计测技术
計測技術
계측기술
METROLOGY & MEASUREMENT TECHNOLOGY
2014年
6期
54-57
,共4页
杜传耀%张春波%马京津%李栋%蔡青
杜傳耀%張春波%馬京津%李棟%蔡青
두전요%장춘파%마경진%리동%채청
积分球%CCD标定%图像处理%线性%面均匀性
積分毬%CCD標定%圖像處理%線性%麵均勻性
적분구%CCD표정%도상처리%선성%면균균성
integrating sphere%CCD calibration%image processing%linear%surface uniformity
积分球对CCD ( Charge-coupled Device)的标定,主要是利用积分球光源具有较好的均匀性、稳定性以及精度高的特点,通过CCD采集积分球光源的图片,将积分球光源亮度转换为CCD拍摄图像的灰度,通过对采集图片灰度信息的提取,得出了CCD标定的各项参数。 CCD线性标定试验的结果表明:在CCD快门值固定时,光源亮度的变化和图像灰度呈线性关系,并且在转换后的不同灰度值区段,线性度有一定的差别:在中值区,线性度最好,低值区次之,高值区最差;在光强一定的情况下, CCD 快门值的改变和采集图像的灰度也呈线性关系。均匀性标定的试验结果表明: CCD的光电响应的非均匀性是随着快门值和灰度的增加而减小的,但最终趋于一个较稳定的值。
積分毬對CCD ( Charge-coupled Device)的標定,主要是利用積分毬光源具有較好的均勻性、穩定性以及精度高的特點,通過CCD採集積分毬光源的圖片,將積分毬光源亮度轉換為CCD拍攝圖像的灰度,通過對採集圖片灰度信息的提取,得齣瞭CCD標定的各項參數。 CCD線性標定試驗的結果錶明:在CCD快門值固定時,光源亮度的變化和圖像灰度呈線性關繫,併且在轉換後的不同灰度值區段,線性度有一定的差彆:在中值區,線性度最好,低值區次之,高值區最差;在光彊一定的情況下, CCD 快門值的改變和採集圖像的灰度也呈線性關繫。均勻性標定的試驗結果錶明: CCD的光電響應的非均勻性是隨著快門值和灰度的增加而減小的,但最終趨于一箇較穩定的值。
적분구대CCD ( Charge-coupled Device)적표정,주요시이용적분구광원구유교호적균균성、은정성이급정도고적특점,통과CCD채집적분구광원적도편,장적분구광원량도전환위CCD박섭도상적회도,통과대채집도편회도신식적제취,득출료CCD표정적각항삼수。 CCD선성표정시험적결과표명:재CCD쾌문치고정시,광원량도적변화화도상회도정선성관계,병차재전환후적불동회도치구단,선성도유일정적차별:재중치구,선성도최호,저치구차지,고치구최차;재광강일정적정황하, CCD 쾌문치적개변화채집도상적회도야정선성관계。균균성표정적시험결과표명: CCD적광전향응적비균균성시수착쾌문치화회도적증가이감소적,단최종추우일개교은정적치。
Integrating sphere calibration of the CCD ( Charge-coupled Device) makes the use of an integrating sphere light source, as it has a good uniformity, stability and high accuracy. CCD is used to capture images of the integrating sphere light source, the integrating sphere light source brightness is converted into image gray scale. Then the parameters of CCD calibration is obtained by extracting the data of gray scale ima-ges. CCD linear calibration test results show that:when the value of the CCD shutter is fixed, the light brightness changes in linear relation with image gray, and the linearity is somewhat different in different gray scale value regions after the conversion. Linearity of the median value region is the best, low value region is second, and high value region is the worst. With some fixed light intensity, shutter changes of CCD are linear with image gray scale. The uniformity calibration test results show that:non-uniformity of CCD photo response decreases with increasing of shutter and image gray, but eventually becomes a more stable value.