IUD%translational force%torque
目的:通过建立实验模型,测试宫内节育器( IUD)与常规医用磁共振( MR)的相互影响,以期确定使用IUD的对象能否进行MRI检查。方法将磁性与非磁性纯铜IUD,分别放置于测试模型上,在0.4 T、1.0 T、1.5 T、3.0 T MR机上测试IUD各点的平移力和扭力。结果磁性金属IUD在0.4 T永磁MR圈内中心磁力最强,平移力最强,力的方向指向圈内中心,与机器是否扫描无关,未能测到扭力。在1.0 T、1.5 T、3.0 T超导型MR外圈平移力明显,近圈缘最高,未测到扭力,圈内保持磁性平衡,合力为零,平移力与扭力为零。非磁性金属铜IUD不存在平移力和扭力。以上数据与机器是否扫描无关。结论金属磁性IUD的位置变化是在进入检查室及进入检查腔的过程中产生,并不是在扫描中产生,在扫描中并无扭力产生;非磁性纯铜IUD在常规医用MR测试无平移力和扭力,检查时不存在受磁力牵拉引起的移位、出血。
目的:通過建立實驗模型,測試宮內節育器( IUD)與常規醫用磁共振( MR)的相互影響,以期確定使用IUD的對象能否進行MRI檢查。方法將磁性與非磁性純銅IUD,分彆放置于測試模型上,在0.4 T、1.0 T、1.5 T、3.0 T MR機上測試IUD各點的平移力和扭力。結果磁性金屬IUD在0.4 T永磁MR圈內中心磁力最彊,平移力最彊,力的方嚮指嚮圈內中心,與機器是否掃描無關,未能測到扭力。在1.0 T、1.5 T、3.0 T超導型MR外圈平移力明顯,近圈緣最高,未測到扭力,圈內保持磁性平衡,閤力為零,平移力與扭力為零。非磁性金屬銅IUD不存在平移力和扭力。以上數據與機器是否掃描無關。結論金屬磁性IUD的位置變化是在進入檢查室及進入檢查腔的過程中產生,併不是在掃描中產生,在掃描中併無扭力產生;非磁性純銅IUD在常規醫用MR測試無平移力和扭力,檢查時不存在受磁力牽拉引起的移位、齣血。
목적:통과건립실험모형,측시궁내절육기( IUD)여상규의용자공진( MR)적상호영향,이기학정사용IUD적대상능부진행MRI검사。방법장자성여비자성순동IUD,분별방치우측시모형상,재0.4 T、1.0 T、1.5 T、3.0 T MR궤상측시IUD각점적평이력화뉴력。결과자성금속IUD재0.4 T영자MR권내중심자력최강,평이력최강,력적방향지향권내중심,여궤기시부소묘무관,미능측도뉴력。재1.0 T、1.5 T、3.0 T초도형MR외권평이력명현,근권연최고,미측도뉴력,권내보지자성평형,합력위령,평이력여뉴력위령。비자성금속동IUD불존재평이력화뉴력。이상수거여궤기시부소묘무관。결론금속자성IUD적위치변화시재진입검사실급진입검사강적과정중산생,병불시재소묘중산생,재소묘중병무뉴력산생;비자성순동IUD재상규의용MR측시무평이력화뉴력,검사시불존재수자력견랍인기적이위、출혈。
Objective To establish experimental model to test the interaction of intrauterine device ( IUD) with conventional med-ical magnetic resonance ( MR) , so as to determine whether the patients using IUD could be checked by MRI or not. Methods Mag-netic IUD and non-magnetic copper IUD were placed on the test model respectively. The translational force and torque of various IUD points were tested on 0. 4T, 1. 0T, 1. 5T and 3. 0T MR machine. Results In the 0. 4 T magnetic metal circle center, magnetic metal IUD showed the strongest magnetic force and translational force. The direction of the force pointed to the circle center and had nothing to do with machine scanning;torque was not tested. Superconductive MR outer translational force was obvious on 1. 0 T, 1. 5 T, 3. 0 MR machines with the highest recent lap margin, torque was not tested, magnetic circle maintained magnetic balance, the forces were zero, translational force and torque were zero. Non-magnetic metal copper IUD didn't have translational and torsional forces. The data above had nothing to do with machine scanning. Conclusion The change of the position of magnetic metal IUD happens in the process of entering into the examination room and cavity instead of scanning;while there is no torque in the process of scanning. Non-magnetic copper IUD didn't have translational force and torque tested by conventional medical magnetic resonance (MR), without shifting and bleeding caused by magnetic pulling during testing.