目的:对院内便携式血糖仪实行统一的质量管理,实施一系列质量管理措施,使院内血葡萄糖的检测结果具有可靠性及可比性.方法:实行行政制度化管理,成立管理委员会,采取质量控制及质量保证手段,实行严格的比对制度以及实施统一的培训工作和考核制度.结果:行政管理和高质量的培训工作可以得到管理层的高度重视,日常质控品的检测可以发现便携式血糖仪及试纸条是否发生变化,室间质量评价两年来葡萄糖和便携式血糖仪PT成绩100%,无失控样本,比对葡萄糖浓度2.8~22.2 mmol/L的7个浓度范围53个样本,全部合格.结论:对院内便携式血糖仪实施一系列的质量管理措施使院内血葡萄糖的检测结果具有可靠性及可比性,取得一定实际应用效果.
目的:對院內便攜式血糖儀實行統一的質量管理,實施一繫列質量管理措施,使院內血葡萄糖的檢測結果具有可靠性及可比性.方法:實行行政製度化管理,成立管理委員會,採取質量控製及質量保證手段,實行嚴格的比對製度以及實施統一的培訓工作和攷覈製度.結果:行政管理和高質量的培訓工作可以得到管理層的高度重視,日常質控品的檢測可以髮現便攜式血糖儀及試紙條是否髮生變化,室間質量評價兩年來葡萄糖和便攜式血糖儀PT成績100%,無失控樣本,比對葡萄糖濃度2.8~22.2 mmol/L的7箇濃度範圍53箇樣本,全部閤格.結論:對院內便攜式血糖儀實施一繫列的質量管理措施使院內血葡萄糖的檢測結果具有可靠性及可比性,取得一定實際應用效果.
목적:대원내편휴식혈당의실행통일적질량관리,실시일계렬질량관리조시,사원내혈포도당적검측결과구유가고성급가비성.방법:실행행정제도화관리,성립관리위원회,채취질량공제급질량보증수단,실행엄격적비대제도이급실시통일적배훈공작화고핵제도.결과:행정관리화고질량적배훈공작가이득도관리층적고도중시,일상질공품적검측가이발현편휴식혈당의급시지조시부발생변화,실간질량평개량년래포도당화편휴식혈당의PT성적100%,무실공양본,비대포도당농도2.8~22.2 mmol/L적7개농도범위53개양본,전부합격.결론:대원내편휴식혈당의실시일계렬적질량관리조시사원내혈포도당적검측결과구유가고성급가비성,취득일정실제응용효과.
Objective:To conduct a series of measures to execute a unified quality management on hospital portable blood glucose meter,in order to make the test results of hospital blood glucose reliable and comparable. Method:To carry out systematic administrative management,setting up management committee,taking quality control and assurance means,rigorous comparison of the system and the implementation of high-quality training programs. Result:Administration and high-quality management training could be highly appreciated by the hospital administrators. The daily quality control testing could find problems if the portable blood glucose meter and test strip changed. For two years,the room quality assessment of the proficiency testing performance of glucose and the portable blood glucose meter were 100%,which had no loss of control specimens. Comparison of 53 specimens which had 7 different glucose concentration,between 2.8 mmol/L and 22.2 mmol/L,the tests all qualified. Conclusion:To implement a series of intervention on quality control of the hospital portable blood glucose meter,to make results of the hospital blood glucose reliable and comparable and have an effect on practical applications.