masses was 91%, the sensitivity was 99.3% and the specificity was 86.7%; Accuracy of elastic strain rate ratio in differentiating breast masses was 92.3%, the sensitivity was 99.3% and the specificity was 73.4%;Accuracy of elastography score combined with elastic strain rate ratio in differentiating breast masses was 98.1%, the sensitivity was 98.6% and the specificity was 93.3%. Conclusion Both the elastography score and elastic strain rate ratio was valuable ultrasonographic method in differntiating benign and malignant breast masses. And combining these two can improve the validity in the diagnosis of malignant breast tumors.