Rockwel hardness (C Gauge) tester%evaluation of uncertainty in measurement%component of standard uncertainty%combined standard uncertainty%expanded uncertainty
目的:对本实验室洛氏硬度计(C标尺)测量结果的不确定度作以评定。方法:按照JJG112-2003《金属洛氏硬度计(ABCDEFGHKNT标尺)检定规程》,以C标尺为例,考虑洛氏硬度试验中标准硬度块、硬度计及试样等因素对洛氏硬度计(C标尺)测量结果的不确定度进行评定,即通过分析、计算洛氏硬度试验中产生不确定度的若干分量,计算出扩展不确定度。结果:洛氏硬度计测量结果的扩展不确定度对32.5HRC标准硬度块为:y1=y1±U95=33.11HRC±1.078HRC,k=2;对62.3HRC标准硬度块为:y2=y2±U95=63.31HRC±0.710HRC,k=2。结论:在洛氏硬度计检定过程中,测量结果的不确定度主要取决于洛氏硬度计的允许误差。评定结果令人满意,符合洛氏硬度(C标尺)的测量要求。
目的:對本實驗室洛氏硬度計(C標呎)測量結果的不確定度作以評定。方法:按照JJG112-2003《金屬洛氏硬度計(ABCDEFGHKNT標呎)檢定規程》,以C標呎為例,攷慮洛氏硬度試驗中標準硬度塊、硬度計及試樣等因素對洛氏硬度計(C標呎)測量結果的不確定度進行評定,即通過分析、計算洛氏硬度試驗中產生不確定度的若榦分量,計算齣擴展不確定度。結果:洛氏硬度計測量結果的擴展不確定度對32.5HRC標準硬度塊為:y1=y1±U95=33.11HRC±1.078HRC,k=2;對62.3HRC標準硬度塊為:y2=y2±U95=63.31HRC±0.710HRC,k=2。結論:在洛氏硬度計檢定過程中,測量結果的不確定度主要取決于洛氏硬度計的允許誤差。評定結果令人滿意,符閤洛氏硬度(C標呎)的測量要求。
목적:대본실험실락씨경도계(C표척)측량결과적불학정도작이평정。방법:안조JJG112-2003《금속락씨경도계(ABCDEFGHKNT표척)검정규정》,이C표척위례,고필락씨경도시험중표준경도괴、경도계급시양등인소대락씨경도계(C표척)측량결과적불학정도진행평정,즉통과분석、계산락씨경도시험중산생불학정도적약간분량,계산출확전불학정도。결과:락씨경도계측량결과적확전불학정도대32.5HRC표준경도괴위:y1=y1±U95=33.11HRC±1.078HRC,k=2;대62.3HRC표준경도괴위:y2=y2±U95=63.31HRC±0.710HRC,k=2。결론:재락씨경도계검정과정중,측량결과적불학정도주요취결우락씨경도계적윤허오차。평정결과령인만의,부합락씨경도(C표척)적측량요구。
Objection: To evaluate the uncertainty in measurement using Rockwell hardness (C Gauge) tester in our laboratory. Method:According to JJG112-2003 Veriifcation Regulation of Metal ic Rockwel Hardness Testers (Scales A, B, C, D, E, F, G, H, K, N, T), the uncertainty in measurement using Rockwel hardness (C Gauge) tester was evaluated by considering the factors in hardness testing, such as standard block of hardness, hardness tester, the sample and so on. That is to say, the expanded uncertainty was achieved by calculating and analysizing several uncertainty components occurring in the hardness testing. Results: For standard block of hardness 32.5 HRC, the expanded uncertainty in hardness testing was y1=y1±U95=33.11 HRC±1.078 HRC, k=2; for standard block of hardness 62.3 HRC, the expanded uncertainty in hardness testing was y2=y2±U95=63.31 HRC±0.710 HRC, k=2. Conclusion:During the veriifcation of Rockwel hardness tester, the uncertainty in measurement was mostly decided by the permissible error. The evaluation result that was in requirement of Rockwel hardness (C Gauge) testing was satisfying.