目的:探讨多效清洗酶在不同水温下用不同的水稀释浸泡对器械清洗的效果影响。方法将污染后的器械均分成A、B、C 3组(n=210), A组用常温的自来水清洗以及稀释;B组用35℃~50℃的温水匹配相同多的水稀释清洗,在清洗过程中确保温度变化不大;C组用>60℃的高温水清洗,同样保持与B组一样的比例稀释,在清洗过程中确保温度变化不大。在浸泡、清洗、干燥后观察。通过5倍的放大镜用肉眼进行观察,再用潜血试验法检测。结果3组在目测方法检查下无统计学差异;潜血测试A组合格率为79.20%,B组合格率为90.10%,C组合格率为71.70%,3组差异有统计学意义(P<0.05);B组的多效清洗酶用不同的水稀释对比合格率,B1组为90.78%、B2组为71.77%,2组差异有统计学意义(P<0.05)。结论多效清洗酶在35℃~50℃浸泡后清洗的效果最好,在匹配中要保持一定的稀释度,并控制好水温。
目的:探討多效清洗酶在不同水溫下用不同的水稀釋浸泡對器械清洗的效果影響。方法將汙染後的器械均分成A、B、C 3組(n=210), A組用常溫的自來水清洗以及稀釋;B組用35℃~50℃的溫水匹配相同多的水稀釋清洗,在清洗過程中確保溫度變化不大;C組用>60℃的高溫水清洗,同樣保持與B組一樣的比例稀釋,在清洗過程中確保溫度變化不大。在浸泡、清洗、榦燥後觀察。通過5倍的放大鏡用肉眼進行觀察,再用潛血試驗法檢測。結果3組在目測方法檢查下無統計學差異;潛血測試A組閤格率為79.20%,B組閤格率為90.10%,C組閤格率為71.70%,3組差異有統計學意義(P<0.05);B組的多效清洗酶用不同的水稀釋對比閤格率,B1組為90.78%、B2組為71.77%,2組差異有統計學意義(P<0.05)。結論多效清洗酶在35℃~50℃浸泡後清洗的效果最好,在匹配中要保持一定的稀釋度,併控製好水溫。
목적:탐토다효청세매재불동수온하용불동적수희석침포대기계청세적효과영향。방법장오염후적기계균분성A、B、C 3조(n=210), A조용상온적자래수청세이급희석;B조용35℃~50℃적온수필배상동다적수희석청세,재청세과정중학보온도변화불대;C조용>60℃적고온수청세,동양보지여B조일양적비례희석,재청세과정중학보온도변화불대。재침포、청세、간조후관찰。통과5배적방대경용육안진행관찰,재용잠혈시험법검측。결과3조재목측방법검사하무통계학차이;잠혈측시A조합격솔위79.20%,B조합격솔위90.10%,C조합격솔위71.70%,3조차이유통계학의의(P<0.05);B조적다효청세매용불동적수희석대비합격솔,B1조위90.78%、B2조위71.77%,2조차이유통계학의의(P<0.05)。결론다효청세매재35℃~50℃침포후청세적효과최호,재필배중요보지일정적희석도,병공제호수온。
Objective To investigate the cleaning enzyme effect on equipment cleaning effect by water dilution and different soaking in different temperature, in order to understand the most effective use of the degree of dilution water cleaning enzyme and what and how.Methods The pollution of the equipment is divided into A, B, C 3 groups, Group A with our daily cleaning in tap water at room temperature and dilution; group B with 35℃-50℃ warm water matching water dilution cleaning as much, and in the process of cleaning in order to ensure the accuracy of the experiment, ensuring that the temperature changed little during the cleaning process; group C with high temperature water cleaning >60℃, the same as before the dilution proportion,and ensure that the temperature changed little during the cleaning process.Observation in immersion, cleaning, drying. With 5 times the magnifying glasswith the naked eye observation, and then detect the occult blood test.Results The 3 groups in the visual method to check no signiifcant difference; occult bloodtest A combination lattice rate was 79.20%, the qualiifed rate of B group was 90.10%, the qualiifed rate of C group was 71.70%, the test results were statistically signiifcant (P<0.05); then the cleaning enzyme B group of water diluted contrast different qualiifed rate, B1 group 90.78%, B2 group 71.77%, there are signiifcant in this two occult blood test (P<0.05).Conclusion The best cleaning enzyme after 35℃-50℃ immersion cleaning effect, andmaintain a certain degree of dilution as temperature control possible in matchingtime.