中国医学装备
中國醫學裝備
중국의학장비
CHINA MEDICAL EQUIPMENT
2014年
12期
56-58
,共3页
迈瑞心电监护仪%误报警%排除方法
邁瑞心電鑑護儀%誤報警%排除方法
매서심전감호의%오보경%배제방법
Mindray electrocardiogram monitor%False alarms%Eliminating the interference
目的:探讨迈瑞心电监护仪的误报警率,并制定干扰排除的措施。方法:选择52例急诊留院观察患者作为对照组,记录心电监护仪误报警的次数及原因,并分析改进措施;另将61例急诊留院观察患者作为实验组,记录改进后记录误报警次数,并与对照组进行比较。结果:对照组患者在心电监护仪的使用中仪器共报警5376次,其中误报警2208次,误报警率为41.07%;实验组患者在心电监护仪使用中仪器共报警2634次,其中误报警798次,误报警率为30.29%,两组相比差异有统计学意义(x2=25.77,P<0.05)。结论:通过对迈瑞心电监护仪进行人工干预,可降低其误报警率,有利于提高医护人员的工作效率,保证患者的治疗安全。
目的:探討邁瑞心電鑑護儀的誤報警率,併製定榦擾排除的措施。方法:選擇52例急診留院觀察患者作為對照組,記錄心電鑑護儀誤報警的次數及原因,併分析改進措施;另將61例急診留院觀察患者作為實驗組,記錄改進後記錄誤報警次數,併與對照組進行比較。結果:對照組患者在心電鑑護儀的使用中儀器共報警5376次,其中誤報警2208次,誤報警率為41.07%;實驗組患者在心電鑑護儀使用中儀器共報警2634次,其中誤報警798次,誤報警率為30.29%,兩組相比差異有統計學意義(x2=25.77,P<0.05)。結論:通過對邁瑞心電鑑護儀進行人工榦預,可降低其誤報警率,有利于提高醫護人員的工作效率,保證患者的治療安全。
목적:탐토매서심전감호의적오보경솔,병제정간우배제적조시。방법:선택52례급진류원관찰환자작위대조조,기록심전감호의오보경적차수급원인,병분석개진조시;령장61례급진류원관찰환자작위실험조,기록개진후기록오보경차수,병여대조조진행비교。결과:대조조환자재심전감호의적사용중의기공보경5376차,기중오보경2208차,오보경솔위41.07%;실험조환자재심전감호의사용중의기공보경2634차,기중오보경798차,오보경솔위30.29%,량조상비차이유통계학의의(x2=25.77,P<0.05)。결론:통과대매서심전감호의진행인공간예,가강저기오보경솔,유리우제고의호인원적공작효솔,보증환자적치료안전。
Objective:To investigate the false alarm rate of electrocardiogram monitor, and to develop methods to exclude interference. Methods:A total of 52 patients in emergency room in hospital in March 2014 were in the control group. The frequency and causes of false alarms of the electrocardiogram monitor were recorded. At the meantime, the improvements were analyzes. A total of 52 patients in emergency room in hospital in May 2014 were chose to be the experimental group. The frequency and causes of false alarms of the electrocardiogram monitor were recorded after some improvements. Results: In control group, the total of alarms rate were 5376 times, including 2208 times of false alarms. The false alarm rate was 41.07%. While in the experimental group, there were 2634 alarms, including 798 times of false alarms. The false alarm rate was 30.29%. Compared to the control group(x2=25.77, P<0.05). It was statistically significant difference. Conclusion:Intervention on Mindray electrocardiogram monitor can reduce the false alarm rate, help to improve the efficiency of nurses and ensure safety to patients.